British Journal of Anaesthesia, 2003, Vol. 91, No. 5 678-683
© 2003 The Board of Management and Trustees of the British Journal of Anaesthesia
Clinical Investigations |
Randomized crossover comparison of the ProSealTM laryngeal mask airway with the Laryngeal Tube® during anaesthesia with controlled ventilation
Royal United Hospital, Combe Park, Bath, UK
Corresponding author
Background. The Laryngeal Tube (LT®) performs similarly to the classic laryngeal mask airway during controlled ventilation but with an improved airway seal. We compared the laryngeal tube with the ProSealTM laryngeal mask airway (PLMA) throughout anaesthesia.
Methods. Thirty-two patients were studied using a randomized cross-over design. The primary outcome measure was airway seal pressure. Secondary outcome measures included peak and plateau airway pressures, time to achieve an airway, ease of insertion, airway manipulations required to achieve a patent airway and grade of fibre-optic laryngoscopy. The proportion of patients in whom good, fair or failed ventilation was achieved was also calculated.
Results. No significant difference was found in regard to seal pressure (PLMA, median 26.5 cm H2O, range 1040; LT, median 24, range 640; P=0.7, 95% confidence interval of the difference 3.5 to 4.0). There were two failures of insertion or ventilation in the LT group and none in the PLMA group. The peak airway pressure with the PLMA was lower than with the LT but the difference was clinically unimportant (PLMA, mean 16.2 cm H2O, SD 3.52; LT, mean 17.9, SD 5.21; P=0.02, 95% confidence interval of the difference 3.1 to 0.28). The PLMA took significantly less time to insert than the LT (PLMA, median 18.5 s, interquartile range 1426; LT, median 22, interquartile range 1536.5; P<0.02, 95% confidence interval of the difference 21.5 to 1.0). The PLMA gave a significantly better view on fibre-optic laryngoscopy than the LT (P<0.001, 95% confidence interval of the difference in grade 2.0 to 1.0). In the 16 patients in whom the PLMA was used during maintenance of anaesthesia ventilation was good in 15, fair in none and failed in one. The equivalent figures for the LT were good in nine, fair in six and failed in one (P=0.009). There was no significant difference in the plateau airway pressure, ease of insertion of the devices, number of manipulations required to achieve or maintain an airway, or in overall complications.
Conclusion. The two devices performed equally well in terms of seal pressure. The PLMA was quicker to insert. Efficacy of ventilation was significantly better with the PLMA than the LT. The PLMA allowed a significantly better view of the larynx with a fibre-optic laryngoscope, and may therefore be of more use in cases where visualization of the larynx is required.
Br J Anaesth 2003; 91: 67883
![]()
CiteULike
Connotea
Del.icio.us What's this?
This article has been cited by other articles:
![]() |
J. T. Kim, S. Y. Jeon, C. S. Kim, S. D. Kim, and H. S. Kim Alternative method for predicting optimal insertion depth of the laryngeal tube in children Br. J. Anaesth., November 1, 2007; 99(5): 704 - 707. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Mihai, G. Knottenbelt, and T. M. Cook Evaluation of the revised laryngeal tube suction: the laryngeal tube suction II in 100 patients Br. J. Anaesth., November 1, 2007; 99(5): 734 - 739. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Asai and K. Shingu The laryngeal tube Br. J. Anaesth., December 1, 2005; 95(6): 729 - 736. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. M. Cook and J. Cranshaw Randomized crossover comparison of ProSeal(R) Laryngeal Mask Airway with Laryngeal Tube Sonda(R) during anaesthesia with controlled ventilation Br. J. Anaesth., August 1, 2005; 95(2): 261 - 266. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. M. Cook Use of the CobraPLATM for Percutaneous Cricothyroidotomy and Tracheostomy Anesth. Analg., March 1, 2005; 100(3): 902 - 902. [Full Text] [PDF] |
||||
![]() |
D. M. Miller A Proposed Classification and Scoring System for Supraglottic Sealing Airways: A Brief Review Anesth. Analg., November 1, 2004; 99(5): 1553 - 1559. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. Agro, G. Barzoi, and B. Galli The CobraPLATM in 110 anaesthetized and paralysed patients: what size to choose? Br. J. Anaesth., May 1, 2004; 92(5): 777 - 778. [Full Text] [PDF] |
||||