BJA Advance Access originally published online on July 27, 2006
British Journal of Anaesthesia 2006 97(4):453-459; doi:10.1093/bja/ael194
Anticoagulation monitoring during vascular surgery: accuracy of the Hemochron® low range activated clotting time (ACT-LR)
1 Department of Anaesthesiology, Hôpital Foch Suresnes, France
2 Department of Biology, Hôpital Foch Suresnes, France
*Corresponding author. E-mail: m.fischler{at}hopital-foch.org
Accepted for publication June 14, 2006.
| Abstract |
|---|
|
|
|---|
Background. Activated clotting time (ACT) is currently used to monitor high concentrations of heparin anticoagulation. A new instrument, the Hemochron® Jr Signature device, has been specifically designed to measure ACT in low-range heparin plasma concentrations (ACT-LR). The purpose of this study was to compare ACT-LR with anti-Xa activity in patients receiving low-dose i.v. heparin during vascular surgery.
Methods. Thirty patients, undergoing arterial vascular surgery, were included in the study and received unfractionated heparin (initial dose 50 u kg1). One hundred and thirty-two pairs of blood samples were simultaneously collected during surgery to determine ACT-LR and anti-Xa activity. Pearson correlation, Kappa test, ROC curve and a specific clinical interpretation of the correlation were performed.
Results. ACT-LR ranged from 68 to 380 s, anti-Xa activity from 0 to 1.45 u ml1. We observed a strong correlation between anti-Xa activity and ACT-LR (r2=0.87; P<0.0001). Accuracy of ACT-LR was good for anti-Xa activity up to 0.6 u ml1 (Kappa, 0.94; accuracy, 97%) and 0.8 u ml1 (Kappa, 0.79; accuracy, 90%), and poor for anti-Xa activity above 1 u ml1 (Kappa, 0.59). A clinical interpretation of the correlation graph found 98% of measured ACT-LR values to be accurate.
Conclusion. Hemochron® Jr Signature provides measurements of ACT-LR, which are accurate for monitoring heparin anticoagulation at anti-Xa activity below 0.8 u ml1.
Keywords: blood, anticoagulants, heparin; blood, coagulation; blood, whole blood coagulation time; surgery, vascular
| Introduction |
|---|
|
|
|---|
Heparin-induced anticoagulation is currently used during arterial vascular surgery to prevent thrombosis and accumulation of thrombi at sites of vascular injury. Unfractioned heparin (UFH) is administered before clamping and blood flow interruption. The goals are to achieve adequate anticoagulation quickly and to maintain a steady anticoagulation concentration until cross-clamp removal. Although the risk of thrombotic events is not well established in this context there is consensus on the necessity of anticoagulation during vascular surgery.
The 7th ACCP Conference on Antithrombotic and Thrombolytic Therapy recently recommended a fairlyhigh concentration of anticoagulation during surgery with an initial dose of 100150 u kg1 of UFH before cross-clamping.1 It was also recommended to supplement this dose every 50 min until circulation is re-established. But these recommendations apply only to situations where there is no monitoring of heparin anticoagulation such as activated clotting time (ACT). Curiously, no anticoagulation monitoring device was specified. With a high variability of heparin responsiveness and plasma elimination half-life,24 anticoagulation monitoring may improve anticoagulation during vascular surgery. Few studies have emphasized the role of anticoagulation monitoring in this setting.59
ACT was first described by Hattersley10 and represents a whole blood clotting time. Accelerated coagulation is obtained by activation of the contact pathway, using either kaolin or celite as an activator. ACT measurement with point-of-care devices is currently used during procedures requiring anticoagulation, such as cardiopulmonary bypass, interventional cardiology and haemodialysis.11 12 Maintaining adequate anticoagulation by monitoring ACT can prevent thrombosis of intravascular devices or extracorporeal circuits.1315 In these settings, ACT has been well correlated with anti-Xa activity.1618 This correlation was found in high heparin anticoagulation conditions (anti-Xa activity between 4 and 6 u ml1). However, traditional ACT measurement devices may not be accurate at lower heparin anticoagulation concentrations typically encountered in vascular surgery.
Recently, a new instrument for ACT assessment has become available, the Hemochron® Jr II Signature device (ITC, International Technidyne Corp., Edison, NJ, USA). It includes specific cartridges which provide good reproducibility to measure ACT in low anti-Xa activity settings (ACT-LR, low range). ACT-LR has already been compared with laboratory-based anti-Xa activity in patients undergoing extracorporeal membrane oxygenation (ECMO).19 Several factors involved in the relationship between ACT and laboratory assay could explain the poor correlation found.
The purpose of our study was 2-fold: (i) to test the relation between ACT-LR measurements and anti-Xa activity; (ii) to evaluate medical decisions based on ACT-LR monitoring during vascular surgery.
| Materials and methods |
|---|
|
|
|---|
Patients and procedure
Thirty consecutive patients, undergoing arterial vascular surgery, were included in the study. Additional inclusion criteria were (i) age (>18 yr); (ii) ASA physical status (IIII). Exclusion criteria were (i) documented heparin allergy; (ii) chronic renal failure with a clearance below 30 ml min1 (Cockroft formula). The study protocol was approved by the Hospital Ethics Committee. Written, informed consent was obtained from all patients.
All patients were premedicated 1 h before surgery with cetirizine (1 mg kg1). Anaesthesia was induced with sufentanil (0.2 µg kg1), propofol (2 mg kg1), atracurium (0.5 mg kg1) and maintained with isoflurane (0.61%) and supplements of these drugs. All patients were equipped with an arterial catheter, according to the standard protocol of our hospital. Anticoagulation was achieved with 50 u kg1 of bovine sodium heparin (Choay Sanofi-Synthélabo, Paris, France) injected before artery clamping and blood flow interruption. Supplemental 25 u kg1 heparin doses were administered every 60 min until cross-clamp removal. Heparin reversal was obtained with protamine, 50 u kg1 (Choay Sanofi-Synthélabo, Paris, France), injected after cross-clamp removal. None of the patients received aminocapric acid or aprotinin.
Blood samples and ACT-LR measurements were collected at various steps as follows: (i) at baseline; (ii) 5 min after heparin bolus; (iii) during cross-clamp: 15, 30, 45, 60 and 90 min after the initial bolus injection; and (iv) at 5 min after protamine infusion. Samples were drawn from an arterial line with no heparin added to the flush solution. Five millilitres of fresh whole blood were collected to measure anti-Xa activity. Vacutainer® tubes (Becton Dickinson, Meylan, France) containing sodium citrate were used and 1 ml of fresh whole blood was collected to measure ACT-LR.
ACT-LR measurement
ACT-LR (low range) measurements were performed according to the manufacturer's instructions, using the portable Hemochron® Jr II Signature device (ITC, NJ, USA). Specific cartridges containing celite (low range) were inserted into the instrument for 1 min prewarming at 37°C. Celite was used as a contact pathway activator to accelerate coagulation.
Measurement of anti-Xa activity
Anti-Xa activity was immediately determined in citrated plasma. Thereby, the intensity of inhibition of the hydrolysis of a chromogenic substrate by factor Xa was determined in presence of heparinantithrombin complexes (Hyphen-Biomed, Andresy, France).
Statistical analysis
Data were expressed as mean (SD). Spearman's correlation was used to assess the relation between ACT-LR and anti-Xa activity. Receiver operating characteristic (ROC) curves were constructed to evaluate the predictive accuracy of ACT-LR for anti-Xa activity. Three different anti-Xa activity settings were defined. As guidelines recommend therapeutic anti-Xa activity of 0.60.7 u ml1 in venous and arterial thrombosis,1 20 0.6, 0.8 and 1 u ml1 were tested. For each anti-Xa activity target, sensitivity (Se), specificity (Sp), positive predictive value (PPV), negative predictive value (NPV) and accuracy were calculated. Areas under the ROC curves were calculated for each threshold. Finally, the degree of concordance between ACT-LR and anti-Xa activity was studied using the Kappa coefficient. Excellent concordance was defined as Kappa coefficient between 0.81 and 1. Good concordance was defined as Kappa coefficient between 0.8 and 0.61. Below these values concordance was considered low.
As statistical correlation neither include clinical significance of accuracy nor describe the type of eventual estimation errors, the clinical relevance of the ACT determination was examined using a graph derived from the error grid analysis (EGA). This concept was initially developed to assess the accuracy of home blood glucometers.21 The EGA takes into account the clinical implications of any treatment decision based on ACT-LR measurements and is based on the assumption represented on Figure 1. The lower threshold was chosen in respect to therapeutic anti-Xa activity used in venous or arterial thrombosis or acute limb ischaemia.20 Six zones are described on the EGA regression graph (Fig. 4). Upper and lower zones A represent ACT values that either deviated less than 10% from the values obtained with the linear regression equation, or were below 200 s when anti-Xa activity was below 0.6 u ml1. An ACT value in zone A is accurate and a clinical decision based on this value would be safe. Upper and lower zones C represent ACT values, not reflecting anti-Xa activity. An ACT value in zone C is not accurate and a clinical decision based on it could be dangerous. Zone C is labelled as erroneous treatment. Upper and lower zones B represent ACT values which deviate more than 10% from values obtained with the linear regression equation, but are outside zone C. An ACT value in zone B is moderately accurate and an erroneous clinical decision based on it would be safe. Zone B is labelled as benign errors.
|
Statistics were computed with StatView 5.0 (SAS Institute, Cary, CA, USA) and SPSS 10.0 (SPSS France, Paris, France). P<0.05 was considered significant.
| Results |
|---|
|
|
|---|
Blood samples were obtained from 30 consecutive patients [mean age, 69 (10) yr, 4 women, 26 men] included in the protocol. Surgical procedures were carotid endarterectomy (12 patients), infrarenal aorta surgery (10 patients) and peripheral vascular surgery (8 patients). Mean duration of arterial cross-clamp was 25 (10) min for carotid endarterectomy and 68 (23) min for aorta and peripheral vascular surgery. Thirteen patients received aspirin and nine patients received heparin up to the morning of surgery. No thrombotic event occurred during or after the intervention.
A total of 264 duplicate anti-Xa activity and Hemochron® ACT-LR measurements were obtained simultaneously from 132 blood samples. Hemochron® ACT-LR ranged from 68 to 380 s, anti-Xa activity from 0 to 1.45 u ml1.
ACT-LR and anti-Xa activity are reported in Table 1. Results for carotid surgery, and aortic and peripheral vascular surgery are presented separately.
|
Taking into account all determinations, significant correlation was observed between ACT-LR obtained with Hemochron® Jr Signature and anti-Xa activity (n=132; y=148.4+99.67 x, r2=0.87, P<0.001) (Fig. 2).
|
The ROC curves for ACT-LR predicting anti-Xa activity are shown in Figure 3 and the different characteristics of each test are presented in Table 2. For an anti-Xa activity above 0.6 u ml1, the optimum ACT-LR cut-off was 200 s (Fig. 3), with an excellent concordance (Kappa, 0.94), 128 blood samples (97%) were correctly classified by the Hemochron® system. For an anti-Xa activity above 0.8 u ml1, the optimal cut-off was 230 s (Fig. 3), with a good concordance (Kappa, 0.79), 119 blood samples (90%) were correctly classified. For an anti-Xa activity above 1 u ml1, however, the optimum cut-off was 250 s (Fig. 3), with poor concordance (Kappa, 0.59) and only 110 blood samples (83%) were classified correctly.
|
|
The EGA-derived graph is represented in Figure 4. One hundred and two ACT values (77%) fell into zone A (clinically accurate), while 28 (21%) fell into zone B (benign errors). Two ACT values (2%) fell into zone C (erroneous treatment).
|
| Discussion |
|---|
|
|
|---|
ACT was initially described to assess the degree of heparin-induced anticoagulation during procedures such as cardiopulmonary bypass interventional, cardiology and haemodialysis.14 22 23 In all these settings high-concentration anticoagulation is mandatory. Anticoagulation monitoring is not systematically used during vascular surgery. Traditionally, ACT measurement devices are not powerful on low concentrations of anticoagulation (anti-Xa activity below 1 u ml1). We report the first study evaluating the Hemochron® ACT-LR monitoring during arterial vascular surgery. Our results show good correlation between ACT-LR and anti-Xa activity. In this particular setting ACT-LR measurements provide correct estimation of anti-Xa activity in more than 95% of total cases. Accuracy of Hemochron® ACT-LR seems to be best for anti-Xa activity below 1 u ml1. Assessment of ACT-LR appears to be appropriate to evaluate heparin-induced anticoagulation during vascular surgery.
The optimum anticoagulation required for vascular surgery with arterial clamping remains to be determined. For percutaneous coronary intervention the optimum concentration of ACT is defined and linked to outcome.13 Target values during cardiopulmonary bypass have also been evaluated.14 15 Target values for anticoagulation, recommended in similar situations, may be useful to estimate the required dose of heparin during surgical cross-clamp. For venous thrombosis or acute arterial ischaemia, the desired anti-Xa activity concentration is 0.30.7 u ml1, with a recommended initial heparin dose of 5080 u kg1.24 After these assumptions, we tested three concentrations of anti-Xa activity (0.6, 0.8 and 1 u ml1) to evaluate the ability of the Hemochron® JR Signature to detect therapeutic ranges. These choices are subjective, but consistent with clinical practice. The assumption of target values for anticoagulation has allowed biological and clinical evaluation of ACT-LR during vascular surgery.
In this study, heparin anticoagulation was determined by assessing the anti-Xa activity. Although the usual therapeutic doses of heparin could be monitored by the activated partial thromboplastin time (APTT),20 anti-Xa activity seems to be more directly linked to heparin-induced anticoagulation. Indeed there is a wide variation in APTT results among different laboratories.25 The APTT results depend on the laboratory method used, and could range from 48 to 108 s at therapeutic heparin plasma concentration.26 27 For this reason, the use of anti-Xa activity instead of APTT would be more relevant in clinical practice.
One major finding in this study was a positive and significant correlation between anti-Xa activity and ACT-LR. Previous studies have already described good correlation between ACT and anti-Xa activity, specially during cardiac surgery.11 On the other hand, no correlation had been reported for low concentrations of anticoagulation between Hemochron® Jr ACT-LR values and anti-Xa activity during ECMO.19 Our different results could be explained by physiological factors during ECMO, such as intense coagulopathy, hypothermia and haemodilution. This study confirms that low anti-Xa activity is the most appropriate setting for Hemochron® Jr II Signature device with low-range cartridges. Secondarily, we tested its ability to differentiate three different concentrations of anticoagulation. With anti-Xa activity above 0.6 or 0.8 u ml1, the corresponding ACT-LR target values were 200 and 230 s, respectively, for adequate anticoagulation. Within these thresholds, 97 and 90% of ACT-LR values were found to be in accordance with anti-Xa activity. PPV (97 and 96%, respectively) and Sp values (both 97%) make ACT-LR a valuable tool to detect dangerous under-anticoagulation. Based on ACT-LR values, rapid correction could be started safely, helping the physician to maintain optimum anticoagulation during surgery. However, low accuracy was observed for anti-Xa activity above 1 u ml1, ACT-LR does not seem to be accurate for monitoring heparin anticoagulation in this range.
Furthermore, analysis of the EGA-derived graph showed ACT-LR measurements to be clinically reliable. ACT-LR values were in perfect agreement with anti-Xa activity in 77% of total cases, making a consequent medical decision safe. In 21% of total cases, ACT-LR values were in zone B, considered as benign errors. Despite the mild inaccuracy of ACT measurements, consequent clinical decisions would not be dangerous. Indeed, the only consequence was an over-anticoagulation which could induce postoperative bleeding in the absence of protamine administration. Nevertheless, as reported,28 the risk of postoperative bleeding is low even if heparin-induced anticoagulation is not routinely reversed. Finally, only 2% of all ACT-LR measurements, in zone C, were failures which could potentially lead to an inappropriate treatment.
There are, however, some reservations concerning our study. The patients included were heterogeneous. Almost half of the patients took platelet antagonist treatment up to the day of surgery. All patients undergoing endarterectomy received aspirin preoperatively without discontinuation, as recommended by recent guidelines.1 Although activation of coagulation during vascular surgery results from stimulation of both intrinsic and extrinsic pathways, local platelet activation may occur after application of cross-clamp. Indeed, increased platelet aggregation has been found after coronary artery flow interruption.29 We hypothesized that platelet antagonist agents decrease this activation and take part in the prevention of thrombus formation. As previously described, Hemochron® ACT is not correlated with any platelet antagonist agent treatment.30 Also, anti-aggregation induced by platelet antagonists is not taken into account in ACT-LR values. For this reason, it seems reasonable to individualize populations of patients who receive antiplatelet agents in future studies.
To achieve anticoagulation before cross-clamp application, we used a heparin dose of 50 u kg1, which is below the recommended dose of 100150 u kg1.1 Consequently, only low anti-Xa activity were studied. If a higher dose had been used, a poorer correlation between ACT-LR measurements and anti-Xa activity might have been found. These choices are arguable, but the optimal dose of heparin during vascular surgery remains undetermined. We have been using this dose for years in our hospital without notable adverse events. We based our choice of concentration of anticoagulation during cross-clamp on therapeutic recommendations for venous and arterial thrombosis. Although heparin-induced anticoagulation is linked to cardiac outcome,31 no study has investigated the links between different anti-Xa activity during vascular surgery and outcome. Also a further study will be needed to evaluate the use of Hemochron® ACT-LR as an anticoagulation monitor for doses higher than 50 u kg1.
Finally, ACT-LR measurements must be directly compared with other methods of point-of-care testing of heparinization such as conventional ACT and APTT.
In conclusion, our study has demonstrated the accuracy of Hemochron® ACT-LR anticoagulation monitoring during vascular surgery with the use of small doses of heparin (50 u kg1). ACT-LR and anti-Xa activity are well correlated and ACT-LR is clinically relevant for low anti-Xa activity. After cross-clamping, target ACT-LR values correlated with anti-Xa activity and can be used to titrate heparin up to the required concentration of anticoagulation.
| Acknowledgments |
|---|
Work should be attributed to the Department of Anaesthesiology, Hôpital Foch, 40 rue Worth, 92151 Suresnes, France, University Paris Ile de France Ouest, France. The study was supported in part by GAMIDA France (Mr E. Brottier) which provided the Hemochron® Jr II Signature device and the specific cartridges. The authors thank Mr E. Brottier for his support of this study.
| References |
|---|
|
|
|---|
1 Clagett GP, Sobel M, Jackson MR, Lip GY, Tangelder M, Verhaeghe R. Antithrombotic therapy in peripheral arterial occlusive disease: the Seventh ACCP Conference on Antithrombotic and Thrombolytic Therapy. Chest 2004; 126:609S26S
2 Despotis GJ, Joist JH, Hogue CW Jr, et al. The impact of heparin concentration and activated clotting time monitoring on blood conservation. A prospective, randomized evaluation in patients undergoing cardiac operation. J Thorac Cardiovasc Surg 1995; 110:4654
3 de Swart CA, Nijmeyer B, Roelofs JM, Sixma JJ. Kinetics of intravenously administered heparin in normal humans. Blood 1982; 60:12518
4 McAvoy TJ. Pharmacokinetic modeling of heparin and its clinical implications. J Pharmacokinet Biopharm 1979; 7:33154[CrossRef][ISI][Medline]
5 Mabry CD, Thompson BW, Read RC, Campbell GS. Activated clotting time monitoring of intraoperative heparinization: our experience and comparison of two techniques. Surgery 1981; 90:88995[ISI][Medline]
6 Poisik A, Heyer EJ, Solomon RA, et al. Safety and efficacy of fixed-dose heparin in carotid endarterectomy. Neurosurgery 1999; 45:43441[CrossRef][ISI][Medline]
7 Martin P, Greenstein D, Gupta NK, Walker DR, Kester RC. Systemic heparinization during peripheral vascular surgery: thromboelastographic, activated coagulation time, and heparin titration monitoring. J Cardiothorac Vasc Anesth 1994; 8:1502[Medline]
8 Szalados JE, Ouriel K, Shapiro JR. Use of the activated coagulation time and heparin doseresponse curve for the determination of protamine dosage in vascular surgery. J Cardiothorac Vasc Anesth 1994; 8:5158[CrossRef][Medline]
9 Castellan L, Causin F, Danieli D, Perini S. Carotid stenting with filter protection. Correlation of ACT values with angiographic and histopathologic findings. J Neuroradiol 2003; 30:1038[ISI][Medline]
10 Hattersley PG. Activated coagulation time of whole blood. JAMA 1966; 196:43640[CrossRef][Medline]
11 Despotis GJ, Gravlee G, Filos K, Levy J. Anticoagulation monitoring during cardiac surgery: a review of current and emerging techniques. Anesthesiology 1999; 91:112251[CrossRef][ISI][Medline]
12 Rath B and Bennett DH. Monitoring the effect of heparin by measurement of activated clotting time during and after percutaneous transluminal coronary angioplasty. Br Heart J 1990; 63:1821
13 Chew DP, Bhatt DL, Lincoff AM, et al. Defining the optimal activated clotting time during percutaneous coronary intervention: aggregate results from 6 randomized, controlled trials. Circulation 2001; 103:9616
14 Bull BS, Korpman RA, Huse WM, Briggs BD. Heparin therapy during extracorporeal circulation. I. Problems inherent in existing heparin protocols. J Thorac Cardiovasc Surg 1975; 69:67484[Abstract]
15 Young JA, Kisker CT, Doty DB. Adequate anticoagulation during cardiopulmonary bypass determined by activated clotting time and the appearance of fibrin monomer. Ann Thorac Surg 1978; 26:23140[Abstract]
16 Culliford AT, Gitel SN, Starr N, et al. Lack of correlation between activated clotting time and plasma heparin during cardiopulmonary bypass. Ann Surg 1981; 193:10511[ISI][Medline]
17 Despotis GJ, Summerfield AL, Joist JH, et al. Comparison of activated coagulation time and whole blood heparin measurements with laboratory plasma anti-Xa heparin concentration in patients having cardiac operations. J Thorac Cardiovasc Surg 1994; 108:107682
18 Paniccia R, Fedi S, Carbonetto F, et al. Evaluation of a new point-of-care celite-activated clotting time analyzer in different clinical settings. The i-STAT celite-activated clotting time test. Anesthesiology 2003; 99:549[CrossRef][ISI][Medline]
19 Ambrose TM, Parvin CA, Mendeloff E, Luchtman-Jones L. Evaluation of the TAS analyzer and the low-range heparin management test in patients undergoing extracorporeal membrane oxygenation. Clin Chem 2001; 47:85866
20 Hirsh J and Raschke R. Heparin and low-molecular-weight heparin: the Seventh ACCP Conference on Antithrombotic and Thrombolytic Therapy. Chest 2004; 126:188S203S
21 Cox DJ, Clarke WL, Gonder-Frederick L, et al. Accuracy of perceiving blood glucose in IDDM. Diabetes Care 1985; 8:52936[Abstract]
22 Cohen M. Monitoring anticoagulation during percutaneous coronary interventions. J Thromb Thrombolysis 1995; 1:2858[CrossRef][Medline]
23 Seifert R, Borchert W, Letendre P, Knutson R, Cipolle R. Heparin kinetics during hemodialysis: variation in sensitivity, distribution volume, and dosage. Ther Drug Monit 1986; 8:326[ISI][Medline]
24 Buller HR, Agnelli G, Hull RD, Hyers TM, Prins MH, Raskob GE. Antithrombotic therapy for venous thromboembolic disease: the Seventh ACCP Conference on Antithrombotic and Thrombolytic Therapy. Chest 2004; 126:401S28S
25 Olson JD, Arkin CF, Brandt JT, et al. College of American Pathologists Conference XXXI on laboratory monitoring of anticoagulant therapy: laboratory monitoring of un-fractionated heparin therapy. Arch Pathol Lab Med 1998; 122:78298[ISI][Medline]
26 Brill-Edwards P, Ginsberg JS, Johnston M, Hirsh J. Establishing a therapeutic range for heparin therapy. Ann Intern Med 1993; 119:1049
27 Bates SM, Weitz JI, Johnston M, Hirsh J, Ginsberg JS. Use of a fixed activated partial thromboplastin time ratio to establish a therapeutic range for un-fractionated heparin. Arch Intern Med 2001; 161:38591
28 Dorman BH, Elliott BM, Spinale FG, et al. Protamine use during peripheral vascular surgery: a prospective randomized trial. J Vasc Surg 1995; 22:24855[CrossRef][ISI][Medline]
29 Gasperetti CM, Gonias SL, Gimple LW, Powers ER. Platelet activation during coronary angioplasty in humans. Circulation 1993; 88:272834
30 Bélisle S. Monitorage de l'héparine durant la circulation extracorporelle. Anticoagulation et réversion. ITBM RBM News 2003; 24:59S66S[CrossRef]
31 Thompson JF, Mullee MA, Bell PR, et al. Intraoperative heparinisation, blood loss and myocardial infarction during aortic aneurysm surgery: a Joint Vascular Research Group study. Eur J Vasc Endovasc Surg 1996; 12:8690[CrossRef][ISI][Medline]
Read all E-letters![]()
CiteULike
Connotea
Del.icio.us What's this?
E-letters:
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||



