British Journal of Anaesthesia, 2001, Vol. 86, No. 1 141-142
© 2001 The Board of Management and Trustees of the British Journal of Anaesthesia
Paradoxical changes in bispectral index during nitrous oxide administration
Department of Anesthesiology, Postgraduate Institute of Medical Education and Research, Chandigarh, India 160012
Accepted for publication: July 18, 2000
| Abstract |
|---|
|
|
|---|
I report two patients undergoing open heart surgery, with nitrous oxide and isoflurane anaesthesia, for whom bispectral index (BIS) monitoring showed high BIS values with nitrous oxide and isoflurane anaesthesia. The BIS decreased immediately after nitrous oxide was stopped and increased again after nitrous oxide was restarted.
Br J Anaesth 2001; 86: 1412
Key words: monitoring, bispectral index, depth of anaesthesia
| Introduction |
|---|
|
|
|---|
The bispectral index (BIS) is a variable, derived from the EEG, that has been reported to have the ability to measure the hypnotic component of the anaesthetic state and has been shown to correlate well with the degree of sedation using several general anaesthetics.1 2 I have been using BIS monitoring (Aspect A-1000 EEG monitor; Aspect Medical System Inc., MA, USA) to titrate anaesthesia in cardiac surgery patients for the last 3 years. During this period, I found two patients (out of more than 300) in whom increasing the concentration of potent inhalational anaesthetic was associated with no change in BIS. Instead, the BIS decreased upon discontinuation of nitrous oxide and increased again after the reintroduction of nitrous oxide.
| Case reports |
|---|
|
|
|---|
Case 1
A 35 yr old woman was scheduled for mitral valve replacement for mitral regurgitation. Anaesthesia was induced with morphine 0.2 mg kg1 and midazolam 0.1 mg kg1 and a titrated sleep dose of thiopental. The BIS was monitored using silver chloride electrodes applied to the forehead, with one on each malar bone, one at the centre of the forehead and a ground electrode on one side of the centre electrode. Tracheal intubation was facilitated with the use of vecuronium 0.1 mg kg1 and the lungs were mechanically ventilated. Anaesthesia was maintained with 66% nitrous oxide in oxygen and with isoflurane. Ten minutes after tracheal intubation, the BIS increased to >90. It did not decrease even with increasing doses of isoflurane (inspired concentrations of
3% isoflurane). At this point the patients arterial pressure began to decrease. Nitrous oxide and isoflurane were stopped and the patient was given 100% oxygen. Two minutes after the nitrous oxide was discontinued, the BIS started to decrease; it fell to <30 after another 5 min. After a further 5 min, the systemic arterial pressure had improved. Nitrous oxide was then reintroduced, whereupon the BIS increased; the BIS fell again after the nitrous oxide was stopped. BIS monitoring was abandoned and anaesthesia controlled using clinical judgement.
Case 2
A 60 yr old man was scheduled for coronary revascularization. Anaesthesia was induced and maintained as in case 1. In spite of adequate concentrations of isoflurane (23% inspired concentrations), the BIS remained at >90. Nitrous oxide was stopped and the BIS decreased to <50 within 5 min. Reintroduction of nitrous oxide again increased the BIS.
| Discussion |
|---|
|
|
|---|
The ability of nitrous oxide to reduce the requirement for other anaesthetic agents has been well described.3 4 Yamamura and colleagues5 studied the effect of nitrous oxide on the EEG of healthy volunteers and showed an increase in high ß-range activity after nitrous oxide administration. Rampil and colleagues6 found that inhalation of nitrous oxide was associated with an increase in high-frequency and
-range activity of the EEG but this did not alter BIS. A reduction in nitrous oxide concentration was associated with a sudden increase in high-amplitude
activity of the EEG along with a decrease in BIS.6 Nitrous oxide in combination with more potent anaesthetics has been studied previously. Porkkala and colleagues7 have shown that nitrous oxide antagonizes the depressant effects of isoflurane on the EEG. Sebel and colleagues8 found that the BIS was higher when isoflurane was used in combination with nitrous oxide than when it was used without nitrous oxide. Analogous effects were observed in a study by Glass and colleagues, who showed that the addition of nitrous oxide to propofol increased the BIS at which patients failed to respond to verbal commands.2 Barr and colleagues found no effect of nitrous oxide on the BIS when given alone or added to isoflurane.9 These conflicting reports of the effect of nitrous oxide on the BIS show that the effects of nitrous oxide are not consistent, reflecting a potential inadequacy of the algorithm used for calculating the BIS from the EEG.
The algorithm that computes the BIS evaluates predominantly three features of the EEG: the ratio of very high ß-range activity to high
plus low ß activity (relative ß ratio), very high ß-range phase relationships and burst suppression phenomena. These features are used sequentially by the algorithm as sedation and anaesthesia increases, with relative ß ratio being the most influential feature during light sedation. As no single pattern of EEG reflects depth of anaesthesia correctly, a number of features are used to calculate the BIS. The paradoxical changes in the BIS of our patients may have resulted from an alteration in the relative ß ratio with nitrous oxide or withdrawal of nitrous oxide. These findings question the utility of monitoring BIS in controlling the depth of anaesthesia in patients receiving nitrous oxide in addition to other anaesthetic agents. Other neurophysiological measures of depth of anaesthesia may be prone to similar problems with nitrous oxide.7 10 11
| References |
|---|
|
|
|---|
1 Lui J, Singh H, White PF. Electroencephalographic bispectral index correlates with intraoperative recall and depth of propofol induced sedation. Anesth Analg 1997; 84: 1859[Abstract]
2 Glass PS, Bloom M, Kearse L, Rosow C, Sebel P, Manberg P. Bispectral analysis measures sedation and memory effects of propofol, midazolam, isoflurane, and alfentanil. Anesthesiology 1997; 86: 83647[Web of Science][Medline]
3 Stevens WC, Dolan WM, Gibbons RT et al. Minimum alveolar concentration (MAC) of isoflurane with and without nitrous oxide in patients of various ages. Anesthesiology 1975; 42: 197200[Web of Science][Medline]
4 Ghouri AF, White PF. Effect of fentanyl and nitrous oxide on the desflurane anesthetic requirement. Anesth Analg 1991; 72: 37781
5 Yamamura T, Fukuda M, Takeya H, Goto Y, Furukoma K. Fast oscillatory EEG activity induced by analgesic concentration of nitrous oxide in man. Anesth Analg 1981; 60: 2838
6 Rampil IJ, Kim JS, Lenhardet R, Negishi C, Sessler DI. Bispectral EEG index during nitrous oxide administration. Anesthesiology 1998; 89: 6717[Web of Science][Medline]
7 Porkkala T, Jantti V, Kaukinen S, Hakkinen V. Nitrous oxide has different effects on the EEG and somatosensory evoked potentials during isoflurane anaesthesia in patients. Acta Anaesthesiol Scand 1997; 41: 497501[Web of Science][Medline]
8 Sebel P, Lang E, Rampil L et al. A multicenter study of bispectral encephalogram analysis for monitoring anesthetic effect. Anesth Analg 1997; 84: 8919[Abstract]
9 Barr G, Jakobsson JG, Owall A, Anderson RE. Nitrous oxide does not alter bispectral index: study with nitrous oxide as sole agent and as an adjunct to i.v. anaesthesia. Br J Anaesth 1999; 82: 82730
10 Newton D, Thornton C, Creagh-Barry P, Dore CJ. Early cortical auditory evoked response in anaesthesia: comparison of the effects of nitrous oxide and isoflurane. Br J Anaesth 1989; 62: 615
11 Thornton C, Creagh-Barry P, Jordan C et al. Somatosensory and auditory evoked responses recorded simultaneously: different effects of nitrous oxide and isoflurane. Br J Anaesth 1992; 68: 50814
![]()
CiteULike
Connotea
Del.icio.us What's this?
This article has been cited by other articles:
![]() |
R. Zanner, S. Pilge, E. F. Kochs, M. Kreuzer, and G. Schneider Time delay of electroencephalogram index calculation: analysis of cerebral state, bispectral, and Narcotrend indices using perioperatively recorded electroencephalographic signals Br. J. Anaesth., September 1, 2009; 103(3): 394 - 399. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Bennett, L. J. Voss, J. P. M. Barnard, and J. W. Sleigh Practical Use of the Raw Electroencephalogram Waveform During General Anesthesia: The Art and Science Anesth. Analg., August 1, 2009; 109(2): 539 - 550. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. A. Wong, R. J. Fragen, P. Fitzgerald, and R. J. McCarthy A comparison of the SNAP IITM and BIS XPTM indices during sevoflurane and nitrous oxide anaesthesia at 1 and 1.5 MAC and at awakening Br. J. Anaesth., August 1, 2006; 97(2): 181 - 186. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. S. Park, E. J. Hur, K. W. Han, H. Y. Kil, and T. H. Han Bispectral index does not correlate with observer assessment of alertness and sedation scores during 0.5% bupivacaine epidural anesthesia with nitrous oxide sedation. Anesth. Analg., August 1, 2006; 103(2): 385 - 9, table of contents. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Zaballos, R. Villazala, S. Agusti, E. Garcia de Lucas, E. Bastida, and J. Navia, MD Can anesthesiologists trust anesthesia depth monitors? Anesth. Analg., August 1, 2006; 103(2): 510 - 510. [Full Text] [PDF] |
||||
![]() |
A. A. Dahaba Different Conditions That Could Result in the Bispectral Index Indicating an Incorrect Hypnotic State Anesth. Analg., September 1, 2005; 101(3): 765 - 773. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. S. Schwab, M. D. Seeberger, E. I Eger II, C. H. Kindler, and M. Filipovic Sevoflurane Decreases Bispectral Index Values More than Does Halothane at Equal MAC Multiples Anesth. Analg., December 1, 2004; 99(6): 1723 - 1727. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. W. Sleigh and J. P. M. Barnard Editorial I: Entropy is blind to nitrous oxide. Can we see why? Br. J. Anaesth., February 1, 2004; 92(2): 159 - 161. [Full Text] [PDF] |
||||
![]() |
T. M. Hemmerling and J. D. Fortier Falsely Increased Bispectral Index Values in a Series of Patients Undergoing Cardiac Surgery Using Forced-Air-Warming Therapy of the Head Anesth. Analg., August 1, 2002; 95(2): 322 - 323. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||