BJA Advance Access published online on June 20, 2008

British Journal of Anaesthesia, doi:10.1093/bja/aen181
This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Supplementary Data
Right arrow All Versions of this Article:
101/3/358    most recent
aen181v1
Right arrow E-Letters: Submit a response to the article
Right arrow Alert me when this article is cited
Right arrow Alert me when E-letters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrowRequest Permissions
Right arrow Disclaimer
Google Scholar
Right arrow Articles by McCahon, R. A.
Right arrow Articles by Hardman, J. G.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by McCahon, R. A.
Right arrow Articles by Hardman, J. G.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?

© The Board of Management and Trustees of the British Journal of Anaesthesia 2008. All rights reserved. For Permissions, please e-mail: journals.permissions@oxfordjournals.org

Validation and application of a high-fidelity, computational model of acute respiratory distress syndrome to the examination of the indices of oxygenation at constant lung-state

R. A. McCahon1, M. O. Columb2, R. P. Mahajan1 and J. G. Hardman1,*

1 University Department of Anaesthesia, Queen's Medical Centre, Nottingham NG7 2UH, UK
2 Intensive Care Unit, Wythenshawe Hospital, Manchester M23 9LT, UK

* Corresponding author. E-mail: j.hardman{at}nottingham.ac.uk

Accepted for publication May 2, 2008.


    Abstract
 Top
 Abstract
 Introduction
 Methods
 Results
 Discussion
 Supplementary material
 Funding
 References
 
Background: Calculated venous admixture (Qs/Qt) is considered the best index of oxygenation; surrogates have been developed (PaO2/FIO2, respiratory index, and arterioalveolar PO2 difference), but these vary with FIO2, falsely indicating a change in lung-state. Using a novel model, we aimed to quantify the behaviour of the indices of oxygenation listed above during physiological and treatment factor variation. The study is the first step in developing an accurate and non-invasive tool to quantify oxygenation defects.

Methods: We present the static and dynamic validation of a novel computational model of gas exchange in acute respiratory distress syndrome (ARDS) based upon the Nottingham Physiology Simulator. Arterial gas tension predictions were compared with data derived from ARDS patients. The subsequent study examined the indices' susceptibility to variation induced by independent changes in FIO2 (0.3–1.0), haemoglobin concentration (Hb: 6–14 g dl–1), oxygen consumption (VO2: 250–350 ml min–1), and PaCO2 (4–8 kPa).

Results: Static validation produced a mean error of –0.3%, a 10-fold improvement over previous models. Dynamic validation produced a mean prediction error of –0.05 kPa for PaO2 and 0.09 kPa for PaCO2. Every parameter, especially FIO2, induced variation in all indices. The least FIO2-dependent index was Qs/Qt (variation: 5.1%). In contrast, PaO2/FIO2 varied by 77% through the range of FIO2.

Conclusions: We have improved simulation of gas exchange in ARDS by using a sophisticated respiratory model. Using the validated model, we have demonstrated that the current indices of oxygenation vary with alteration in Hb, PaCO2, and VO2 in addition to their previously well-documented dependence on FIO2.

Keywords: lung, respiratory distress syndrome; measurement techniques, gas exchange; measurement techniques, lung shunting; model, computer simulation; model, respiratory failure


    Introduction
 Top
 Abstract
 Introduction
 Methods
 Results
 Discussion
 Supplementary material
 Funding
 References
 
On a daily basis, clinicians use patients' arterial gas tensions to communicate and interpret the severity of gas exchange defects; however, these values are dependent upon factors outside the patient's lungs, such as inspired oxygen fraction (FIO2), minute ventilation, and oxygen consumption, meaning that arterial gas tensions must be viewed in the context of other physiological values and the treatment provided. For many years, clinicians have sought to describe patients' oxygenation without reliance on such contextual descriptors.15

Unfortunately, the complexity of the human lung's ventilation–perfusion (VQ) and the behaviour of whole blood, all integrated with a dynamic and autoregulating cardiovascular system, cannot easily be described. Thus, the tension-based indices of oxygenation (i.e. PaO2/FIO2, respiratory index, and arterioalveolar PO2 difference) are not as reliable as the gold standard, content-based venous admixture (Qs/Qt).412 In particular, the tension-based indices vary as factors outside the patient's pathophysiology change, falsely implying a change in the patient's lung-state. This has far-reaching implications as the indices of oxygenation are used for communicating and monitoring a patient's illness severity, assessing the efficacy of interventions, stratification for research purposes, resource allocation, and determination of clinical management pathways. However, the continued popularity of the tension-based indices must be due, in part, to the fact that Qs/Qt calculation requires mixed venous blood sampling via a pulmonary artery catheter, which is not without risk.13 Therefore, there is a need for an accurate, clinically useful, and non-invasive tool to quantify patients' lung-states.

The use of simulated acute respiratory distress syndrome (ARDS) lung provides opportunities for research unavailable using standard clinical approaches. In 2001, Hahn14 called for a detailed investigation of this issue using sophisticated modelling; in particular, hypoxic pulmonary vasoconstriction (HPV) and tidal ventilation were regarded as essential but, at the time, missing. The Nottingham Physiology Simulator (NPS) is a sophisticated computational multi-compartmental lung model that incorporates tidal ventilation, pulsatile pulmonary blood flow, HPV, and a realistic and validated oxygen–haemoglobin model.1517

Our intention was to validate and apply an NPS ARDS lung model to an assessment of the behaviour of the current indices of oxygenation during variation in diverse physiological factors within the model; these included haemoglobin concentration, oxygen consumption, inspired oxygen concentration, and arterial CO2 tension. The study represents the first step in a programme of work intended to produce a robust index of oxygenation whose calculation does not require mixed-venous blood sampling.


    Methods
 Top
 Abstract
 Introduction
 Methods
 Results
 Discussion
 Supplementary material
 Funding
 References
 
The Nottingham Physiology Simulator
The NPS has been validated in a number of situations and configurations.1520 The principles underlying the modelling are summarized in the appendix (see Supplementary material at British Journal of Anaesthesia online).

NPS version 060406 was used in this investigation and it is available for download via the corresponding author.

Model calibration
A detailed description of the model configuration and validation exercise is available in the appendix (see Supplementary material at British Journal of Anaesthesia online).

The ARDS VQ defect was calibrated in the NPS using data published by Nirmalan and colleagues7 in 2001; in this investigation, the authors measured arterial and mixed venous gas tensions, haemoglobin concentration, and cardiac output in 10 patients known to have ARDS. The NPS ARDS model VQ defect was configured using the 10 data sets from Patients 3 and 4. These patients were randomly selected and subsequently excluded from the validation investigation. The data set from two patients was considered adequate to configure the model VQ defect; this allowed validation against the remaining eight patients' data.

Model validation
In brief, a static validation of the configured NPS model was performed using the remaining 45 data sets (Patients 1, 2, 5, 6, 7, 8, 9, and 10) from Nirmalan's study. Within-subject and whole-group PaO2 prediction errors were calculated.

Subsequently, a dynamic validation of the configured NPS model was performed using published data on ARDS patient responses to changing FIO2, respiratory rate (RR), and tidal volume (Tv).15 For the purposes of validation, the quality of matching was judged by the error in predicting the resulting PaO2 and PaCO2.

Evaluation of indices of oxygenation
The values of tidal volume, RR, PEEP, and inspiratory to expiratory ratio were chosen to represent typical ARDS patients.7 13 21 HPV and dynamic oxygen–haemoglobin association–dissociation were enabled within the model.

The following indices of oxygenation were evaluated in this investigation: calculated shunt fraction (Qs/Qt), alveolar–arterial oxygen tension gradient (PA–aO2), respiratory index (PA–aO2/PaO2), and PF ratio (PaO2/FIO2).

Each index was recorded in a virtual patient with the configured VQ defect held constant (Table 1). The inspired oxygen concentration (FIO2), haemoglobin concentration (Hb), arterial carbon dioxide tension (PaCO2), and oxygen consumption (VO2) were varied in isolation, whereas the other variables were clamped (held constant) at baseline (Table 2). Alveolar oxygen tension (PAO2) was calculated using the alveolar gas equation and arterial oxygen tension (PaO2) and mixed venous oxygen tension (PvO2) were recorded.


View this table:
[in this window]
[in a new window]

 
Table 1 NPS configuration for the virtual patient used in this investigation. RER varied between 0.47 and 1.03 to maintain PaCO2 during variation in VO2 or alteration of PaCO2 independent of other variables

 


View this table:
[in this window]
[in a new window]

 
Table 2 Variables adjusted during examination of the indices of oxygenation. Three of the four variables were held at baseline whereas the remaining variable was adjusted through the examined range using the increment

 
Tidal volume (Vt) and RR were constant throughout the investigation (Table 1), thus keeping lung-state constant. Respiratory exchange ratio (RER) was kept constant except to induce change in (or maintain) PaCO2 while VO2 was altered.

Statistical analysis
Comparison was made between the NPS model and the simpler pulmonary model of Nirmalan and colleagues. Model performances were assessed by comparing the predicted estimates with the measured PaO2 (kPa) values using within-subject Pearson correlation (r) and agreement was quantified using intraclass correlation (ri). Biases (predicted–measured) for actual and proportional errors in PaO2 data were estimated with 95% confidence intervals (CI) and within-subject 95% limits of agreement (LA). Error biases (proportional differences or prediction errors) were calculated as: (predicted value–measured value)/measured value. In addition, the data were subjected to regression analysis to estimate the within-subject 95% prediction intervals across the range of PaO2 values used in the validations. Analyses were carried out using Excel 2000 (Microsoft Inc., Redmond, WA, USA), Number Cruncher Statistical Software (NCSS 2004, Inc., Kaysville, UT, USA), and GraphPad Prism 4.03 (GraphPad Software Inc., San Diego, CA, USA).


    Results
 Top
 Abstract
 Introduction
 Methods
 Results
 Discussion
 Supplementary material
 Funding
 References
 
Model calibration
The NPS ARDS VQ defect was configured by setting bronchial resistance vertical slider-1 to 400/2500 and vascular resistance vertical slider-5 to 16/2500. Sliders 1–5 on each scale represent the anatomical progression from lung apex to base, respectively. The resulting configuration of the NPS ARDS lung model incorporated a substantial VQ defect and an absolute shunt of 40% of cardiac output. All other sliders stayed at their default settings of 1. This resulted in an average PaO2 prediction error against the calibration data set of 1.48% (SD 14.0%).

Static validation
Within-subject correlation coefficients (r) for measured and predicted PaO2 were 0.897 for the Nirmalan model and 0.903 for the NPS model. Overall agreements, as assessed by intraclass correlation coefficients (ri) were 0.968 and 0.975, respectively, showing further improvements in model fit with the NPS model. The predicted–measured biases with 95% CI and within-subject 95% LA were –0.58 kPa (95% CI –1.57 to 0.41, 95% LA –2.89 to 1.73) and –0.20 kPa (95% CI –1.13 to 0.73, 95% LA –2.38 to 1.98) for the models, respectively. These represent error biases of –3.0% (95% CI –9.3 to 3.39; 95% LA –17.7 to 11.7) and –0.3% (95% CI –5.7 to 5.1; 95% LA –12.8 to 12.2), respectively, in favour of the NPS model. The within-subject SD of residuals of 1.11 and 1.07 kPa (representing the corresponding 95% prediction intervals of 2.18 and 2.10 kPa) for the models, respectively, show that differences in predicted and measured estimates are not clinically important.

Dynamic validation
The settings for extrapulmonary shunt and physiological dead space used to match the NPS ARDS model to patients before intervention are given in the appendix (Supplementary Table 3a). The post-intervention PaO2 and PaCO2 predicted by the NPS ARDS model are given in the appendix (Supplementary Table 2a). The mean magnitude of change observed in patients' PaO2 after intervention was 6.05 kPa; the 95% CI of the error in predicting PaO2 was –0.53 to 0.43 kPa (error bias –0.05 kPa). The mean magnitude of change observed in patients' PaCO2 after intervention was 0.94 kPa; the 95% CI of the error in predicting PaCO2 was –0.10 to 0.27 kPa (error bias 0.09 kPa).

Evaluation of indices of oxygenation
Variations in PAO2, PaO2, and PvO2 induced by changing Hb, FIO2, PaCO2, and VO2 are presented in Figure 1.


Figure 1
View larger version (13K):
[in this window]
[in a new window]
[Download PowerPoint slide]
 
Fig 1 Induced variation in alveolar, arterial, and mixed-venous oxygen tensions.

 
Variations in PaO2/FIO2, Qs/Qt, PA–aO2, and PA–aO2/PaO2 induced by changing Hb, FIO2, PaCO2, and VO2 are presented in Figures 25, respectively. Maximal variation in each index is presented in Table 3.


Figure 2
View larger version (8K):
[in this window]
[in a new window]
[Download PowerPoint slide]
 
Fig 2 Induced variation in the PF ratio (PaO2/FIO2) with constant pulmonary VQ configuration.

 


Figure 3
View larger version (9K):
[in this window]
[in a new window]
[Download PowerPoint slide]
 
Fig 3 Induced variation in the calculated venous admixture with constant pulmonary VQ configuration.

 


Figure 4
View larger version (9K):
[in this window]
[in a new window]
[Download PowerPoint slide]
 
Fig 4 Induced variation in the alveolar–arterial oxygen tension gradient with constant pulmonary VQ configuration.

 


Figure 5
View larger version (8K):
[in this window]
[in a new window]
[Download PowerPoint slide]
 
Fig 5 Induced variation in the respiratory index (PA–aO2/PaO2) with constant pulmonary VQ configuration.

 


View this table:
[in this window]
[in a new window]

 
Table 3 Maximal induced variation, expressed as a percentage of the smallest observed value

 
Variation in FIO2 caused the largest variation in each index; Qs/Qt had the smallest FIO2-induced variability of 5.1%, whereas PaO2/FIO2 varied by 77%. PaO2/FIO2 was also susceptible to variation induced by changing VO2, Hb, and PaCO2.

Induced changes were linear except in PaO2/FIO2 and PA–aO2/PaO2 during FIO2 variation. Each of these indices was less prone to FIO2-induced variation between FIO2 0.6 and 1.0; in this range, PaO2/FIO2 varied by 12.3% and PA–aO2/PaO2 varied by 16.3%.


    Discussion
 Top
 Abstract
 Introduction
 Methods
 Results
 Discussion
 Supplementary material
 Funding
 References
 
The two validation exercises generated a credible, high-fidelity model that can be applied reliably to the clinical context of ARDS throughout the relevant ranges of treatment and pathophysiological variation. The increased sophistication of the NPS model most likely explains its improved performance over that of Nirmalan and colleagues.7 The addition of HPV, tidal ventilation, pulsatile pulmonary blood flow, and dynamic oxygen–haemoglobin binding may reasonably be expected to make the model more credible and more life-like than earlier, simple models.

Using the validated NPS ARDS model, the current indices of oxygenation showed significant variability without a change in lung-state (Table 3), generating false impressions of the patient's disease. Overall, the content-based Qs/Qt varied the least, with PaO2/FIO2 performing the best among the less reliable tension-based indices. In particular, PaO2/FIO2 varied little within the ARDS-relevant range of FIO2 60–100%, which is in agreement with previous studies.6 7 22

Previous modelling studies have indicated that Hb does not influence gas exchange in ARDS.6 However, in those investigations, PaO2 and PvO2 were held unrealistically constant. In this study, a reduction in Hb decreased PvO2 in the shunt-rich NPS ARDS model. It is apparent that for a given PAO2, reduced PvO2 may induce diffusion-limitation of oxygen transfer, thereby reducing PaO2.23 24 Therefore, haemoglobin concentration has a predictable effect on the calculated tension-based indices (Table 3) via this mechanism. In the same way, alterations in VO2 caused changes in PvO2 and thus PaO2. Although there was small variation in PAO2, the gradient of the line (Fig. 1) was almost identical to that of PaO2, generating little variation in PA–aO2.

Indices that incorporate PAO2 assume the equality of PaCO2 and PACO2, that is, PaCO2 is easily measurable and acts as a surrogate for PACO2 in the alveolar gas equation. This assumption may not be appropriate within the ARDS lung. Indeed, unique values for PAO2 and PACO2 do not exist; such averaged values inadequately represent areas of the lung that are poorly ventilated, perfused, or both. The variation of PaO2 with PaCO2 is explained by the NPS ARDS model's dynamic oxygen–haemoglobin dissociation model;15 both Haldane and Bohr effects may be observed within the NPS. The incorporation into our ARDS model of a high-fidelity model of the dynamic relationship between oxygen and haemoglobin is crucially important, and represents an advance over previous investigations in this field.

In this investigation, we used a shunt-rich model of ARDS lung-state. However, in disease states with less absolute shunt (where perfused alveoli receive zero ventilation) and more venous admixture (where perfused alveoli receive inadequate but non-zero ventilation), both content- and tension-based indices are likely to behave quite differently from that observed in this study. Previous clinical and modelling investigations have shown a variable effect of FIO2 on Qs/Qt.2428 Plausible explanations for the apparent contradictions include blunting of HPV, absorption atelectasis, and the diversity of VQ distributions. Furthermore, previous studies have found that in the presence of smaller shunts and lower FIO2, PaO2/FIO2 was dependent on FIO2, thus rendering it misleading in clinical practice.6 22 This could be explained by the different responsiveness of PaO2 to changing FIO2 in the presence of poorly ventilated compared with unventilated alveoli.

Despite the improvements over previous models of respiratory pathophysiology, the current study is limited in that the results are representative of a single virtual patient rather than a heterogeneous patient population. While our virtual patient subject is broadly representative of the ARDS population, compliant, impervious to harm, and free of ethical constraint, it is a single subject, and does not represent every lung-state; future modelling studies should aim to create virtual populations to allow the broader applicability of findings and a greater demonstration of model robustness.

In conclusion, we have improved simulation of gas exchange in ARDS by using a sophisticated non-linear respiratory model based upon the NPS. The increased sophistication of the NPS ARDS model has permitted physiological clamping, that is, the ability to hold a physiological parameter constant. Thus, we have shown that the indices of oxygenation also vary with alteration in Hb, PaCO2, and VO2 in addition to their well-documented dependence on FIO2; such dependence causes apparent changes in lung-state during alterations in physiological and treatment factors when no change in lung-state has occurred. Future work should examine the robustness of these indices in other pulmonary disease states and in the population variation seen within disease states. The factors that vary with external variation should now be isolated and mathematically ‘clamped’ so that we may develop new, independent indices of gas exchange and, in particular, oxygenation.


    Supplementary material
 Top
 Abstract
 Introduction
 Methods
 Results
 Discussion
 Supplementary material
 Funding
 References
 
Supplementary material is available at British Journal of Anaesthesia online.


    Funding
 Top
 Abstract
 Introduction
 Methods
 Results
 Discussion
 Supplementary material
 Funding
 References
 
This work was supported in part by a grant from the European Society of Anesthesiology.


    References
 Top
 Abstract
 Introduction
 Methods
 Results
 Discussion
 Supplementary material
 Funding
 References
 
1 Filley GF, Bigelow DB, Olson DE, Lacquet LM. Pulmonary gas transport. A mathematical model of the lung. Am Rev Respir Dis (1968) 98:480–9.[Web of Science][Medline]

2 Riley RL. Analysis of factors affecting partial pressures of O2 and CO2 in gas and blood of lungs. Methods. J Appl Physiol (1951) 4:102–20.[Free Full Text]

3 West JB. Effects of ventilation–perfusion inequality on overall gas exchange studied in computer models of the lung. J Appl Physiol (1969) 202:116.

4 Cane RD, Shapiro BA, Templin R, Walther K. Unreliability of oxygen tension-based indices in reflecting intrapulmonary shunting in critically ill patients. Crit Care Med (1988) 16:1243–5.[Web of Science][Medline]

5 Fletcher R. Relationship between alveolar deadspace and arterial oxygenation in children with congenital cardiac disease. Br J Anaesth (1989) 62:168–76.[Abstract/Free Full Text]

6 Gowda MS, Klocke RA. Variability of indices of hypoxemia in adult respiratory distress syndrome. Crit Care Med (1997) 25:41–5.[CrossRef][Web of Science][Medline]

7 Nirmalan M, Willard T, Columb MO, Nightingale P. Effects of changes in arterial mixed venous oxygen content difference (C(a-v)O2) on indices of pulmonary oxygen transfer in a model ARDS lung. Br J Anaesth (2001) 86:477–85.[Abstract/Free Full Text]

8 Gilbert R. Arterial–alveolar oxygen partial pressure ratio. Crit Care Med (1987) 15:540–1.[Web of Science][Medline]

9 Gilbert R, Auchincloss JH Jr, Kuppinger M, Thomas MV. Stability of the arterial/alveolar oxygen partial pressure ratio. Effects of low ventilation/perfusion regions. Crit Care Med (1979) 7:267–72.[Web of Science][Medline]

10 Sganga G, Siegel JH, Coleman B, Giovannini I, Boldrini G, Pittiruti M. The physiologic meaning of the respiratory index in various types of critical illness. Circ Shock (1985) 17:179–93.[Web of Science][Medline]

11 Zetterstrom H. Assessment of the efficiency of pulmonary oxygenation. The choice of oxygenation index. Acta Anaesthesiol Scand (1998) 32:579–84.

12 Rasanen J, Downs JB, Malec DJ, Oates K. Oxygen tensions and oxyhaemoglobin saturations in the assessment of pulmonary gas exchange. Crit Care Med (1987) 15:1058–61.[Web of Science][Medline]

13 Wheeler AP, Bernard GR, Thompson BT, et al. National Heart, Lung, and Blood Institute Acute Respiratory Distress Syndrome (ARDS) Clinical Trials Network. Pulmonary-artery versus central venous catheter to guide treatment of acute lung injury. N Engl J Med (2006) 354:2213–24.[Abstract/Free Full Text]

14 Hahn CEW. Editorial 1; KISS and indices of pulmonary oxygen transfer. Br J Anaesth (2001) 86:465–7.[Free Full Text]

15 Hardman JG, Bedforth NM, Ahmed AB, Mahajan RP, Aitkenhead AR. A physiology simulator: validation of its respiratory components and its ability to predict the patient's response to changes in mechanical ventilation. Br J Anaesth (1998) 81:327–32.[Abstract/Free Full Text]

16 Hardman JG, Wills JS, Aitkenhead AR. Investigating hypoxemia during apnea: validation of a set of physiological models. Anesth Analg (2000) 90:614–8.[Abstract/Free Full Text]

17 Wills JS, Hardman JG. The development of hypoxemia during apnoea in children: a computational modelling investigation. Br J Anaesth (2006) 97:564–70.[Abstract/Free Full Text]

18 Hardman JG, Aitkenhead AR. Validation of an original mathematical model of CO2 elimination and dead space ventilation. Anesth Analg (2003) 97:1840–5.[Abstract/Free Full Text]

19 Hardman JG, Wills JS, Aitkenhead AR. Factors determining the onset and course of hypoxaemia during apnoea; an investigation using physiological modelling. Anesth Analg (2000) 90:619–24.[Abstract/Free Full Text]

20 McNamara MJ, Hardman JG. Hypoxaemia during open-airway apnoea; a computational modelling analysis. Anaesthesia (2005) 60:741–6.[CrossRef][Web of Science][Medline]

21 Brower RG, Matthay MA, Morris A, Schoenfeld D, Thompson BT, Wheeler A. Ventilation with lower tidal volumes as compared with traditional tidal volumes for acute lung injury and the acute respiratory distress syndrome. N Engl J Med (2000) 342:1301–8.[Abstract/Free Full Text]

22 Ingram K, Hardman JG. The variability of indices of oxygenation. Eur J Anaesthesiol (2005) 22:A232.

23 Wagner PD. Influence of mixed venous PO2 on diffusion of O2 across the pulmonary blood:gas barrier. Clin Physiol (1982) 2:105–15.[Web of Science][Medline]

24 Rossaint R, Hahn SM, Pappert D, Falke KJ, Radermacher P. Influence of mixed venous PO2 and inspired O2 fraction on intrapulmonary shunt in patients with severe ARDS. J Appl Physiol (1995) 78:1531–6.[Abstract/Free Full Text]

25 Douglas ME, Downs JB, Dannemiller FJ, Hodges MR, Munson ES. Change in pulmonary venous admixture with varying inspired oxygen. Anesth Analg (1976) 55:688–95.[Abstract/Free Full Text]

26 Quan SF, Kronberg GM, Schlobohm RM, Feeley TW, Don HF, Lister G. Changes in venous admixture with alterations of inspired oxygen concentration. Anesthesiology (1980) 52:477–82.[Web of Science][Medline]

27 Santos C, Ferrer M, Roca J, Torres A, Hernandez C, Rodrguez-Roisin R. Pulmonary gas exchange response to oxygen breathing in acute lung injury. Am J Respir Crit Care Med (2000) 161:26–31.[Abstract/Free Full Text]

28 Whiteley JP, Gavaghan DJ, Hahn CEW. Variation of venous admixture, SF6 shunt, PaO2, and the PaO2/FIO2 ratio with FIO2. Br J Anaesth (2002) 88:771–8.[Abstract/Free Full Text]


Add to CiteULike CiteULike   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us    What's this?


This article has been cited by other articles:


Home page
Br J AnaesthHome page
A. Kathirgamanathan, R. A. McCahon, and J. G. Hardman
Indices of pulmonary oxygenation in pathological lung states: an investigation using high-fidelity, computational modelling
Br. J. Anaesth., August 1, 2009; 103(2): 291 - 297.
[Abstract] [Full Text] [PDF]


This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Supplementary Data
Right arrow All Versions of this Article:
101/3/358    most recent
aen181v1
Right arrow E-Letters: Submit a response to the article
Right arrow Alert me when this article is cited
Right arrow Alert me when E-letters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrowRequest Permissions
Right arrow Disclaimer
Google Scholar
Right arrow Articles by McCahon, R. A.
Right arrow Articles by Hardman, J. G.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by McCahon, R. A.
Right arrow Articles by Hardman, J. G.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?