BJA Advance Access published online on January 19, 2007
British Journal of Anaesthesia, doi:10.1093/bja/ael361
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
The coiling length of thoracic epidural catheters: the influence of epidural approach angle
Department of Anesthesiology, Seoul National University Hospital, Seoul National University College of Medicine, Seoul, Republic of Korea
* Corresponding author: Department of Anesthesiology, Seoul National University Hospital, 28 Yongon-dong Chongno-gu, Seoul, Republic of Korea 110-744. E-mail: bahkjh{at}snu.ac.kr
BACKGROUND: Thoracic epidural catheters are used for anaesthesia and postoperative analgesia. Usually, epidural catheters are placed without confirmation of their position despite frequent reports of complications as a result of malposition. In this study, we evaluated the threading length of thoracic epidural catheters without coiling and assessed the influence of two different epidural approach angles on the threading length without coiling.
METHODS: Eighty-three patients scheduled for thoracotomy were enrolled and randomly allocated into the acute angle group and the obtuse angle group. In both groups, skin insertion was performed at the T89 intervertebra level. Epidural access was performed under fluoroscopy using a paramedian approach at the T78 level in the acute angle group and at the T67 level in the obtuse angle group, and an end-hole 19-gauge epidural catheter was inserted. Coiling length, defined as the length of the catheter within the epidural space when any part of the catheter just begins to head caudally, was measured in both groups.
RESULTS: The coiling length was 7.4(4.4) cm (95% CI 6.08.7 cm) in the obtuse angle group compared with 4.9(3.3) cm (95% CI 3.86.0 cm) in the acute angle group (P = 0.005).
CONCLUSIONS: Approaching the thoracic epidural space with an obtuse approach angle provides longer coiling length. We recommend that an obtuse approach angle should be used to maximize the chance of the catheter reaching the intended level with minimum risk of coiling.
Keywords: anaesthetic techniques, epidural; approach angle; coiling length; equipment; catheters epidural
This study was presented for poster discussion as an abstract at the 2006 European Society of Anesthesiologists meetings.