BJA Advance Access originally published online on April 30, 2007
British Journal of Anaesthesia 2007 98(6):828-834; doi:10.1093/bja/aem090
Systematic errors and susceptibility to noise of four methods for calculating anatomical dead space from the CO2 expirogram
Department of Anaesthetics, Royal Prince Alfred Hospital, University of Sydney, NSW 2050, Sydney, Australia
* Corresponding author: Department of Anaesthetics, Royal Prince Alfred Hospital, University of Sydney, Building 89 Level 4, Missenden Rd, Camperdown, NSW 2050, Australia. E-mail: ytan7763{at}usyd.edu.au
Accepted for publication February 16, 2007.
| Abstract |
|---|
|
|
|---|
Background: Anatomical dead space is usually measured using the Fowler equal area method. Alternative methods include the Hatch, Cumming, and Bowes methods, in which first, second, and third order polynomials, respectively, fitted to an expired CO2 volume vs expired volume curve, intercept the x-axis at the anatomical dead space. This study assessed systematic errors and susceptibility to noise of the Fowler, Hatch, Cumming, and Bowes dead spaces calculated over 4080% of the CO2 expirogram.
Methods: Simulated CO2 expirograms with 220 ml anatomical dead space and varying alveolar plateau slopes were generated digitally and zero-mean Gaussian noise added. CO2 expirograms were recorded in 10 anaesthetized human subjects. Anatomical dead space was calculated by the Fowler, Hatch, Cumming, and Bowes methods.
Results: The Fowler, Hatch, Cumming, and Bowes methods displayed systematic biases of 1.8%, 13.2%, 2.4%, and 1.3%, respectively, at a normalized simulated alveolar plateau slope of 1.6 litre1. At a noise level of 0.0066 vol/vol, the standard deviations of recovered simulated dead spaces were 70.6, 1.8, 2.4, and 3.7 ml, respectively. The Hatch, Cumming, and Bowes methods applied to human expirograms differed significantly from that of Fowler by 13, 4, and 11 ml, respectively. In the human study, the Hatch and Cumming methods yielded the lowest intra-individual dead space variability.
Conclusions: The Fowler method shows greatest susceptibility to measurement noise and the Hatch method exhibits the largest systematic error. The Cumming method, which exhibits both low bias and low noise susceptibility, is preferred for estimating anatomical dead space from CO2 expirograms.
Keywords: airway, dead space; carbon dioxide, measurement; lung, dead space
| Introduction |
|---|
|
|
|---|
The measurement of respiratory dead space, which provides important insight into the efficiency of gas exchange, has found wide application in respiratory physiology, clinical anaesthesia, and critical care medicine.14 Classically, respiratory dead space is divided into two parts: anatomical dead space (VDF) and alveolar dead space. Alveolar dead space is calculated as the difference between physiological dead space (VDphys) and anatomical dead space,5 and can be affected by changes in the distribution of ventilation to perfusion ratio in the lungs.4 Anatomical dead space is thus important because the measurement of alveolar dead space depends on the accurate measurement of anatomical dead space.
Anatomical dead space is commonly calculated by the Fowler equal area method,6 which uses the expired CO2 concentration or partial pressure vs expired volume curve. An alternative is the Hatch7 or Langley8 method in which a straight line is fitted to the most linear part of the expired CO2 volume (VECO2) vs expired volume (VE) curve by linear regression. The intersection of this regression line and the x-axis is an estimate of the anatomical dead space.8 9 Recognizing that a non-zero slope in the alveolar plateau (phase III) of the CO2 expirogram causes a quadratic phase III VECO2VE relationship, Cumming and Guyatt10 and Wolff and colleagues11 fitted a second order polynomial to the VECO2VE data to calculate anatomical dead space. Bowes and colleagues12 used a third order polynomial which allows a changing slope in phase III. Because capnograms are always contaminated with noise,13 and the integration required to calculate VECO2 reduces noise, the Hatch, Cumming, and Bowes methods may yield results of greater precision than the Fowler equal area method.
Since the recommendation of the Hatch method by Fletcher,14 15 numerous authors have used this method to calculate anatomical dead space in various physiological and clinical settings.13 14 1619 The Cumming and Bowes methods, however, have not been used extensively. In this study, we investigated the systematic errors and susceptibilities to measurement noise of the Fowler, Hatch, and modified forms of the Cumming and Bowes methods, using simulated CO2 expirograms and CO2 expirograms recorded in anaesthetized patients without clinically significant lung disease.
| Theoretical considerations |
|---|
|
|
|---|
A CO2 expirogram (Fig. 1A) comprises three distinct phases. In phase I, CO2-free gas is exhaled from the airway dead space. Phase II is a sigmoidal curve with a rapidly increasing CO2 fraction caused by asymmetry in the airway structure and mixing of fresh gas in the airways with CO2-rich gas from the alveoli. Phase III represents alveolar gas and usually has an increasing CO2 concentration. If the lungs are normal, a straight line can be fitted to a selected part of phase III of the curve by linear regression (Fig. 1A):
|
|
|
|
|
|
|
Thus, the part of the idealized VECO2VE curve corresponding to phase III of the CO2 expirogram is a second order polynomial that intersects the x-axis at VDF. This analysis suggests that the part of the VECO2VE curve corresponding to phase III of the CO2 expirogram should preferably be approximated by a second order polynomial rather than by a straight line or a third order polynomial curve, and that the x-axis intercept of a second order polynomial fitted to an appropriately selected part of the VECO2VE curve may be a good estimate of the Fowler dead space.
| Methods |
|---|
|
|
|---|
Simulation study: systematic errors
Idealized CO2 expirograms, F'CO2(VE), were generated by defining
|
|
In all idealized expirograms, the simulated anatomical dead space (V'DF) and tidal volume were 220 and 750 ml, respectively, and the volume of CO2 expired in one breath was 25.2 ml. The normalized slope of phase III, a (absolute slope divided by mean phase III F'CO2),20 was varied from 0 to 1.6 litre1 in steps of 0.4 litre1. The constant b was selected such that all the idealized expirograms passed through the point [(VT+V'DF)/2, 0.05] to obtain an average alveolar FCO2 of 0.05. Idealized numerical expirograms were calculated with 0.35 ml increments in expired volume. Each expirogram comprised 2000 points. The effects of airway asymmetry and gas mixing in the airways were simulated by smoothing the idealized expirogram with a weighted moving average filter. Weighting factors comprised a 701 order Blackman window.21 The smoothed curves were used for estimation of the anatomical dead space by the Fowler equal area,6 Hatch (first order polynomial),7 Cumming (second order polynomial),10 and Bowes (third order polynomial)12 methods.
Dead space calculation methods
Fowler dead space was calculated by the equal area method (Fig. 1A), assuming that the linear section of the alveolar plateau extended from 40% to 80% of expired volume.22 In the Hatch method, a first order polynomial was fitted to the VECO2VE curve between 40% and 80% of the expired volumes by linear regression. The Cumming and Bowes methods were modified by fitting second and third order polynomials, respectively, to the VECO2VE curve between 40% and 80% of the expired volumes. The intersections of the polynomials with the x-axis formed the respective estimates of anatomical dead space. The resulting anatomical dead spaces are referred to as VDH1, VDC2, and VDB3 for the first order polynomial curve method (Hatch method), second order polynomial curve method (modified Cumming method), and third order polynomial curve method (modified Bowes method), respectively. All calculations were performed using computer programs written in Matlab (Mathworks, Natick, MA, USA).
Monte Carlo simulation study: effect of noise in the CO2 measurement
A Monte Carlo simulation study of the effect of CO2 analyser noise on dead space calculation was performed as follows. A smoothed CO2 expirogram was generated using the same parameters described earlier and a normalized phase III slope of 1.2 litre1. Zero-mean normally distributed random noise (SD 0.0013, 0.0026, 0.0039, 0.0053, and 0.0066 vol/vol, corresponding to CO2 partial pressures of 1, 2, 3, 4, and 5 mm Hg, respectively, at a barometric pressure of 760 mm Hg) representing measurement noise, the effects of the limited resolution of the CO2 analyser, and variability in airway CO2 fraction, was added to the expirogram.13 The four dead spaces, VDF, VDH1, VDC2, and VDB3, were estimated using the methods described earlier. This procedure was repeated 1000 times, and the means and standard deviations of the four dead space values were calculated.
Clinical study
After approval by the local ethics committee of the Royal Prince Alfred Hospital and with written informed patient consent, 10 American Society of Anesthesiologists physical status IIIII patients aged 5276 presenting for vascular surgery were enrolled in this study. All patients received a radial artery cannula for haemodynamic monitoring. Routine monitoring included pulse oximetry, ECG, pharyngeal temperature, and capnometry. Anaesthesia was induced with propofol 2 mg kg1, fentanyl 2 µg kg1, and rocuronium 0.8 mg kg1 given i.v. and a cuffed endotracheal tube was inserted. Anaesthesia was maintained with inhalation isoflurane 11.5%, bolus rocuronium 0.3 mg kg1 and fentanyl 12 µg kg1. The lungs were mechanically ventilated with a volume-controlled mechanical ventilator (Cato, Dräger, Germany). Arterial blood pressure was maintained within 20% of baseline with a low-dose infusion of metaraminol (00.05 mg min1). Baseline ventilation parameters were as follows: tidal volume, 10 ml kg1; respiratory frequency, 10 bpm; inspiration to expiration ratio (I:E ratio), 1:1.7; end-inspiratory hold, 10%; end-expiratory pressure, 0 cm H2O; and inspired O2 concentration, 35%.
Airway gas flow, CO2, and pressure were measured by a NICO monitor (Novametrix Medical Systems Inc., New Haven, CT, USA). The response time of the mainstream CO2 analyser was less than 60 ms and the resolution was 2 mm Hg. The standard deviation of noise in the PCO2 data logging system (including the analogue-to-digital converter) was approximately 0.3 mm Hg when the input PCO2 was constant. The airway configuration from proximal to distal was endotracheal tube, catheter mount, airway filter, mainstream infrared CO2 analyser, pneumotachograph, pressure monitor, sidestream O2 analyser, and Y piece of the anaesthetic circuit. Partial pressure of CO2, O2, gas flow and pressure signals in the airway were digitized at a frequency of 300 Hz using a 12-bit analogue-to-digital converter (DAQPad-1200, National Instruments Corporation, TX, USA) and recorded by a computer running Matlab. The CO2 analyser was calibrated with known concentrations of CO2 and the flowmeter was calibrated using a second order polynomial method with a calibrated 3 litre syringe.23 All signals were digitally filtered with a 51st order low pass filter (3 dB frequency 5.0 Hz).
To vary the dead space in each patient,24 25 the respiratory parameters were randomly adjusted one at a time to the following settings: tidal volume: 64%, 80%, 100%, 120%, and 144% of baseline; end-expiratory pressure: 0, 5, and 10 cm H2O; I:E ratio: 1:1.7, 1:1, and 2:1; end-inspiratory hold: 10%, 30%, and 50% of the inspiratory time. Under each condition, 4653 breaths were recorded for off-line analysis. When one parameter was adjusted the other ventilatory parameters were kept at baseline values. A total of 11 different ventilatory settings were used in this study. To reduce variations in the CO2 expirogram caused by external interference, only undisturbed breaths were included in the study. Surgical activity and manual interference with the respiratory circuit caused airway pressure and PCO2 perturbations. When airway pressure increased to more than half the average peak inspiratory pressure and corresponding FE'CO2 was more than half the average FE'CO2 for that ventilatory condition, a valid breath was identified and analysed. For each breath, the four dead spaces (VDF, VDH1, VDC2, and VDB3) were calculated using the methods described earlier. Data are reported as mean (SD) where applicable. Mean values were compared by single factor ANOVA with Bonferroni post hoc test for multiple comparisons. F-tests were used to compare variances. We used non-parametric tests (Friedman and Dunn's multiple comparison tests) to analyse the data which were not normally distributed. Statistical analysis was performed using Prism V3.0 (GraphPad Software, San Diego, CA, USA). P < 0.05 was considered to be statistically significant.
| Results |
|---|
|
|
|---|
Simulation study
Simulated CO2 expirograms with and without added noise (0.0026 vol/vol corresponding to 5% of indicated FCO2 at an FCO2 of 0.05) and corresponding VECO2VE curves are shown in Figure 2A and B. The high slope of phase III caused a high curvature in the VECO2VE curve (dotted curves in Fig. 2A and B). Integration visibly reduced noise in the VECO2VE curve (solid line in Fig. 2B).
|
Simulation study: systematic errors
When the slope of phase III was zero, all four methods recovered anatomical dead spaces within 2.5% of the model dead space (Fig. 3). As the normalized phase III slope increased from 0 to 1.6 litre1, calculated VDH1 increased from 220 to 249 ml, corresponding to a bias of 0.1% to 13.2% compared with the true anatomical dead space of 220 ml. Over the same range of phase III slope, VDF decreased from 220 to 216 ml (0% to 1.8% bias) and VDC2 increased from 218 to 225 ml (0.8% to 2.4% bias). VDB3 slightly underestimated anatomical dead space when the phase III slope was 1.6 litre1 (217 ml, bias 1.3%). In the normal range of normalized phase III slope (0.41.2 litre1), the systematic errors of VDF and VDC2 were low (0.9% to 0.8%).
|
Monte Carlo simulation study: effect of noise in the CO2 measurement
Noise in the CO2 measurement caused a significant change in the mean dead space calculated by the Fowler method (0.0001<P < 0.001) for noise amplitudes of 0.0053 and 0.0066 vol/vol compared with zero-noise (Fig. 4). The standard deviation of the Fowler method increased significantly (0.01<P < 0.05) to 71 ml as the noise increased from 0.0013 to 0.0066 vol/vol. At all noise levels, VDH1 exhibited the lowest standard deviation (range 0.41.8 ml), followed by VDC2 (range 0.52.4 ml), VDB3 (range 0.73.7 ml), and VDF (range 12.070.6 ml). VDH1 also shows a bias of 9.5% at all noise levels corresponding to the bias of the Hatch method at a normalized phase III slope of 1.2 litre1 (Fig. 3). The standard deviations of VDH1, VDC2, and VDB3 were significantly smaller than those of VDF for all non-zero noise levels (0.01<P < 0.05).
|
Clinical study
All patients had oxygen saturations higher than 95% throughout the study. A total of 6903 expirograms were obtained from 10 patients with various ventilator settings. From calculations based on monitoring times and ventilator rates, we estimate that approximately 100 ventilator cycles were not identified as valid breaths and therefore not analysed. The initial non-parametric ANOVA shows that there are significant differences among the dead spaces measured by the four methods (P < 0.0001). Differences between VDH1 and VDF ranged from 40 to 171 ml (mean 13 ml, limits of agreement 37 to 63 ml). Differences between VDC2 and VDF ranged from 75 to 112 ml (mean 4 ml, limits of agreement 36 to 28 ml). Differences between VDB3 and VDF ranged from 158 to 174 ml (mean 11 ml, limits of agreement 60 to 39 ml). All the mean differences were significantly different from zero (0.0001<P < 0.001, Dunn's multiple comparison test). Mean intra-individual standard deviations (95% confidence intervals)26 of dead space calculated from expirograms acquired during baseline ventilation were: VDF: 11 (7.914.1) ml, VDH1: 3.3 (2.34.3) ml, VDC2: 4.3 (3.25.4) ml, and VDB3: 8.3 (5.910.7) ml (Fig. 5). In the 10 patients studied, the intra-individual standard deviations of VDH1 and VDC2 were significantly smaller than those of both VDF and VDB3 (0.001<P < 0.05).
|
The difference between VDH1 and VDC2 was positively correlated with the normalized slope of phase III (R = 0.6667, 0.01<P < 0.05) (Fig. 6). The difference was similar in magnitude to the bias in our simulation study and had a similar dependence on normalized phase III slope. At a normalized phase III slope of zero, the difference between VDC2 and VDH1 predicted by the regression equation was 3.4 ml (95% confidence intervals 2.93.9 ml), which agreed well with our simulation results.
|
| Discussion |
|---|
|
|
|---|
This study using simulated data and clinical measurements shows that anatomical dead space calculated by the Hatch method, in which a straight line is fitted to the phase III portion of the VECO2VE curve, systematically overestimates anatomical dead space when phase III of the expirogram has a positive slope. The bias of the Hatch method increases with increasing phase III slope and reaches clinically significant values (more than 10%) when the phase III slope is greater than approximately 1 litre1. The x-axis intercepts of second and third order polynomials fitted to the phase III portion of the VECO2VE curve (modified Cumming and Bowes methods, respectively) show negligible mean bias compared with the classical Fowler method in both simulation and clinical studies. Our simulation and clinical studies show similar relationships between the bias of the Hatch method and the phase III slope (Figs 3 and 6, respectively). The Hatch (VDH1) and modified Cumming (VDC2) and Bowes (VDB3) methods are significantly less sensitive to noise than the Fowler method in the simulation study and yield significantly lower intra-individual variance in the clinical study. If the PCO2 measurement contains noise with a standard deviation greater than approximately 1 mm Hg, the random errors in anatomical dead space calculated by the Fowler method may be clinically significant (>10%) in 5% or more of measurements. The modified Cumming method in which anatomical dead space is estimated as the x-intercept of a second order polynomial fitted to the VECO2VE curve between 40% and 80% of expired volume yields both low mean bias (4 ml compared with Fowler equal area method in the clinical study) and low noise susceptibility.
In the original Cumming method,10 a second order polynomial is fitted to the VECO2VE data by adding data points to the regression sequentially from the top of the curve downwards, until the addition of further points produces no change in these parameters. We modified the Cumming method using VECO2VE data only between 40% and 80% of the expired volume in the polynomial regressions. This part of the expirogram includes only the approximately linear part of phase III of the CO2 expirogram and does not include the increase in VECO2 associated with closing volume (phase IV) that occasionally occurs. The more pronounced cardiogenic oscillations that are occasionally seen at the end of an expirogram are also omitted. These modifications reduce bias and may explain why this study found good agreement between the Fowler and the modified Cumming methods (4 ml mean difference), whereas Cumming and Guyatt10 found a larger (20 ml) mean difference. We used the same section of the VECO2 vs VE curves in the linear and third order polynomial regressions for the same reasons and for comparative purposes.
Integration reduces noise in a signal; hence, anatomical dead space, estimated from the VECO2 VE curve, is expected to exhibit lower variability than the Fowler dead space. A non-zero phase III slope, however, leads to a quadratic relationship between VECO2 and VE, which results in systematic errors in dead space estimates when a straight line is fitted to the VECO2VE curve (Hatch method). This mechanism is also the reason for the bias of the Hatch method in the noise simulation study. If phase III of the expirogram is not a straight line then a third order polynomial should fit the VECO2VE curve between 40% and 80% of expired volume better than a second order polynomial. Our clinical study suggests that the third order polynomial results in small but statistically significant under-estimation of Fowler dead space, and the additional degree of freedom increases its susceptibility to noise.
The x-axis intercept of a straight line fitted to a VECO2VE curve, as an estimate of anatomical dead space, was first reported by Hatch and colleagues7 in 1953 and apparently independently developed by Langley and colleagues8 in 1975. Fletcher14 15 recommended and used this method in a series of studies on respiratory dead space in 1984. The Hatch method has been widely used and referred to in most citations as the Langley method.13 14 1619 Among the available methods for calculating anatomical dead space (Fowler method, Hatch method,7 8 a pre-interface expirate method,16 27 a threshold method,28 the Lanczos method,10 the Romero method,29 substitution of end-tidal PCO2 in the Bohr equation30), the Hatch method is possibly the second most commonly used after the Fowler method. The results and conclusions of studies that used the Hatch method should be interpreted with caution, as this study suggests that the Hatch method introduces a positive bias in anatomical dead space estimates when phase III of the expirogram has a positive slope.
Our theoretical analysis shows that the Cumming second order polynomial method produces lowest errors independent of the ventilator setting. In the clinical study, we obtained different dead spaces by varying ventilator setting. It is possible that dead space measurement errors might depend slightly on the ventilator setting. As the four methods were applied to the same CO2 expirograms for comparative purposes, we expect the effect of ventilator setting on dead space errors to be negligible. Although the finite resolution of the CO2 analyser and gas flowmeter affects the accuracy of dead space estimates, these errors are very small and are random rather than systematic.
In conclusion, the Fowler equal area method for the calculation of anatomical dead space is susceptible to noise and the Hatch method yields results with systematic errors. The modified Cumming method, in which anatomical dead space is estimated as the x-intercept of a second order polynomial fitted to the VECO2VE curve between 40% and 80% of expired volume, produces both low bias and low noise susceptibility. This study suggests that of the four methods assessed, the modified Cumming method is preferred for estimating anatomical dead space from CO2 expirograms.
| Acknowledgements |
|---|
|
|
|---|
The authors are pleased to acknowledge the financial support of an Australian Research Council Strategic Partnership with IndustryResearch and Training (ARC-SPIRT) grant, Dräger Australia Pty Ltd, The Joseph Fellowship, The Jobson Foundation, The University of Sydney and the Australian National Health and Medical Research Council (NHMRC).
| References |
|---|
|
|
|---|
1 Kline JA, Israel EG, Michelson EA, et al. Diagnostic accuracy of a bedside D-dimer assay and alveolar dead-space measurement for rapid exclusion of pulmonary embolism: a multicenter study. JAMA (2001) 285:7618.
2 Kline JA, Kubin AK, Patel MM, Easton EJ, Seupal RA. Alveolar dead space as a predictor of severity of pulmonary embolism. Acad Emerg Med (2000) 7:6117.[Web of Science][Medline]
3 Noble MI, Langley F, Buckman M, et al. Comparison of ventilation/perfusion lung-imaging and dead-space measurements in airway disease. Clin Sci (Lond) (1981) 60:1723.[Medline]
4 Tang Y, Turner MJ, Baker AB. Effects of alveolar dead-space, shunt and V/Q distribution on respiratory dead-space measurements. Br J Anaesth (2005) 95:53848.
5 Severinghaus JW, Stupfel M. Alveolar dead space. Am J Physiol (1955) 183:660.
6 Fowler WS. Lung function studies. II: The respiratory dead space. Am J Physiol (1948) 154:40516.
7 Hatch T, Cook KM, Palm PE. Respiratory dead space. J Appl Physiol (1953) 5:3417.
8 Langley F, Even P, Duroux P, Nicolas RL, Cumming G. Ventilatory consequences of unilateral pulmonary artery occlusion. Coll Inst Natl Sante Rech Med (1975) 51:20912.
9 Fletcher R, Jonson B, Cumming G, Brew J. The concept of deadspace with special reference to the single breath test for carbon dioxide. Br J Anaesth (1981) 53:7788.
10 Cumming G, Guyatt AR. Alveolar gas mixing efficiency in the human lung. Clin Sci (Lond) (1982) 62:5417.[Medline]
11 Wolff G, Brunner JX, Weibel W, Bowes R, Bertschmann MW. Anatomical and series dead space volume: concept and measurement in clinical praxis. Appl Cardiopulm Pathophysiol (1989) 2:299307.
12 Bowes CL, Richardson JD, Cumming G, Horsfield K. Effect of breathing pattern on gas mixing in a model with asymmetrical alveolar ducts. J Appl Physiol (1985) 58:1826.
13 Raemer DB, Calalang I. Accuracy of end-tidal carbon dioxide tension analyzers. J Clin Monit (1991) 7:195208.[CrossRef][Web of Science][Medline]
14 Fletcher R. Airway deadspace, end-tidal CO2, and Christian Bohr. Acta Anaesthesiol Scand (1984) 28:40811.[Web of Science][Medline]
15 Fletcher R, Jonson B. Deadspace and the single breath test for carbon dioxide during anaesthesia and artificial ventilation. Effects of tidal volume and frequency of respiration. Br J Anaesth (1984) 56:10919.
16 Wolff G, Brunner JX. Series dead space volume assessed as the mean value of a distribution function. Int J Clin Monit Comput (1984) 1:17781.[CrossRef][Medline]
17 Ream RS, Schreiner MS, Neff JD, et al. Volumetric capnography in children. Influence of growth on the alveolar plateau slope. Anesthesiology (1995) 82:6473.[CrossRef][Web of Science][Medline]
18 De Robertis E, Servillo G, Jonson B, Tufano R. Aspiration of dead space allows normocapnic ventilation at low tidal volumes in man. Intensive Care Med (1999) 25:6749.[CrossRef][Web of Science][Medline]
19 Williams EM, Hamilton RM, Sutton L, Viale JP, Hahn CE. Alveolar and dead space volume measured by oscillations of inspired oxygen in awake adults. Am J Respir Crit Care Med (1997) 156:18349.
20 Breen PH, Mazumdar B, Skinner SC. Comparison of end-tidal PCO2 and average alveolar expired PCO2 during positive end-expiratory pressure. Anesth Analg (1996) 82:36873.[Abstract]
21 Mathworks. In: Signal Processing Toolbox-user's Guide (2004) Natick, MA: The Math Works Inc.
22 Schulz H, Schulz A, Eder G, Heyder J. Labeled carbon dioxide (C18O2): an indicator gas for phase II in expirograms. J Appl Physiol (2004) 97:175562.
23 Tang Y, Turner MJ, Yem JS, Baker AB. Calibration of pneumotachographs using a calibrated syringe. J Appl Physiol (2003) 95:5716.
24 Baker AB, Colliss JE, Cowie RW. Effects of varying inspiratory flow waveform time in intermittent positive pressure ventilation. II: various physiological variables. Br J Anaesth (1977) 49:122134.
25 Williams MH Jr, Rayford CM. Effect of variation of tidal volume on size of physiological dead space in dogs. J Appl Physiol (1956) 9:302.
26 Dixon WJ, Massey FJ. Introduction to Statistical Analysis (1983) 4th edn. New York: McGraw-Hill.
27 Astrom E, Niklason L, Drefeldt B, Bajc M, Jonson B. Partitioning of dead spacea method and reference values in the awake human. Eur Respir J (2000) 16:65964.[Abstract]
28 Olsson SG, Fletcher R, Jonson B, Nordstom L, Prakash O. Clinical studies of gas exchange during ventilatory supporta method using the Siemens-Elema CO2 analyzer. Br J Anaesth (1980) 52:4919.
29 Romero PV, Lucangelo U, Lopez Aguilar J, Fernandez R, Blanch L. Physiologically based indices of volumetric capnography in patients receiving mechanical ventilation. Eur Respir J (1997) 10:130915.[Abstract]
30 Riou Y, Leclerc F, Neve V, et al. Reproducibility of the respiratory dead space measurements in mechanically ventilated children using the CO2SMO monitor. Intensive Care Med (2004) 30:14617.[Web of Science][Medline]
![]()
CiteULike
Connotea
Del.icio.us What's this?
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||





Means of VDF are significantly smaller than mean of VDF at zero noise.
