British Journal of Anaesthesia 2006 97(2):262-263; doi:10.1093/bja/ael163
This Article
Right arrow Full Text (PDF)
Right arrow E-Letters: Submit a response to the article
Right arrow Alert me when this article is cited
Right arrow Alert me when E-letters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrow Search for citing articles in:
ISI Web of Science (2)
Right arrowRequest Permissions
Right arrow Disclaimer
Google Scholar
Right arrow Articles by Severinghaus, J. W.
Right arrow Articles by De Wolf, A. M.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Severinghaus, J. W.
Right arrow Articles by De Wolf, A. M.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?


© The Board of Management and Trustees of the British Journal of Anaesthesia 2006. All rights reserved. For Permissions, please e-mail: journals.permissions@oxfordjournals.org

Can large volume N2O uptake alone explain the second gas effect?

*E-mail: jwseps{at}comcast.net

Editor—Hendrickx and colleagues1 report that the second gas effect of N2O, increasing the rate of uptake of sevoflurane during induction of general anaesthesia, is not diminished even when the sevoflurane is started after an hour of saturation of the body with N2O. This conclusion resulted from demonstration that the ratio of expired to inspired sevoflurane was about 5% higher in the presence of N2O than with oxygen as the other inspired gas. This conflicts with theoretical analysis of the second gas effect, which should depend on mass transfer into the lung of an additional soluble agent such as sevoflurane.

This surprise finding casts suspicion on the analytical accuracy of the Datex analyser. The authors do not report tests of the possible interactions of the CO2 or water vapour in expired air with N2O on sevoflurane sensitivity in the device. Almost all clinical anaesthetic gas analysers use the infrared (IR) absorption method and attempt to compensate for all the possible interactions. Some of the interactions are physical effects known as pressure broadening. These may be sufficiently complex to cause incomplete compensation in mixtures of many gases as during anaesthesia. CO2 and N2O share some IR absorption wavelengths.

By mass spectrometry or gas chromatography (GC), the authors should rule out the possibility that their IR analyser exposed to a mixture containing 5% CO2, 25% O2, 70% N2O and sevoflurane reads about 5% higher sevoflurane than 5% CO2 in O2 or N2 with a known sevoflurane concentration. A possible effect of water vapour might also be worth testing.

The authors speculate that pulmonary ventilation to perfusion inhomogeneity might explain this effect, and provide some background evidence in support. The idea is interesting, but I cannot rationalize it being so independent of the uptake of N2O.

J. W. Severinghaus*

San Francisco, USA


 

Editor—We are delighted to have captured Prof. Severinghaus' interest in our findings on the second gas effect of N2O on sevoflurane.1 As described by Prof. Severinghaus himself, IR analysers are known to be affected by the presence of other gases.2 The M-CAiOV compact gas analyser used in our study [General Electric (Datex-Ohmeda), Helsinki, Finland] is supposedly less affected by the presence of other gases. First, the M-CAiOV uses an IR wavelength of 8–9 µm to eliminate the effect of CO2 and to minimize (but not totally eliminate) the effect of N2O (older analysers used 3.3 µm). In addition, five different wavelengths between 8 and 9 µm are used for automated agent detection, and a correction factor has been added for the effect of N2O for each of these five wavelengths. Also, NafionTM tubing equalizes the water vapour pressure in the sample to that in the atmosphere. All this recent technology should have improved the accuracy of IR analysis and reduced the errors introduced by the presence of other gases. However, following Prof. Severinghaus' sage advice, we decided to examine the analytic accuracy of the IR analyser in the M-CAiOV by comparing its measurements to those of GC (Gow-Mac 580, Bethlehem, PA, USA) under various conditions.

To examine whether the IR response was linear over the sevoflurane concentration range relevant to the study (1–2%), gas samples with sevoflurane {approx}1 and {approx}2% in O2 were analysed by IR and GC. To examine the effect of carrier gas composition, various mixtures of sevoflurane 1 or 2% in O2, N2O, CO2, N2 and H2O were prepared (see below), and analysed by both techniques. To examine whether different M-CAiOV modules differ in their IR analyser performance, four different analysers were tested.

Using a conventional anaesthesia machine, CO2 from an E-cylinder and exhaled air to provide water vapour, the following gas mixtures were prepared with the following approximate concentrations: sevoflurane 1 and 2% in O2 (O2 groups); sevoflurane 1% in 30% O2 and 70% N2O (N2O group); sevoflurane 1% in 28% O2, 66% N2O, and 5% CO2 (N2O + CO2 group); sevoflurane 1% in air (N2 group); and sevoflurane 1 and 2% in O2+N2O+N2+CO2+H2O (H2O group). For the O2+N2O+CO2 mixture, 5 ml CO2 was added to a 95 ml gas mixture with sevoflurane ~1%, 30% O2, and 70% N2O, yielding 5% CO2 in the final mixture. To examine the effect of water vapour, 40–50 ml of the middle portion of an exhaled breath of one of the investigators (J.F.A.H.) was mixed with 3 ml CO2 and 40–50 ml of either sevoflurane 2% (analyser I and IV) or 4% (analyser II and III) in O2+N2O, yielding the following approximate concentrations: sevoflurane 1 or 2%, 21% O2, 31% N2O, 39% N2, 5% CO2, 3% H2O. All samples were drawn into 100 ml glass syringes. To ensure adequate mixing of the gases, the mixtures were injected back and forth at least four times into another 100 ml glass syringe. Immediately after mixing, the first and last 10 ml were injected into the gas chromatograph; the middle portion, 80 ml, was sampled by the IR analyser. The GC was calibrated with a calibration gas with sevoflurane 1.409%. The four IR analysers were calibrated using the appropriate calibration gas from the company.

Because new mixtures were made for each of the four analysers tested, none of the mixtures contained exactly the same sevoflurane concentration; this is illustrated by the different results for the GC analysis for samples with supposedly the same concentration (Table 1). We therefore calculated the ratio of the GC over the IR value, expressed in %. These GC/IR ratios were statistically analysed using the t-test to compare O2 groups and two-way ANOVA to determine the effect of the constitution of the carrier gas. The results are presented in Table 1. The response of the analysers was linear between 1 and 2% (P>0.05), carrier gas composition had no significant effect, but the analysers differed from one another, except for analyser II and IV.


View this table:
[in this window]
[in a new window]
 
Table 1 Sevoflurane concentrations (%) measured by gas chromatography (GC) and infrared (IR) analysis in different carrier gases

 
How do the results pertain to the interpretation of the results of our previous study? Individual modules differ slightly but unpredictably in their accuracy, but this is not an issue because the same module was used for all patients. Second, the module's response is linear in the range of the sevoflurane concentrations relevant to our study, 1–2%. Third, carrier gas composition has minimal effect on IR analysis, making it unlikely that the in- and end-expired sevoflurane concentrations in the O2 and O2/N2O groups would be differently affected. Within each of the O2/N2O groups, the inspired and end-expired N2O concentrations differ by less than 10% (absolute value), further reducing the possibility of N2O differently affecting the in- and expired sevoflurane concentrations in the N2O groups. Finally, we reported FA/FI, not FA and FI themselves, a distinction that is important. Indeed, if a particular IR analyser would overestimate the ‘true’ sevoflurane concentration by 5%, the ratio of FA/FI would become FA*1.05/FI*1.05; in other words, IR technology should give us a virtually identical FA/FI ratio. Therefore, in our opinion, the potential small error introduced by the presence of N2O on IR measurement of sevoflurane concentration does not annihilate the results of our study.

We conclude that the accuracy of the IR analyser was sufficient to support the conclusion of our study: the second gas effect cannot be adequately explained only by large amounts of N2O uptake. However, we concur with Prof. Severinghaus that GC remains the gold standard to measure absolute concentrations of inhaled anaesthetics.

J. F. A. Hendrickx1,*, R. Carette2, H. J. M. Lemmens1 and A. M. De Wolf3

1 Stanford, CA, USA
2 Aalst, Belgium
3 Chicago, IL, USA

*E-mail: jcnwahendrickx{at}yahoo.com

References

1 Hendrickx JFA, Carette R, Lemmens HJM, De Wolf AM. Large volume N2O uptake alone does not explain the second gas effect of N2O on sevoflurane during constant inspired ventilation. Br J Anaesth 2006; 96:391–5[Abstract/Free Full Text]

2 Severinghaus JW, Larson CP, Eger EI. Correction factors for infrared carbon dioxide pressure broadening by nitrogen, nitrous oxide and cyclopropane. Anesthesiology 1961; 22:429–32[Medline]


Add to CiteULike CiteULike   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us    What's this?


This article has been cited by other articles:


Home page
Anesth. Analg.Home page
P. J. Peyton, M. Chong, C. Stuart-Andrews, G. J. B. Robinson, R. Pierce, and B. R. Thompson
Measurement of Anesthetics in Blood Using a Conventional Infrared Clinical Gas Analyzer
Anesth. Analg., September 1, 2007; 105(3): 680 - 687.
[Abstract] [Full Text] [PDF]


This Article
Right arrow Full Text (PDF)
Right arrow E-Letters: Submit a response to the article
Right arrow Alert me when this article is cited
Right arrow Alert me when E-letters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrow Search for citing articles in:
ISI Web of Science (2)
Right arrowRequest Permissions
Right arrow Disclaimer
Google Scholar
Right arrow Articles by Severinghaus, J. W.
Right arrow Articles by De Wolf, A. M.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Severinghaus, J. W.
Right arrow Articles by De Wolf, A. M.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?