Background. The bispectral index (BIS) and the rapidly extracted auditory evoked potentials index (A-line ARX Index or AAI) have been proposed as methods to measure the depth of sedation. A prospective study was designed to assess the performance of both these methods for measuring the depth of sedation induced by propofol or midazolam under epidural block.
Methods. Thirty-two ASA I and II adult patients undergoing elective gynaecological surgery under low-thoracolumbar epidural block were studied. Eighteen patients received propofol (Group P: 20 mg bolus every 3 min) and 14 received midazolam (Group M: 0.5 mg bolus every 5 min) until an observers assessment of alertness/sedation (OAA/S) scale score of 1 was achieved. AAI and BIS were monitored for different OAA/S scores.
Results. AAI and BIS decreased and increased following the changes on the patients OAA/S scores and correlated with sedation significantly. During the onset phase, the coefficients of Spearmans rank correlation for AAI and BIS were respectively 0.958 and 0.898 (P<0.001) for Group P, and 0.973 and 0.945 (P<0.001) for Group M. During the recovery phase in Group P, the coefficients were respectively 0.946 and 0.702 (P<0.001). Linear regression analysis showed that both AAI and BIS were linearly related to the OAA/S scores. The coefficients of Spearmans rank correlation and linear regression for AAI were all greater than those for BIS (P<0.05).
Conclusions. Both AAI and BIS correlated well with the depth of sedation induced by propofol or midazolam under epidural block. AAI may be more valuable when monitoring depth of sedation.
Br J Anaesth 2002; 89: 2604
![]()
CiteULike
Connotea
Del.icio.us What's this?
This article has been cited by other articles:
![]() |
D. Hohener, S. Blumenthal, and A. Borgeat Sedation and regional anaesthesia in the adult patient Br. J. Anaesth., January 1, 2008; 100(1): 8 - 16. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. Weber, M. Zimmermann, and T. Bein The Impact of Acoustic Stimulation on the AEP Monitor/2 Derived Composite Auditory Evoked Potential Index Under Awake and Anesthetized Conditions Anesth. Analg., August 1, 2005; 101(2): 435 - 439. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Kakinohana, S. Nakamura, Y. Miyata, and K. Sugahara Emergence from Propofol Anesthesia in a Nonagenarian at a Bispectral Index of 52 Anesth. Analg., July 1, 2005; 101(1): 169 - 170. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Ekman, L. Brudin, and R. Sandin A Comparison of Bispectral Index and Rapidly Extracted Auditory Evoked Potentials Index Responses to Noxious Stimulation During Sevoflurane Anesthesia Anesth. Analg., October 1, 2004; 99(4): 1141 - 1146. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Nishiyama, T. Matsukawa, and K. Hanaoka A Comparison of the Clinical Usefulness of Three Different Electroencephalogram Monitors: Bispectral Index, Processed Electroencephalogram, and Alaris Auditory Evoked Potentials Anesth. Analg., May 1, 2004; 98(5): 1341 - 1345. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. N. Schmidt, P. Bischoff, T. Standl, A. Gerhardt, G. Lankenau, and J. S. am Esch Alaris AEPTM Monitor's "Click Detection" Does Not Help to Detect Inadvertent Disconnection of Headphones During Anesthesia Anesth. Analg., January 1, 2004; 98(1): 123 - 127. [Abstract] [Full Text] [PDF] |
||||