Comparison of stroke volume (SV) and stroke volume respiratory variation (SVV) measured by the axillary artery pulse-contour method and by aortic Doppler echocardiography in patients undergoing aortic surgery
1 Department of Anesthesiology and Critical Care, Centre Hospitalo-Universitaire Pitié-Salpêtrière, APHP, Université Pierre et Marie Curie Paris, France
2 Department of Anaesthesia and Intensive Care, Sheba Medical Center, Tel Aviv University Tel Hashomer, 52621, Israel
*Corresponding author: Département d'anesthésie-réanimation, CHU Pitié-Salpêtrière, 47 Boulevard de l'Hôpital, 75651 Paris Cedex 13, France. E-mail: jean-pierre.goarin{at}psl.aphp.fr
Accepted for publication July 9, 2006.
| Abstract |
|---|
|
|
|---|
Background. The goal of the study was to compare stroke volume (SV) and respiratory stroke volume variation (SVV) measured by pulse-contour analysis and aortic Doppler.
Methods. These were measured by pulse-contour analysis and thermodilution (PiCCO) and by aortic pulsed wave Doppler with transoesophageal echocardiography in patients undergoing abdominal aortic surgery. Simultaneous measurements were done at different times of surgery. All data were recorded on PiCCOwin software and videotape and analysed off-line by a blinded investigator.
Results. A total of 114 measurements were achieved in 20 patients. There was a good correlation and small bias between the PiCCO and the echo-Doppler values of the mean SV [r=0.885; bias=0.2 (8) ml], and between the minimum [r=0.842; bias=1 (9) ml] and maximum SV [r=0.840; bias=2 (10) ml] values.
Conclusions. There is a fair correlation between pulse-contour analysis and aortic Doppler for beat-by-beat measurement of SV but not for calculation of SV respiratory ventilation.
Keywords: heart, cardiac output; monitoring, cardiac; transpulmonary thermodilution, transoesophageal echocardiography
| Introduction |
|---|
|
|
|---|
In mechanically ventilated patients the systolic pressure variation (SPV) and the pulse pressure variation (PPV) have been shown to be sensitive predictors of fluid responsiveness.14 However, both SPV and PPV serve in this regard as surrogate measurements of the stroke volume variation (SVV). The respiratory variations of stroke volume (SV), estimated by the measurement of aortic blood flow velocity5 or velocitytime intergral,6 7 have also been shown to be sensitive indicators of fluid responsiveness. The introduction of a pulse-contour technique for the measurement of continuous cardiac output (CO) has, however, enabled the real-time measurement of individual SV values, and SVV during mechanical ventilation. SVV, as measured by the PiCCO monitor (Pulsion Medical Systems, Munich, Germany), was also found to be a good indicator of fluid responsiveness,810 although one report found a lack of correlation between the SVV and the response of the CO to fluid loading.11 Pinsky12 13 has advised caution in the clinical use of SVV, reasoning that though the mean SV measured by the pulse-contour method has been adequately validated, the arterial pulse-contour technique has not been validated to monitor rapid changes in SV, as may occur over a single breath. Such validation is further complicated by the fact that different monitors use different algorithms for pulse-contour analysis, and therefore the results of any validation process are relevant only to the method studied, and not to the pulse-contour method in general. Although the validation of PiCCO method for measurement of SV using combined pulse-contour and transpulmonary thermodilution has already been done by comparison with pulmonary thermodilution, both methods are equally based on the same StewartHamilton principle. Using a different method, based on completely different physiological principle (StewartHamilton vs Doppler), may improve the validation of the PiCCO system.
We have therefore compared mean values of SV, rapid changes of SV (minimum and maximum SV) and the derived SVV, as measured by the PiCCO monitor, to similar data measured by aortic Doppler during transoesophageal echocardiography (TOE). The study was done in patients undergoing vascular surgery and axillary artery thermistor-tipped catheters were used to measure CO and to obtain the arterial pressure waveform for pulse-contour analysis.
| Methods |
|---|
|
|
|---|
After approval of the study protocol by the institutional review board, 23 consecutive patients undergoing major aortic (aorto-aortic or aorto-femoral bypass) surgery were included in the study. Patients were informed of the use of continuous CO monitoring with the PiCCO system and assessment of cardiac function with TOE during surgery. Exclusion criteria included the presence of atrial fibrillation, aortic valve disease (detected after operation by transthoracic echocardiography) and significant stenosis of the subclavian artery detected preoperatively as part of the Doppler examination of the carotid arteries.
Preoperatively and intraoperatively patients were managed according to the standard clinical practice of the vascular anaesthesia unit. Our routine intraoperative monitoring for major vascular surgery includes the insertion of a central venous catheter, an arterial catheter for continuous arterial pressure recording and the use of TOE throughout the procedure. A 4F thermistor-tipped catheter (Pulsion Medical System, Munich, Germany) was introduced into the axillary artery as described previously.14 The canulation of the axillary artery was done under local anaesthesia in the awake patient before induction of anaesthesia, using a sterile technique. The arterial catheter was then connected to the PiCCO monitor for haemodynamic measurement. After induction of anaesthesia and endotracheal intubation, the patient's lungs were ventilated with a tidal volume of 8 ml kg1 and no PEEP, while maintaining anaesthesia with a continuous infusion of propofol and sufentanil. The ventilatory frequency was initially set up in order to keep end-tidal
between 3.99 and 4.66 kPa and was unchanged during the study period. The range of ventilatory frequencies among patients was 1216 min1. A transoesophageal multiplane probe was then inserted for routine haemodynamic monitoring. Simultaneous measurements by both PiCCO and TOE were done after induction of anaesthesia, before and after aortic cross-clamping, before and after aortic declamping and after fluid challenges consisting of 500 ml of plasma expander (Plasmion, Kabi Pharma, France), which were administered based on clinical judgement and repeated whenever they induced an increase of
15% in the Doppler aortic CO value.
PiCCO measurements
CO was measured in triplicate by the injection of NaCl solution 0.9% (15 ml) at a temperature below 8°C through the central venous catheter, with the thermodilution curve being measured by the axillary artery thermistor-tipped catheter.14 This CO value calibrates automatically the pulse-contour based measurement of continuous SV, with the latter being displayed in the dedicated data acquisition software (PiCCOwin, Pulsion Medical Systems) as individual, mean, maximal or minimal values. In addition, the SVV is automatically calculated using the formula:
![]() |
TOE measurements
Before each measurement, the TOE probe was positioned to obtain the deep trans-gastric longitudinal view of the heart and an alignment with the aortic root. All echo-Doppler measurements were recorded on videotape with the simultaneous airway pressure trace. The echo-Doppler data were recorded by a trained observer and analysed off-line by another investigator blinded to the PiCCO results. Synchronicity of both Doppler and PiCCO measurements was ensured by making a signal on the PiCCOwin software at the end of each echo measurement.
The aortic valve area was calculated from the left ventricular outflow tract diameter (LVOTd) measured at the insertion of the aortic cusp as LVOTa (cm2)=
*(LVOTd2/4). The mean echo-Doppler SV was obtained by multiplying the LV outflow tract area (LVOTa) by the aortic flow velocitytime integral (VTIao) using the formula SV (ml)=LVOTa (cm2)*VTIao (cm). The mean echo-Doppler SV (SV-Emean) was calculated as the average of six consecutive beats. Echo-Doppler calculation of CO was obtained by multiplying the SV-Emean by the heart rate. The maximum and minimum SV values were identified during four consecutive respiratory cycles and averaged to obtain SV-Emax and SV-Emin, respectively.
All results are expressed as mean (SD). Statistical analysis was performed using statistical software (Staview 5.0; SAS Institute, Cary, NC). The echo-Doppler and PiCCO measurements were compared by linear regression using least square variables to evaluate proportional (slope of curve) and constant (y-intercept) errors and bias (agreement), the limits of agreement (SD 2) according to the BlandAltman method and the precision (absolute value of difference between variables). Statistical significance was considered to be at P<0.05.
| Results |
|---|
|
|
|---|
Of the 23 patients, one patient was excluded because of a suboptimal arterial pressure trace, and two others were excluded because of suboptimal alignment of the TOE probe with the LVOT, because of anatomical causes.
A total of 114 measurements were obtained in the 20 patients who were included in the final study population. The median number (range) of measurements per patient was 6 (57).
There was a fair correlation and small bias between the PiCCO and the echo-Doppler values of the minimum and maximum SV (Table 1, Figs 1 and 2) and with the mean SV (Table 1, Fig. 3). Although the mean SVV values were similar when measured by the echo-Doppler and by the PiCCO, the correlation coefficient (r) of SVV measured with echo-Doppler (SVV-E) and with PiCCO (SVV-P), though significant, was only 0.661 with a bias of 1 (5)% (Table 1, Fig. 4).
|
|
|
|
|
| Discussion |
|---|
|
|
|---|
The recent re-introduction of arterial pressure pulse-contour analysis enables the measurement of continuous CO and beat-by-beat SV. This method has been validated by studies that have shown good agreement between CO and mean SV measured by pulse-contour analysis and those measured by the pulmonary thermodilution method. Both thermodilution methods are based on the StewartHamilton principle. The results of our study show for the first time that the mean SV determined by the pulse-contour analysis of the PiCCO system are in good agreement with echo-Doppler measurements.
In addition, the ability to measure individual SV by the pulse-contour method has enabled the introduction of continuous measurement of SVV on some monitors. SVV is a measurement of the change of SV during a mechanical breath. Normally, the main haemodynamic effect of a mechanical breath is a transient decrease in LV preload. However, a more preload-dependent patient will demonstrate a larger decrease in SV and will have, as a consequence, a larger SVV.15 This principle has been applied successfully, both experimentally and clinically, using SPV19 and PPV,20 21 which have been shown to reflect fluid responsiveness much better than filling pressures or volumetric measures of preload such as the global end-diastolic volume (GEDV) and the LV end-diastolic area. However, variations in systolic pressure and pulse pressure serve only as surrogates of SV and, potentially, a direct measurement of SVV may be more accurate.12 13
Nevertheless, like any other new monitored variable, the pulse contour-derived SVV needs to be adequately validated so that its accuracy and limits of performance are well recognized. Such validation process can and should be done by comparing the SVV to a gold standard measurement, by assessing its individual performance as a fluid-predictor, and by comparing it to similar methods (e.g. SPV, PPV) that have been tested previously. Moreover, as different pulse-contour methods use different algorithms, each method has to be individually validated.12 13 SVV measured by the PiCCO monitor has been shown to perform well as predictor of fluid responsiveness.8 10 Two separate groups of investigators have found an SVV cut-off value of 9.5% to predict a positive response of the CO to fluid loading in patients undergoing neurosurgery8 and cardiac surgery.9 In one of these studies, however, SVV had a lesser predictive value in patients with a low preoperative ejection fraction and higher intraoperative LV end-diastolic dimensions compared with patients with normal preoperative cardiac function.16 Another study, in cardiac surgery patients, found no significant relationship between baseline SVV values and the percentage change in CO after volume loading, although there was a significant correlation between baseline SVV values and the change in SVV after volume loading.11 In addition, SVV has been shown to compare favourably with PPV and SPV.17 18
This gradual validation process has not, however, included a comparison of the PiCCO-derived SVV to another method, such as the echo-Doppler, which can measure individual SV values. This point was raised in two separate editorials, leading their author to advise caution in the clinical use of the pulse-contour SVV.12 13
Our results offer only partial proof as to the validity of the PiCCO SVV algorithm. We have shown for the first time that there is a very good agreement between the values of the minimal and maximal SV measured by the PiCCO pulse-contour method and by the echo-Doppler. This is of significance, as the main potential problem of any pulse-contour algorithm is the failure to follow rapid changes in SV, such as those occurring during a mechanical breath. However, there was considerable variation between SVV calculated from the echo-Doppler SVmax and SVmin values and SVV automatically derived by the PiCCO monitor.
The reason for a good agreement of the minimum and maximum SV values on the one hand, and a poorer agreement of the derived SVV on the other, is not immediately apparent. The echo-Doppler derived SVV was calculated using the mean values of the manually measured inspiratory SVmax and expiratory SVmin during four respiratory cycles, as identified by the simultaneously recorded aortic flow and airway pressure. The PiCCO monitor uses a different approach to the calculation of SVV, which is based on the averaging of maximum and minimum SV values during four consecutive periods of 7.5 s. With a ventilatory frequency of at least 8 bpm, each of these time frames should include the changes that occur in SV during at least one respiratory cycle. It is possible that the observed difference in the calculated SVV values stem from a different timing and lack of synchronization of the two methods. We tried to synchronize both measurements by marking the echo-Doppler measurement period on the PiCCOwin data acquisition software, the latter is set by default to save data at intervals of 1 min only. Hence, it may well be that the 1 min recordings of the PiCCOwin data do not overlap with the averaging period of the echo-Doppler data. Thus, the SVmin and SVmax values of the two different methods are not calculated at the exact same time, so that any small change in SV may cause a difference in the respective measurements of SVmax and SVmin. Because of error propagation, the small differences in SVmin and SVmax between the two methods may be enhanced when SVV is calculated. Respiratory movements may also affect the Doppler determination of SVV because of the changing alignment of the Doppler beam. The lesser degree of agreement between echo-Doppler and PiCCO SVV data may also stem from the fact the PiCCO SVV data output has an unexplained variability, which can be seen by the frequent changes in SVV values on the monitor screen. This reason for this variability is not clear but may be the main reason for the differences observed between PiCCO and echo-Doppler SVV data.
There is a fair correlation between pulse-contour analysis and aortic Doppler for beat-by-beat measurement of minimal, maximal and mean SV during aortic surgery but not for calculation of SV respiratory variation.
| References |
|---|
|
|
|---|
1 Coriat P, Vrillon M, Perel A, et al. A comparison of systolic blood pressure variations and echocardiographic estimates of end-diastolic left ventricular size in patients after aortic surgery. Anest Analg 1994; 78:4653
2 Tavernier B, Makhotine O, Lebuffe G, Dupont J, Scherpereel P. Systolic pressure variation as a guide to fluid therapy in patients with sepsis-induced hypotension. Anesthesiology 1998; 89:131321[CrossRef][Web of Science][Medline]
3 Michard F, Chemla D, Richard C, et al. Clinical use of respiratory changes in arterial pulse pressure to monitor the hemodynamic effects of PEEP. Am J Respir Crit Care Med 1999; 159:9359
4 Michard F, Boussat S, Chemla D, et al. Relation between respiratory changes in arterial pulse pressure and fluid responsiveness in septic patients with acute circulatory failure. Am J Respir Crit Care Med 2000; 62:1348
5 Feissel M, Michard F, Mangin I, Ruyer O, Faller JP, Teboul JL. Respiratory changes in aortic blood velocity as an indicator of fluid responsiveness in ventilated patients with septic shock. Chest 2001; 119:86773
6 Beaussier M, Coriat P, Perel A, et al. Determinants of systolic pressure variation after vascular surgery. J Cardiothoracic Vasc Anesth 1995; 9:54751[CrossRef][Web of Science][Medline]
7 Slama M, Masson H, Teboul JL, et al. Respiratory variations of aortic VTI: a new index of hypovolemia and fluid responsiveness. Am J Physiol Heart Circ Physiol 2002; 283:H172933
8 Berkenstadt H, Margalit N, Hadani M, et al. Stroke volume variation as a predictor of fluid responsiveness in patients undergoing brain surgery. Anesth Analg 2001; 92:9849
9 Reuter DA, Felbinger TW, Schmidt C, et al. Stroke volume variations for assessment of cardiac responsiveness to volume loading in mechanically ventilated patients after cardiac surgery. Intensive Care Med 2002; 28:3928[CrossRef][Web of Science][Medline]
10 Marx G, Cope T, McCrossan L, et al. Assessing fluid responsiveness by stroke volume variation in mechanically ventilated patients with severe sepsis. Eur J Anaesth 2004; 21:1328[CrossRef][Web of Science][Medline]
11 Wiesenack C, Prasser C, Rodig G, Keyl C. Stroke volume variation as an indicator of fluid responsiveness using pulse contour analysis in mechanically ventilated patients. Anesth Analg 2003; 96:12547
12 Pinsky M. Functional hemodynamic monitoring. Intensive Care Med 2002; 28:3868[CrossRef][Web of Science][Medline]
13 Pinsky M. Probing the limits of arterial pulse contour analysis to predict preload responsiveness. Anesth Analg 2003; 96:12457
14 Segal E, Berkenstadt H, Katzenelson R, Perel A. Transpulmonary thermodilution cardiac output measurement using the axillary artery in critically ill patients. J Clin Anesth 2002; 14:21013[CrossRef][Web of Science][Medline]
15 Buhre W, Weyland A, Kazmaier S, et al. Comparison of cardiac output assessed by pulse-contour analysis and thermodilution in patients undergoing minimally invasive direct coronary artery bypass grafting. J Cardiothorac Vasc Anesthesia 1999; 13:43740[CrossRef][Web of Science][Medline]
16 Goedje O, Hoeke K, Lichtwark-Aschoff M, et al. Continuous cardiac output by femoral arterial thermodilution calibrated pulse contour analysis: comparison with pulmonary arterial thermodilution. Crit Care Med 1999; 27:240712[CrossRef][Web of Science][Medline]
17 Sakka SG, Reinhart K, Meier-Hellmann K. Comparison of pulmonary artery and arterial thermodilution cardiac output in critically ill patients. Intensive Care Med 1999; 25:8436[CrossRef][Web of Science][Medline]
18 Della Rocca G, Costa MG, Pompei L, et al. Continuous and intermittent cardiac output measurement: pulmonary artery catheter versus aortic transpulmonary technique. Br J Anesth 2002; 88:3506
19 Perel A, Pizov R, Cotev C. Systolic blood pressure variation is a sensitive indicator of hypovolemia in ventilated dogs subjected to graded hemorrhage. Anesthesiology 1987; 67:498502[Web of Science][Medline]
20 Reuter DA, Kirchner A, Felbinger TW, et al. Usefulness of left ventricular stroke volume variation to assess fluid responsiveness in patients with reduced cardiac function. Crit Care Med 2003; 31:1399404[CrossRef][Web of Science][Medline]
21 Reuter DA, Felbinger T, Kilger F, Schmidt C, Lamm P, Goetz AE. Optimizing fluid therapy in mechanically ventilated patients after cardiac surgery by on-line monitoring of left ventricular stroke volume variations: a comparison to aortic systolic pressure variations. Br J Anaesth 2002; 88:1246
![]()