Validation of pulse contour derived stroke volume variation during modifications of cardiac afterload


1 Department of Anesthesiology
2 Institute for Surgical Research, Ludwig Maximilians University Munich, Klinikum Großhadern, Marchioninistr. 15, D-81377 Munich, Germany
3 Department of Anesthesiology, University of Hamburg, Hamburg-Eppendorf University-Hospital, Martinistr. 52, 20246 Hamburg, Germany
* Corresponding author. E-mail: dreuter{at}uke.uni-hamburg.de
Background: Left ventricular stroke volume variation (SVV) or its surrogates are useful tools to assess fluid responsiveness in mechanically ventilated patients. So far it is unknown, how changes in cardiac afterload affect SVV. Therefore, this study compared left ventricular SVV derived by pulse contour analysis with SVV measured using an ultrasonic flow probe and investigated the influence of cardiac afterload on left ventricular SVV.
Methods: In 13 anaesthetized, mechanically ventilated pigs [31(SD 6) kg], we compared cardiac output (CO), stroke volume (SV), and SVV determined by pulse contour analysis and by an ultrasonic aortic flow signal (BlandAltman analysis). After obtaining baseline measurements, cardiac afterload was increased using phenylephrine and decreased using adenosine (both continuously administered). Measurements were performed with a constant tidal volume (12 ml kg1) without PEEP.
Results: Neither increasing mean arterial pressure (MAP) [from 59 (7) to 116 (19)] nor decreasing MAP [from 63 (7) to 39 (4)] affected CO, SV, and SVV (both methods). Method comparison revealed a bias for SVV of 0.1% [standard error of the mean (SE) 0.8] at baseline, 1.2% (SE 0.8) during decreased and 4.0% (SE 0.7) during increased afterload, the latter being significantly different from the others (P < 0.05). Thereby, pulse contour analysis tended to underestimate SVV during decreased afterload and to overestimate SVV during increased afterload. Limits of agreement were approximately 6% for all points of measurement.
Conclusions: Left ventricular SVV is not affected by changes in cardiac afterload. There is a good agreement of pulse contour with flow derived SVV. The agreement decreases, if afterload is extensively augmented.
Keywords: heart, cardiac afterload, positive end-expiratory pressure; heart, cardiac preload; heart, stroke volume variation, validation
Declaration of interest. A.E.G. and D.A.R. are members of the medical advisory board of Pulsion Medical Systems. This study was exclusively supported by a research grant of the University of Munich.
![]()
CiteULike
Connotea
Del.icio.us What's this?
This article has been cited by other articles:
![]() |
D. Lahner, B. Kabon, C. Marschalek, A. Chiari, G. Pestel, A. Kaider, E. Fleischmann, and H. Hetz Evaluation of stroke volume variation obtained by arterial pulse contour analysis to predict fluid responsiveness intraoperatively Br. J. Anaesth., September 1, 2009; 103(3): 346 - 351. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. J. Funk, E. W. Moretti, and T. J. Gan Minimally Invasive Cardiac Output Monitoring in the Perioperative Setting Anesth. Analg., March 1, 2009; 108(3): 887 - 897. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. L. Ryan, W. H. Cooke, C. A. Rickards, K. G. Lurie, and V. A. Convertino Breathing through an inspiratory threshold device improves stroke volume during central hypovolemia in humans J Appl Physiol, May 1, 2008; 104(5): 1402 - 1409. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. Bein, J. Renner, P. Meybohm, J. Scholz, J. C. Kubitz, and D. A. Reuter Validation of pulse contour derived stroke volume variation Br. J. Anaesth., November 1, 2007; 99(5): 745 - 746. [Full Text] [PDF] |
||||
E-letters:
Read all E-letters
- VALIDATION OF PULSE CONTOUR DERIVED STROKE VOLUME VARIATION
- Berthold Bein, et al.
- British Journal of Anaesthesia, 19 Jul 2007 [Full text]
- Response to the letter of Dr. Bein
- Daniel A. Reuter, et al.
- British Journal of Anaesthesia, 14 Aug 2007 [Full text]