British Journal of Anaesthesia, 1991, Vol. 67, No. 1 36-40
© 1991 The Board of Management and Trustees of the British Journal of Anaesthesia
research-article |
COMPARISON OF THE FINAPRES AND DIRECT ARTERIAL PRESSURE MONITORING DURING PROFOUND HYPOTENSIVE ANAESTHESIA
Department of Anaesthesia, Western Infirmary Dumbarton Road, Glasgow G11 6NT
University Department of Anaesthesia, Glasgow Royal Infirmary Alexandra Parade, Glasgow G31 2ER
The Finapres was compared with direct intraarterial pressure monitoring in 10 patients undergoing local resection of choroidal melanoma, an operation that requires a period of profound hypotension. Good agreement was recorded for systolic arterial pressure and heart rate over a range of pressures. However, agreement of mean and diastolic pressures was poor, with the Finapres tending to overestimate these values. In cases requiring profound hypotension, direct arterial pressure monitoring remains the method of choice.
![]()
CiteULike
Connotea
Del.icio.us What's this?
This article has been cited by other articles:
![]() |
B. P.M. Imholz, W. Wieling, G. A. van Montfrans, and K. H. Wesseling Fifteen years experience with finger arterial pressure monitoring:: assessment of the technology Cardiovasc Res, June 1, 1998; 38(3): 605 - 616. [Abstract] [Full Text] [PDF] |
||||