* To whom correspondence should be addressed. Background. The USCOM ultrasonic cardiac output monitor (USCOM Pty Ltd, Coffs Harbour, NSW, Australia) is a non-invasive device that determines cardiac output by continuous-wave Doppler ultrasound. The aim of this study was to evaluate the accuracy of the USCOM device compared with the thermodilution technique in intensive care patients who had just undergone cardiac surgery. Methods. We conducted a prospective study in the 18-bed intensive care unit of a 600-bed tertiary referral hospital. Twenty-four mechanically ventilated patients were studied immediately following cardiac surgery. We evaluated the USCOM monitor by comparing its output with paired measurements obtained by the standard thermodilution technique using a pulmonary artery catheter. Results. Forty paired measurements were obtained in 22 patients. We were unable to obtain an acceptable signal in the remaining two patients. Comparison of the two techniques showed a bias of 0.18 and limits of agreement of -1.43 to 1.78. The agreement may not be as good between techniques at higher cardiac output values. Conclusions. The USCOM monitor has a place in intensive care monitoring. It is accurate, rapid, safe, well-tolerated, non-invasive and cost-effective. The learning curve for skill acquisition is very short. However, during the learning phase the USCOM monitor measurements are rather operator dependent. Its suitability for use in high and low cardiac output states requires further validation.
Accepted November 11, 2004
Clinical Investigation
Clinical evaluation of USCOM ultrasonic cardiac output monitor in cardiac surgical patients in intensive care unit
P. V. van Heerden, E-mail: peter.vanheerden{at}health.wa.gov.au
![]()
Abstract ![]()
CiteULike
Connotea
Del.icio.us What's this?
This article has been cited by other articles:
![]() |
W. H. L. Teoh and A. T. H. Sia Colloid Preload Versus Coload for Spinal Anesthesia for Cesarean Delivery: The Effects on Maternal Cardiac Output Anesth. Analg., May 1, 2009; 108(5): 1592 - 1598. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Phillips, P. Lichtenthal, J. Sloniger, D. Burstow, M. West, and J. Copeland Noninvasive Cardiac Output Measurement in Heart Failure Subjects on Circulatory Support Anesth. Analg., March 1, 2009; 108(3): 881 - 886. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. J. Funk, E. W. Moretti, and T. J. Gan Minimally Invasive Cardiac Output Monitoring in the Perioperative Setting Anesth. Analg., March 1, 2009; 108(3): 887 - 897. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. Siu, A. Tucker, S. K. Manikappa, and J. Monagle Does Patient Position Influence Doppler Signal Quality from the USCOM Ultrasonic Cardiac Output Monitor? Anesth. Analg., June 1, 2008; 106(6): 1798 - 1802. [Abstract] [Full Text] [PDF] |
||||
E-letters:
Read all E-letters
- Acceptable limits of agreement need to be defined a priori
- Anna Lee, et al.
- British Journal of Anaesthesia, 22 Mar 2005 [Full text]