BJA Advance Access originally published online on September 19, 2006
British Journal of Anaesthesia 2006 97(6):878-882; doi:10.1093/bja/ael252
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
An evaluation of poor LMA CTrachTM
views with a fibreoptic laryngoscope and the effectiveness of corrective measures,
1 Department of Anaesthesia, Yong Loo Lin School of Medicine, National University of Singapore Singapore
2 Department of Anaesthesia, National University Hospital Singapore
*Corresponding author: Department of Anaesthesia, National University Hospital, 5 Lower Kent Ridge Road, Singapore 119074. E-mail: analiue{at}nus.edu.sg
Background. The LMA CTrachTM
system has integrated fibreoptics to enable viewing of the glottis to aid tracheal intubation via a laryngeal mask conduit. Earlier evaluations of this system had shown high success rates of insertion, ventilation and tracheal intubation, but frequent difficulty or failure to view the glottis.
Method. We evaluated CTrach views in 100 patients. In 69 patients with partial or no views of the glottis, we evaluated the causes with the aid of an Olympus LF-DP fibreoptic laryngoscope passed through the CTrach. In those with poor views, we tested the effectiveness of simple corrective measures to improve the views and optimize CTrach placement. These included pushing the CTrach deeper in, partial withdrawal by 6 cm and reinsertion to correct epiglottic downfolding, withdrawal by 1 cm and applying forward lift to correct obstruction by the arytenoids, and complete removal to clean the fibreoptic ports followed by reinsertion when secretions were the problem.
Results. Downfolding of the epiglottis caused poor views in 57 patients, which was correctable in 51. Obstruction of the view by the arytenoids occurred in seven patients and by secretions in five, all of which were correctable. With the Ctrach, we were able to ventilate the lungs in all patients, view the glottis in 94 and intubate the trachea in 97 patients.
Conclusion. Despite the ability to ventilate the lungs in all patients with the CTrach, poor initial views were very frequent, indicating suboptimal placement, although most poor views can be improved, leading to a high intubation success rate.
Declaration of interest. The LMA Company Singapore supplied the LMA CTrachTM equipment for this study and supported the costs of fibreoptic laryngoscopy and sterilization of equipment. The Department of Anaesthesia, National University Hospital, Singapore has been paid by the LMA Company Singapore for teaching on other laryngeal mask devices in Singapore, China and Malaysia. The authors have evaluated other airway equipment. The Department of Anaesthesia, National University Hospital, Singapore and the authors have not accepted any payment for their evaluations of the LMA CtrachTM or any other airway equipment. None of the authors has any personal financial relationship with or stock holdings in any company whose product is mentioned in this report or with any competing company. Part of this work was presented at the European Society of Anaesthesiology Annual Meeting in June 2006 in Madrid, Spain.
LMA CTrachTM is the property of LMA International N.V.
![]()
CiteULike
Connotea
Del.icio.us What's this?
This article has been cited by other articles:
![]() |
F. Martin and D. J. Buggy New airway equipment: opportunities for enhanced safety Br. J. Anaesth., June 1, 2009; 102(6): 734 - 738. [Full Text] [PDF] |
||||
![]() |
A. Ojeda, A. M. Lopez, X. Borrat, and R. Valero Failed Tracheal Intubation with the LMA-CTrachTM in Two Patients with Lingual Tonsil Hyperplasia Anesth. Analg., August 1, 2008; 107(2): 601 - 602. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. H. C. Liu A LMA CTrachTM for large patients Br. J. Anaesth., August 1, 2008; 101(2): 284 - 285. [Full Text] [PDF] |
||||
![]() |
K. B. Greenland FastrachTM tubes: modifying the design for use with the LMA CTrachTM? Br. J. Anaesth., July 1, 2007; 99(1): 146 - 147. [Full Text] [PDF] |
||||
![]() |
D. Cattano, B. Pesetti, C. Di Salvo, F. Giunta, E. H. C. Liu, R. W. L. Goy, and F. G. Chen Evaluation of the LMA CTrach Br. J. Anaesth., March 1, 2007; 98(3): 409 - 409. [Full Text] [PDF] |
||||
E-letters:
Read all E-letters
- cTrach: the size 6 could be better
- Davide Cattano, et al.
- British Journal of Anaesthesia, 7 Dec 2006 [Full text]
- Response to "CTrach: the size 6 could be better"
- Eugene H.C. Liu
- British Journal of Anaesthesia, 15 Dec 2006 [Full text]
- Fastrach tubesTM – modifying the design for use with the LMA CTrachTM?
- Keith B. Greenland
- British Journal of Anaesthesia, 1 May 2007 [Full text]