Semi-invasive monitoring of cardiac output by a new device using arterial pressure waveform analysis: a comparison with intermittent pulmonary artery thermodilution in patients undergoing cardiac surgery
Department of Anaesthesiology and Intensive Care Medicine, Klinikum Ludwigshafen, Bremserstr. 79, 67063 Ludwigshafen, Germany
* Corresponding author. E-mail: j-mayer{at}gmx.de
BACKGROUND: Thermodilution technique using a pulmonary artery catheter (PAC) is a widely used method to determine cardiac output (CO). It is increasingly criticized because of its invasiveness and its unclear riskbenefit ratio. Thus, less invasive techniques for measuring CO are highly desirable. We compared a new, semi-invasive device (FloTrac/VigileoTM) using arterial pressure waveform analysis for CO measurement in patients undergoing cardiac surgery with bolus thermodilution measurements.
METHODS: Forty patients undergoing coronary artery bypass grafting or valve repair were enrolled. A PAC was inserted and routine radial arterial access was used for semi-invasive determination of CO with the Vigileo. CO was measured simultaneously by bolus thermodilution and the Vigileo technique after induction of anaesthesia (T1), before cardiopulmonary bypass (CPB) (T2), after CPB (T3), after sternal closure (T4), on arrival in the intensive care unit (ICU) (T5), and 4 h (T6), 8 h (T7), and 24 h after surgery (T8). CO was indexed to the body surface area (cardiac index, CI).
RESULTS: A total of 244 pairs of CI measurements were analysed. Bias and precision (1.96 SD of the bias) were 0.46 litre min1 m2 and 1.15 litre min1 m2 (r = 0.53) resulting in an overall percentage error of 46%. Subgroup analysis revealed a percentage error of 51% for data pairs obtained intraoperatively (T1T4), 42% in ICU (T5T8), and 56% for values obtained during low CI (T1T8).
CONCLUSIONS: In cardiac surgery patients, CO measured by a new semi-invasive arterial pressure waveform analysis device showed only moderate agreement with intermittent pulmonary artery thermodilution measurement.
Keywords: measurement techniques; thermodilution; monitoring; cardiopulmonary; surgery; cardiovascular
![]()
CiteULike
Connotea
Del.icio.us What's this?
This article has been cited by other articles:
![]() |
J. Mayer, J. Boldt, R. Beschmann, A. Stephan, and S. Suttner Uncalibrated arterial pressure waveform analysis for less-invasive cardiac output determination in obese patients undergoing cardiac surgery Br. J. Anaesth., August 1, 2009; 103(2): 185 - 190. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. R. Concha, V. F. Mertz, L. I. Cortinez, K. A. Gonzalez, and J. M. Butte Pulse Contour Analysis and Transesophageal Echocardiography: A Comparison of Measurements of Cardiac Output During Laparoscopic Colon Surgery Anesth. Analg., July 1, 2009; 109(1): 114 - 118. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Chatti, S. de Rudniki, S. Marque, A. S. Dumenil, A. Descorps-Declere, A. Cariou, J. Duranteau, M. Aout, E. Vicaut, and B. P. Cholley Comparison of two versions of the Vigileo-FloTracTM system (1.03 and 1.07) for stroke volume estimation: a multicentre, blinded comparison with oesophageal Doppler measurements Br. J. Anaesth., April 1, 2009; 102(4): 463 - 469. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. Compton and J.-H. Schafer Noninvasive Cardiac Output Determination: Broadening the Applicability of Hemodynamic Monitoring Seminars in Cardiothoracic and Vascular Anesthesia, March 1, 2009; 13(1): 44 - 55. [Abstract] [PDF] |
||||
![]() |
M. Cannesson, H. Musard, O. Desebbe, C. Boucau, R. Simon, R. Henaine, and J.-J. Lehot The Ability of Stroke Volume Variations Obtained with Vigileo/FloTrac System to Monitor Fluid Responsiveness in Mechanically Ventilated Patients Anesth. Analg., February 1, 2009; 108(2): 513 - 517. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Biancofiore, L. A. H. Critchley, A. Lee, L. Bindi, M. Bisa, M. Esposito, L. Meacci, R. Mozzo, P. DeSimone, L. Urbani, et al. Evaluation of an uncalibrated arterial pulse contour cardiac output monitoring system in cirrhotic patients undergoing liver surgery Br. J. Anaesth., January 1, 2009; 102(1): 47 - 54. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. W. L. Scheeren, C. Wiesenack, F. D. Compton, B. Zukunft, C. Hoffmann, W. Zidek, and J.-H. Schaefer Performance of a minimally invasive cardiac output monitoring system (Flotrac/Vigileo) Br. J. Anaesth., August 1, 2008; 101(2): 279 - 280. [Full Text] [PDF] |
||||
![]() |
B. Matthieu, N.-G. Karine, C. Vincent, V. Alain, C. J. Francois, R. Philippe, and S. Francois Cardiac Output Measurement in Patients Undergoing Liver Transplantation: Pulmonary Artery Catheter Versus Uncalibrated Arterial Pressure Waveform Analysis Anesth. Analg., May 1, 2008; 106(5): 1480 - 1486. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. D. Compton, B. Zukunft, C. Hoffmann, W. Zidek, and J.-H. Schaefer Performance of a minimally invasive uncalibrated cardiac output monitoring system (FlotracTM/VigileoTM) in haemodynamically unstable patients Br. J. Anaesth., April 1, 2008; 100(4): 451 - 456. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Davison and C. Junker Advances in Critical Care for the Nephrologist: Hemodynamic Monitoring and Volume Management Clin. J. Am. Soc. Nephrol., March 1, 2008; 3(2): 554 - 561. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Mayer, J. Boldt, M. W. Wolf, J. Lang, and S. Suttner Cardiac Output Derived from Arterial Pressure Waveform Analysis in Patients Undergoing Cardiac Surgery: Validity of a Second Generation Device Anesth. Analg., March 1, 2008; 106(3): 867 - 872. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Button, L. Weibel, O. Reuthebuch, M. Genoni, A. Zollinger, and C. K. Hofer Clinical evaluation of the FloTrac/VigileoTM system and two established continuous cardiac output monitoring devices in patients undergoing cardiac surgery Br. J. Anaesth., September 1, 2007; 99(3): 329 - 336. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. Guarracino, M. Stefani, F. Lapolla, C. Cariello, L. Doroni, A. Danella, R. Baldassarri, J. Mayer, J. Boldt, T. Schollhorn, et al. Monitoring cardiac output with Flo Trac VigileoTM Br. J. Anaesth., July 1, 2007; 99(1): 142 - 143. [Full Text] [PDF] |
||||
E-letters:
Read all E-letters
- Monitoring cardiac output with Flo Trac VigileoTM
- Fabio Guarracino, et al.
- British Journal of Anaesthesia, 13 Mar 2007 [Full text]
- Author reply: Monitoring cardiac output with Flo Trac VigileoTM
- Jochen Mayer, et al.
- British Journal of Anaesthesia, 30 Mar 2007 [Full text]
- Arterial pressure wave form and cardiac output in dilated cardiomyopathy
- Usha Kiran, et al.
- British Journal of Anaesthesia, 6 Jun 2007 [Full text]