BJA Advance Access originally published online on May 4, 2006
British Journal of Anaesthesia 2006 96(6):708-714; doi:10.1093/bja/ael093
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Tissue oxygenation response to mild hypercapnia during cardiopulmonary bypass with constant pump output
1 Outcomes Research Institute, University of Louisville Louisville, KY, USA
2 Department of Anesthesiology and Perioperative Medicine, Neuroscience Intensive Care Unit, University of Louisville Louisville, KY, USA
3 Department of Surgery, Division of Cardiothoracic and Vascular Surgery, University of Louisville Louisville, KY, USA
4 Department of Outcomes Research, Cleveland Clinic Foundation Cleveland, OH, USA
5 Cardiothoracic and Vascular Surgery Team, Jewish Hospital Louisville, KY, USA
*Corresponding author: Outcomes Research Institute, 501 E. Broadway, Suite 210, Louisville, KY 40202, USA. E-mail: ozan.akca{at}louisville.edu
Accepted for publication March 15, 2006.
| Abstract |
|---|
|
|
|---|
Background. Tissue oxygenation is the primary determinant of wound infection risk. Mild hypercapnia markedly improves cutaneous, subcutaneous (s.c.), and muscular tissue oxygenation in volunteers and patients. However, relative contributions of increased cardiac output and peripheral vasodilation to this response remains unknown. We thus tested the hypothesis that increased cardiac output is the dominant mechanism.
Methods. We recruited 10 ASA III patients, aged 4065 yr, undergoing cardiopulmonary bypass for this crossover trial. After induction of anaesthesia, a Silastic tonometer was inserted s.c. in the upper arm. S.C. tissue oxygen tension was measured with both polarographic electrode and fluorescence-based systems. Oximeter probes were placed bilaterally on the forehead to monitor cerebral oxygenation. After initiation of cardiopulmonary bypass, in random order patients were exposed to two arterial CO2 partial pressures for 30 min each: 35 (normocapnia) or 50 mm Hg (hypercapnia). Bypass pump flow was kept constant throughout the measurement periods.
Results. Hypercapnia during bypass had essentially no effect on
, mean arterial pressure, or tissue temperature.
and pH differed significantly. S.C. tissue oxygenation was virtually identical during the two
periods [139 (50163) vs 145 (38158), P=0.335] [median (range)]. In contrast, cerebral oxygen saturation (our positive control measurement) was significantly less during normocapnia [57 (2867)%] than hypercapnia [64 (3789)%, P=0.025].
Conclusions. Mild hypercapnia, which normally markedly increases tissue oxygenation, did not do so during cardiopulmonary bypass with fixed pump output. This suggests that hypercapnia normally increases tissue oxygenation by increasing cardiac output rather than direct dilation of peripheral vessels.
Keywords: carbon dioxide, hypercapnia; carbon dioxide, hypercarbia; complications, acidosis, respiratory; heart, cardiac output; oxygenation, tissue, cutaneous; oxygenation, tissue, subcutaneous
| Introduction |
|---|
|
|
|---|
Wound infections are among the most common serious complications of anaesthesia and surgery.1 All wounds become contaminated. Contamination progresses to a clinical infection when host defence fails. This process occurs during a decisive period lasting several hours after contamination;2 in other words, during the immediate perioperative period. Oxidative killing by neutrophils is the primary determinant of host resistance to bacterial contamination.3 It is thus unsurprising that factors which increase subcutaneous (s.c.) tissue oxygenation reduce infection risk.4
Determinants of s.c. oxygen availability include arterial oxygen pressure,4 haemoglobin concentration,5 cardiac output,6 local perfusion,7 fraction of fat in s.c. tissue,8 hydration status9 and autonomic responses to pain.10 Mild-to-moderate hypercapnia also improves tissue oxygen availability.11 For example, s.c. oxygenation increased from 63 (14) to 89 (19) mm Hg (P=0.01) when arterial PCO2 increased from 35 to 50 mm Hg.11 This 26 mm Hg increase is likely to be clinically important.12
There are two mechanisms by which hypercapnia might improve tissue oxygenation. The first is activation of central13 and peripheral14 chemoreceptors, which promotes release of catecholamines.15 16 Catecholamines, in turn, increase cardiac output,11 17 enhance systemic vascular conductance14 18 and increase venous return.13 19 20 Interestingly, hypercapnia is not associated with increased blood pressure,11 at least not during anaesthesia, because baroreflex-mediated parasympathetic activity reduces systemic vascular resistance (SVR) and heart rate.17 21 The second mechanism by which hypercapnia might improve tissue oxygenation is direct, locally medicated peripheral vasodilation that may overwhelm sympathetic vasoconstriction.22 23
The relative contribution of increased cardiac output and direct peripheral vasodilation remains unknown; that is, peripheral capacitance vessels could dilate to absorb increased cardiac output or cardiac output could increase in response to systemic vasodilation. Normally, it would be difficult to separate peripheral vasodilation caused by the local action of CO2 from mechanisms that involve increased cardiac output and central autonomic regulation of the circulation. An exception, though, is during cardiopulmonary bypass when the bypass pump can control systemic perfusion. We, therefore, tested the hypothesis that mild hypercapnia increases peripheral tissue and cerebral oxygenation during cardiopulmonary bypass even when the pump flow (cardiac output) remains constant.
Among the various systems currently used for tissue oxygen monitoring, the Clark-type electrode-based method is considered the gold standard. The disadvantage of this method is that the Clark-type electrode consumes oxygen itself, which can decrease the system's accuracy, especially at lower tissue oxygen tensions. Ruthenium-based fluorescent oxygen probe (optode) systems do not have this disadvantage, but have not been fully validated for clinical use. We thus took this opportunity to validate a new fluorescence optode oxygen-monitoring system by comparing it with the standard polarographic electrode (Clark-type) method and evaluating practical factors related to its use.
| Methods |
|---|
|
|
|---|
With the approval of Human Studies Committee at the University of Louisville and written informed consent of patients, we enrolled 10 ASA physical status III patients aged 4065 yr, undergoing heart surgery on cardiopulmonary bypass. Patients were excluded if they were morbidly obese (BMI>35), had diabetes or peripheral vascular disease of any type, or were taking
2-agonists or other vasoactive drugs. Patients who smoked were included, but were not permitted to smoke during the 8 h before surgery because smoking reduces tissue oxygenation for 1 h.24 Using a crossover study design, we compared s.c. tissue oxygen pressure and cerebral oxygen saturation during normocapnia and mild hypercapnia. Based on previous data,11 we expected that mild hypercapnia (a 1520 mm Hg increase in arterial PCO2) would increase s.c. tissue oxygen pressure by at least 15 mm Hg. Assuming, as in our previous studies, that the SD would be 15 mm Hg, we estimated that a sample size of 10 would provide an 80% power to detect a statistically significant effect of hypercapnia at an alpha level of 0.05.
Protocol
Anaesthesia was induced with etomidate (0.3 mg kg1), succinylcholine (
1 mg kg1), and fentanyl (34 µg kg1), and maintained with pancuronium (0.01 mg kg1 as needed), fentanyl (
10 µg kg1 bolus before sternotomy then 23 µg kg1as needed) and isoflurane. Hypovolaemia reduces s.c. perfusion;9 therefore, patients were kept well hydrated, especially during the pre-bypass period, by an infusion of crystalloid at a rate of 68 ml kg1 h1. Patients were actively warmed to maintain core temperature near 36°C before and after the bypass procedure. Because local warming influences tissue oxygenation, the upper arm tonometer site (described below) was protected from active surface warming.
After initiation of cardiopulmonary bypass, pump flow was set to maintain sufficient perfusion with a constant flow at a target mean arterial pressure of 60 mm Hg. Once set, pump outflow was kept constant until the end of the bypass period. Patients were alternately exposed to a
of 35 and 50 mm Hg during bypass, using a crossover design. Each designated
tension was maintained for about 30 min. CO2 was added to the oxygenator circuit by the perfusionist, as necessary, to maintain the designated
. Treatment order was randomly assigned and based on computer-generated codes that were maintained in sealed, sequentially numbered opaque envelopes until the beginning of the bypass procedure. Intraoperatively (including the bypass period), the inspired oxygen concentration (
) was kept constant for each patient [average of all patients was 0.85 (0.10)].
Measurements
Demographic and morphometric characteristics and potential confounding factors, including preoperative haemoglobin, smoking status, type and duration of surgery, total amount of crystalloids given and intraoperative core temperature were recorded. Core temperature was measured from the distal oesophagus (Mallinckrodt Anaesthesiology Products, St Louis, MO, USA) and Swan-Ganz catheter. Bladder temperature, which was recorded from the urinary catheter, was used to represent core temperature during the bypass period because the chest cavity was open.
After induction of anaesthesia, 56 cm of a 15 cm-long Silastic tonometer (1 mm outer diameter, 0.8 mm inner diameter) was inserted s.c. on the lateral aspect of one upper arm, which was then gently tucked at the patient's side. S.C. tissue oxygen tension (PsqO2) was measured with a polarographic electrode system (Licox Medical Systems, Corp., Greenvale, NY, USA).25 The polarographic oxygen electrode was calibrated on room air at the beginning of each surgery (154 mm Hg), and then positioned within the Silastic tonometer described above. A thermocouple was inserted into the opposite lumen of the tonometer and positioned approximately 1 cm from the oxygen electrode. The system was filled with hypoxic saline to remove air from the catheter.
Calibration and stabilization of the polarographic system required 2030 min. In vitro accuracy of the polarographic electrodes (in a water bath at 37°C) is ±3 mm Hg for 0100 mm Hg O2 and ±5% for 100360 mm Hg O2. Temperature sensitivity is 0.25% per degree centigrade, but temperature-compensation is included in the
calculations. Polarographic electrode calibration remains stable (within 8% of baseline value for room air) in vivo for at least 8 h. Consequently, polarographic electrodes measure oxygen tension accurately and reliably over a broad range of s.c. temperatures and PO2 values.25
S.C. tissue oxygen tension was also measured from the same silicon tubing with the FOXY fluorescence system (Ocean Optics Inc., Dunedin, FL, USA), a relatively new technology. The FOXY-PI600 oxygen probe was simultaneously inserted into the silicon catheter. The fibre-optic probe conveys excitation light produced by a blue light emitting diode (LED) to a solgel thin film coating on the membrane at the probe tip. Fluorescence quenching of this probe is related to the partial pressure of oxygen by the SternVolmer equation.26 Fluorescence generated at the tip is collected by the probe and carried by optical fibre to the detector.27 We performed a two-point calibration of the optode at 0 and 154 mm Hg (
: 0 and 0.21) before insertion.
Cerebral oxygen saturation was monitored by a non-invasive near-infrared reflectance spectroscopic oximeter (INVOS 3100, Somanetics, Troy, MI, USA).28 Cerebral oximetry reflects the balance between regional oxygen supply and demand.29
Data analysis
The major outcomes of our study were s.c. oxygen tension, as measured by two different systems, and cerebral oxygen saturation.
Intraoperative values were recorded at 5 min intervals. Tissue oxygenation values were monitored continuously and recorded at the end of each 30 min interval. The average values of the tissue oxygen gathered at the end of each 30 min interval were then compared between the periods. Demographic and morphometric factors were recorded.
Outcomes were compared between the groups with paired, two-tailed t-tests. Non-normally distributed results were analysed by Wilcoxon signed-rank tests. Results are presented as means (SDs or 95% confidence intervals), actual values, median (range), or percentages; P<0.05 was considered statistically significant. Additionally, two-period crossover methodology described by Hills and Armitage30 was used to investigate treatment effects and possible period effects for the three major outcomes.
Linear regression and BlandAltman analysis were used to compare tissue oxygen tensions obtained from the polarographic electrode (Licox) and fluoroscopic optode (Foxy) systems.
| Results |
|---|
|
|
|---|
The patients were 56 (4262) yr old and three were women. Three of the ten were obese (BMI between 30 and 35 kg m2). The duration of surgery averaged 3.5 h, which included 11.5 h of cardiopulmonary bypass. Five patients had coronary artery bypass graft (CABG) surgery, five had valve replacement surgery, and one had both arterial grafting and a valve replacement. All but one of the patients had a history of cigarette smoking. Preoperative haemoglobin and other laboratory values were within normal range (Table 1).
|
During bypass, measurements of
,
, pump outflow, mean arterial pressure, core (bladder) temperature, and s.c. tissue temperature were similar during the normocapnic and hypercapnic periods (Table 2). Arterial CO2 pressures (
) and pH differed significantly, as expected (Table 2). No vasopressor agents were administered before or during bypass.
|
was virtually identical during each study period. Although the relationship is not known to be linear, s.c. oxygen partial pressure (PsqO2) was approximately one-third that of the
(
120 mm Hg vs 365 mm Hg).
during normocapnia (PCO2 of 35 mm Hg) was similar to that during hypercapnia (PCO2 of 50 mm Hg), whether measured by the Licox or Foxy system (Table 2 and Fig. 1).
|
On the other hand, hypercapnia caused a clinically important and statistically significant increase in the cerebral oxygen saturation despite constant bypass pump flow and mean arterial pressure [55 (14)% vs 64 (17)%, P=0.025, Table 2 and Fig. 2].
|
Using the two-period crossover methodology,30 PsqO2, as measured with Licox, had a non-significant treatment effect of 4.9 mm Hg (P=0.213, 95% CI 3.3 to 13.1) and a non-significant period effect of 5.1 mm Hg (P=0.194, 95% CI 3.1 to 13.3). PsqO2, as measured with the Foxy system, had similar results with a non-significant treatment effect of 3.8 mm Hg (P=0.562, 95% CI 10.3 to 17.9) and a non-significant period effect of 11.4 mm Hg (P=0.100, 95% CI 2.7 to 25.5). On the contrary, in cerebral oxygen saturation there was a statistically significant treatment effect of 9.5% (P=0.007, 95% CI 4.114.9) with a non-significant period effect of 3.4% (P=0.204, 95% CI 2.0 to 8.8).
Tissue oxygenation recorded from the tissue oxygen-monitoring systems correlated well (r=0.92, Fig. 3). The difference between the polarographic and fluoroscopic systems (accuracy, bias) averaged 6.8 (15.0) mm Hg. The precision (2 SDs of the difference between the systems) was 30 mm Hg (Fig. 4). The Foxy system proved difficult to use because it was light sensitive, subject to substantial motion artifact, and lacked a small-calibre temperature sensor.
|
|
| Discussion |
|---|
|
|
|---|
It is well established that mild hypercapnia improves peripheral perfusion and increases tissue PO2. Our main question was whether peripheral vasodilation is a direct effect of the CO2 or a secondary result of increased cardiac output and related central autonomic homeostatic responses. We found that mild hypercapnia did not increase s.c. tissue oxygenation when systemic blood flow and mean arterial pressure remained constant during cardiopulmonary bypass. Increased tissue oxygenation during mild hypercapnia thus most likely results from a hyperdynamic circulatory response and shifting oxyhaemoglobin dissociation curve rather than direct peripheral vasodilation.
Even mild systemic hypercapnia provokes sympatho-excitation15 which increases cardiac output, enhances systemic vascular conductance,18 19 andbecause of active venoconstrictionincreases venous return.19 20 The increase in capacitance vessel tone is mediated via peripheral and central chemoreceptors and does not necessarily parallel changes in overall vascular resistance18and, as our results demonstrate, does not increase s.c. perfusion. Venoconstriction nonetheless increases venous return and is, thus, invariably associated with a sustained increase in cardiac output,14 18 19 31 an effect that dominates a small direct depressant effect of hypercapnic acidosis on the isolated myocardium.32 33
Hypercapnia increases both sympathetic and cardiac vagus nerve activity in anaesthetized dogs. Such co-activation of vagus and sympathetic systems, which can be initiated reflexively or by action on higher centres, has been shown to be of distinct physiological benefit in controlling reactions that relate cardiac function to body need. As the sympathetic and parasympathetic systems are co-activated during systemic hypercapnia, blood pressure and heart rate response depends on the functional balance between these two systems. We were unable to evaluate heart rate during bypass, but have previously shown that both mean arterial blood pressure and heart rate remained essentially unchanged during hypercapnia even though cardiac output increases 25%.11
Recent evidence indicates that both the sympathetic and parasympathetic nervous systems contribute to regulation of peripheral microcirculation.34 The presence of parasympathetic-related, high-frequency microvascular oscillations has been recently documented in adrenergically rich peripheral regionseven during nonpulsatile perfusion34and has important implications for local flow homeostasis. Interestingly, nonpulsatile flow in the systemic circulation, as during cardiopulmonary bypass, progressively increases sympathetic nerve activity,35 leading to arterial vasoconstriction and increased SVR.35 This increase in sympathetic nerve activity is attributable to suppression of baroreflex-mediated inhibition of the vasomotor centre and is associated with local parasympathetic activation34 in an effort to preserve tissue perfusion. Mild hypercapnia during constant systemic blood flow did not increase tissue oxygenation; however, we cannot exclude the possibility of a differential action of CO2 in regional vascular resistance,36 which could have diverted systemic perfusion to areas other than s.c. tissues.37 Our results provided data only from the s.c. tissue. If any perfusion improvements occurred because of a change in vascular resistance, we missed them because of a lack of monitoring. Potential redistribution cannot be excluded either.
In striking contrast to the unchanged s.c. oxygenation during constant systemic blood flow, mild hypercapnia markedly improved cerebral regional oxygen saturation. And the magnitude of the increase was similar to that we have previously observed in patients undergoing elective non-cardiac surgery.31 This finding is consistent with the fact that sympathetic control of the cerebral vessels is weaker than of other vascular beds,38 and the contractile state of the cerebrovascular smooth muscle appears to depend mainly on local metabolic factors including the partial pressure of CO2. Thus, the vasodilator effect of CO2 is particularly marked in the cerebral circulation where a CO2 concentration of 710% nearly doubles cerebral blood flow (CBF) in humans,39 while mild hypercapnia (
50 mm Hg) impairs autoregulation of CBF and is associated with an overall increase in cerebral oxygenation.11 40 A similar cerebrovascular response during cardiopulmonary bypass leads to an increase in CBF41 that is associated with a reduction in cerebral oxygen consumption.42 On the other hand, peripheral vasomotor tone during hypercapnia is essentially the result of a balance between the direct effects of CO2 and the level of sympathetic activity.38 Of course, because neither cerebral nor other vital organ blood flows were monitored in the study, it was not possible to exclude any role of redistribution of blood flow between the organ systems.
The mechanism by which CO2 exerts its direct effects on the cerebral vasculature seems to involve nitric oxide (NO), ATP-sensitive potassium channels, and cyclooxygenase-dependent pathways. The CO2NO axis is considered a cardinal pathway for CBF regulation in humans. Thus, although ATP-sensitive and Ca2+-activated potassium channels are also major systems that respond to hypercapnic acidosis, their response is incomplete in the absence of NO donors. In both animals43 and humans,44 hypercapnic vasodilatation is mediated by inhibition of NO synthasethe enzyme responsible for NO synthesis. It is probable that the vasodilation to hypercapnic acidosis is mediated either by increased synthesis of NO or increased sensitivity to NO.
A secondary goal of our study was to compare a new fluoroscopic optode based oxygen-monitoring system (Foxy) with the current standard, polarographic electrode system (Licox). The fluorescence optode system is known to be more accurate in lower tissue oxygen environments than the polarographic one, because it does not consume oxygen. However, increased bias and impaired linearity of the relationship between the oxygen-monitoring systems when PsqO2 was above 150 mm Hg indicates reduced accuracy of the fluorescence optode at high tissue oxygenation. Other drawbacks of the fluorescence method include substantial light sensitivity, sensitivity to movement and the lack of a small-calibre temperature probe. These limitations make its use in the clinical setting impractical.
Another potential limitation of our study was that we allowed only a 30 min equilibration period at each designated arterial PCO2 concentration. (A period of 30 min was chosen because the bypass period sometimes lasted only an hour.) However, 2030 min is sufficient to obtain stable tissue oxygen values with tonometric systems that accommodate tissue oxygen probes. A further potential limitation of our study was that three of our 10 patients were obese and obesity is known to influence tissue oxygenation.8 However, because of our crossover design, the effect on tissue oxygenation should have been the same at both concentrations of CO2. And lastly, we did not provide a positive control by increasing the bypass pump outflow during the study, which would have simulated increased cardiac output. Although we initially thought about putting this in our protocol, because we were working with patients as study subjects, we decided that it would be risky to include it. The clinical importance of the relationship between the bypass pump outflow and weaning the patient off the pump is just too fragile.
In summary, mild hypercapniawhich normally markedly increases s.c. tissue oxygenationfailed to do so during cardiopulmonary bypass when the bypass pump controlled systemic blood flow. In contrast, cerebral oxygen saturation increased as usual. The increase in s.c. oxygen partial pressure that normally accompanies mild hypercapnia thus results largely from increased cardiac output and compensatory autonomic circulatory responses. Potential contribution of shifting of oxyhaemoglobin dissociation curve and decreased oxygen consumption should also be considered.
| Acknowledgments |
|---|
Supported by NIH Grants GM 61655 and DE 14879-01A1 (Bethesda, MD), the Joseph Drown Foundation (Los Angeles, CA, USA), the Oesterreichische Nationalbank Jubiläumsfonds, the Gheens Foundation (Louisville, KY, USA), Ocean Optics Inc. Spectroscopy Educational Grant program, and the Commonwealth of Kentucky Research Challenge Trust Fund (Louisville, KY, USA). O.A. is the recipient of a Research Training Grant from the Foundation for Anesthesia Education and Research. Somanetics, Inc. (Troy, MI, USA) donated the INVOS cerebral oximeter probes. Tyco-Mallinckrodt Anesthesiology Products, Inc. (St Louis, MO, USA) donated the thermocouples we used. We thank Gilbert Haugh, MS, for help with the statistical analysis, Nancy Alsip, PhD, for editorial assistance (both from the Outcomes Research Institute, University of Louisville), and Dr Harvey Edmonds for suggestions about cerebral oxygenation and perfusion monitoring (Department of Anesthesiology, University of Louisville).
| References |
|---|
|
|
|---|
1 Haley R. Infection-control strategies save $250,000 annually. Hospitals 1985; 59:635[Web of Science][Medline]
2 Burke JF. The effective period of preventive antibiotic action in experimental incisions and dermal lesions. Surgery 1961; 50:1618[Web of Science][Medline]
3 Knighton DR, Fiegel VD, Halverson T, et al. Oxygen as an antibiotic. The effect of inspired oxygen on bacterial clearance. Arch Surg 1990; 125:97100
4 Greif R, Akça O, Horn E-P, et al. Supplemental perioperative oxygen to reduce the incidence of surgical wound infection. N Engl J Med 2000; 342:1617
5 Gosain A, Rabin J, Reymond JP, et al. Tissue oxygen tension and other indicators of blood loss or organ perfusion during graded hemorrhage. Surgery 1991; 109:52332[Web of Science][Medline]
6 Shoemaker W, Appel P, Kram H. Oxygen transport measurements to evaluate tissue perfusion and titrate therapy: dobutamine and dopamine effects. Crit Care Med 1991; 19:67288[Web of Science][Medline]
7 Ikeda T, Tayefeh F, Sessler DI, et al. Local radiant heating increases subcutaneous oxygen tension. Am J Surg 1998; 175:337[CrossRef][Web of Science][Medline]
8 Kabon B, Nagele A, Reddy D, et al. Obesity decreases perioperative tissue oxygenation. Anesthesiology 2004; 100:27480[CrossRef][Web of Science][Medline]
9 Arkilic C, Tagushi A, Ahluwalia A, Sessler D, Kurz A. The effect of supplemental fluid administration on tissue perfusion and tissue oxygen pressure. Anesthesiology 2000; 93:A207
10 Akça O, Melischek M, Scheck T, et al. Postoperative pain and subcutaneous oxygen tension. Lancet 1999; 354:412[CrossRef][Medline]
11 Akça O, Doufas A, Morioka N, et al. Hypercapnia improves tissue oxygenation. Anesthesiology 2002; 97:8016[CrossRef][Web of Science][Medline]
12 Hopf HW, Hunt TK, West JM, et al. Wound tissue oxygen tension predicts the risk of wound infection in surgical patients. Arch Surg 1997; 132:9971004 discussion 1005
13 Gaddis ML, MacAnespie CL, Rothe CF. Vascular capacitance responses to hypercapnia of the vascularly isolated head. Am J Physiol 1986; 251:H16470
14 Shoemaker JK, Vovk A, Cunningham DA. Peripheral chemoreceptor contributions to sympathetic and cardiovascular responses during hypercapnia. Can J Physiol Pharmacol 2002; 80:113644[CrossRef][Web of Science][Medline]
15 Low JM, Gin T, Lee TW, Fung K. Effect of respiratory acidosis and alkalosis on plasma catecholamine concentrations in anaesthetized man. Clin Sci (Lond) 1993; 84:6972[Medline]
16 Somers VK, Mark AL, Zavala DC, Abboud FM. Contrasting effects of hypoxia and hypercapnia on ventilation and sympathetic activity in humans. J Appl Physiol 1989; 67:21016
17 Andersen M and Mouritzen C. Effect of acute respiratory and metabolic acidosis on cardiac output and peripheral resistance. Ann Surg 1966; 63:1618
18 Rothe CF, Maass-Moreno R, Flanagan AD. Effects of hypercapnia and hypoxia on the cardiovascular system: vascular capacitance and aortic chemoreceptors. Am J Physiol 1990; 259:H9329
19 Braunwald E, Ross J Jr, Kahler R, et al. Reflex control of the systemic venous bed. Effects on venous tone of vasoactive drugs, and of baroreceptor and chemoreceptor stimulation. Circ Res 1963; 2:53952[Medline]
20 Tominaga M, Stekiel TA, Bosnjak ZJ, Kampine JP. Contribution of carotid chemoreceptors to mesenteric venoconstriction during acute hypercapnia in rabbits. Am J Physiol 1999; 277:H230510
21 Oikawa S, Hirakawa H, Kusakabe T, Nakashima Y, Hayashida Y. Autonomic cardiovascular responses to hypercapnia in conscious rats: the roles of the chemo- and baroreceptors. Auton Neurosci 2005; 117:10514[CrossRef][Web of Science][Medline]
22 Morii S, Ngai AC, Winn HR. Reactivity of rat pial arterioles and venules to adenosine and carbon dioxide: with detailed description of the closed cranial window technique in rats. J Cereb Blood Flow Metab 1986; 6:3441[Web of Science][Medline]
23 Yamaguchi K, Takasugi T, Fujita H, et al. Endothelial modulation of a pH-dependent pressor response in isolated perfused rabbit lungs. Am J Physiol 1996; 270:H2528
24 Jensen JA, Goodson WH, Hopf HW, Hunt TK. Cigarette smoking decreases tissue oxygen. Arch Surg 1991; 126:11314
25 Gottrup F, Firmin R, Chang N, Goodson WH III, Hunt TK. Continuous direct tissue oxygen tension measurement by a new method using an implantable silastic tonometer and oxygen polarography. Am J Surg 1983; 146:399403[CrossRef][Web of Science][Medline]
26 Torres Filho IP and Intaglietta M. Microvessel PO2 measurements by phosphorescence decay method. Am J Physiol 1993; 265:H14348
27 OOISensors, Fiber Optic Sensors System Operating Instructions Ocean Optics I. 2004; In vol. 20012005. Ocean Optics, Inc, 2004
28 Akca O and Sessler D. Use of cerebral oximetry to detect and manage cerebral desaturation with a rapidly expanding neck hematoma. Acta Anaesthesiol Scand 2002; 46:6078[CrossRef][Web of Science][Medline]
29 Nemoto E, Yonas H, Kassam A. Clinical experience with cerebral oximetry in stroke and cardiac arrest. Crit Care Med 2000; 28:10524[CrossRef][Web of Science][Medline]
30 Hills M and Armitage P. The two-period cross-over clinical trial. Br J Clin Pharmacol 1979; 8:720[Web of Science][Medline]
31 Akca O, Liem E, Suleman M, et al. Effect of intra-operative end-tidal carbon dioxide partial pressure on tissue oxygenation. Anaesthesia 2003; 58:53642[CrossRef][Web of Science][Medline]
32 Kazmaier S, Weyland A, Buhre W, et al. Effects of respiratory alkalosis and acidosis on myocardial blood flow and metabolism in patients with coronary artery disease. Anesthesiology 1998; 89:8317[CrossRef][Web of Science][Medline]
33 Walley KR, Lewis TH, Wood LD. Acute respiratory acidosis decreases left ventricular contractility but increases cardiac output in dogs. Circ Res 1990; 67:62835
34 Podgoreanu MV, Stout RG, El-Moalem HE, Silverman DG. Synchronous rhythmical vasomotion in the human cutaneous microvasculature during nonpulsatile cardiopulmonary bypass. Anesthesiology 2002; 97:111017[CrossRef][Web of Science][Medline]
35 Toda K, Tatsumi E, Taenaka Y, Masuzawa T, Takano H. Impact of systemic depulsation on tissue perfusion and sympathetic nerve activity. Ann Thorac Surg 1996; 62:173742
36 Lioy F, Blinkhorn MT, Garneau C. Regional hemodynamic effects of changes in PaCO2 in the vagotomized, sino-aortic de-afferented rat. J Auton Nerv Syst 1985; 12:30113[CrossRef][Web of Science][Medline]
37 Hoka S, Arimura H, Bosnjak ZJ, Kampine JP. Regional venous outflow, blood volume, and sympathetic nerve activity during hypercapnia and hypoxic hypercapnia. Can J Physiol Pharmacol 1992; 70:10329[Web of Science][Medline]
38 Ainslie PN, Ashmead JC, Ide K, Morgan BJ, Poulin MJ. Differential responses to CO2 and sympathetic stimulation in the cerebral and femoral circulations in humans. J Physiol 2005; 566:61324
39 Kety S and Schmidt C. The effect of altered tensions of carbon dioxide and oxygen on cerebral blood flow and cerebral oxygen consumption of normal young men. J Clin Invest 1948; 27:48492
40 Rostrup E, Law I, Pott F, Ide K, Knudsen GM. Cerebral hemodynamics measured with simultaneous PET and near-infrared spectroscopy in humans. Brain Res 2002; 954:18393[CrossRef][Web of Science][Medline]
41 Prough DS, Stump DA, Roy RC, et al. Response of cerebral blood flow to changes in carbon dioxide tension during hypothermic cardiopulmonary bypass. Anesthesiology 1986; 64:57681[CrossRef][Web of Science][Medline]
42 Prough DS, Rogers AT, Stump DA, et al. Hypercarbia depresses cerebral oxygen consumption during cardiopulmonary bypass. Stroke 1990; 21:11626
43 Iadecola C, Pellegrino D, Moskowitz M, Lassen N. Nitric oxide synthase inhibition and cerebrovascular regulation. J Cereb Blood Flow Metab 1994; 14:17592[Web of Science][Medline]
44 Schmetterer L, Findl O, Strenn K, et al. Role of NO in the O2 and CO2 responsiveness of cerebral and ocular circulation in humans. Am J Physiol 1997; 273:R200512
![]()