Performance of a minimally invasive cardiac output monitoring system (Flotrac/Vigileo)
Rostock and Regensburg, Germany
* E-mail: thomas.scheeren{at}uni-rostock.de
Editor—We read with great interest the study by Compton and colleagues1 comparing minimal invasive measurements of cardiac output obtained with two devices, the PiCCOTM and the FloTracTM/VigileoTM system. The latter has been introduced into clinical practice in 2005, and early studies comparing this device with established methods to determine cardiac output have shown limited agreement only. As a consequence, the algorithm for calculating stroke volume, heart rate, and hence cardiac output derived from the arterial pressure waveform (pulse contour analysis) has been modified and several software updates have been released. Compton and colleagues claim that they used the new, improved software version named 1.10, which was released in November 2006, and not in spring 2006 as stated. On the other hand, patients were included between July 2006 and May 2007. Thus, the question arises which software version was really used or how many of their 25 patients were studied using the latest software update. This question is of importance as the updated software resulted in an improved percentage error of measurement values compared with standard techniques (Table 1, which updates Table 4 from the original article). One of the software updates refers to the detection of arrhythmias, eliminating one of the study limitations raised by the authors. A second point refers to the data presentation. In the Method section, the authors claim that both the transpulmonary thermodilution-derived cardiac output and the pulse contour-derived continuous cardiac output values of the PiCCOTM system were used for comparison with the new device. However, in the Results section, no comparison between both pulse contour-derived methods is presented. Instead, PiCCOTM data obtained from transpulmonary thermodilution and pulse contour analysis were pooled. It would be interesting to see the agreement between data obtained from transpulmonary thermodilution and pulse contour analysis obtained with the PiCCOTM. Finally, we wonder if transpulmonary thermodilution is a good choice for obtaining reference values for cardiac output, particularly in patients with acute lung injury, as the femoral thermodilution curve is likely affected by pulmonary oedema. In the Compton study, 19 of the 25 patients suffered from acute respiratory failure. The authors state that re-calibration of the PiCCOTM device was done at least every 24 h. A recent paper has shown that the accuracy of obtained values exceeds the accepted percentage error of 30%2 as early as 60 min after re-calibration.3 Hence, a frequent recalibration of the PiCCOTM device (every 20–30 min) is required to obtain reliable results, particularly in the presence of vasopressor therapy.4 In their discussion, the authors state that arrhythmia was a limitation, but provide no information regarding its incidence. In this respect, it is noteworthy that the key feature of the latest software version (v1.10) is the improved performance during arrhythmia.5 6
|
In conclusion, we challenge the conclusions presented by the authors concerning the (lack of) accuracy of the FloTracTM/VigileoTM system since they obviously did not consistently use the improved new software version throughout their study as stated. Furthermore, the inappropriate choice and use of the reference method precludes expecting a close agreement between the two techniques.
Berlin, Germany
* E-mail: friederike.compton{at}charite.de
Editor—We would like to thank Drs Scheeren and Wiesenack for their interest in our work. It is correct that two versions of the FloTracTM/VigileoTM software were used in the study. Despite our timely inquiries, we were only informed after publication that six of our patients were not measured using the latest version (1.10). According to Edwards Lifesciences, the software was updated sometime in August 2006. Omitting those six patients (57 measurement pairs) and recalculating the data, however, yield comparable lack of success, as can be seen below. Our results and thus our conclusions remain unchanged.
Agreement between arterial pressures: MAPfem/MAPrad bias 3 mm Hg, limits of agreement (LOA) –12 to 18 mm Hg.
Comparison of radial and femoral artery pressures: systolic arterial pressures P=0.2517, mean arterial pressures P<0.001, and diastolic arterial pressures P<0.001.
Agreement between cardiac indexes: reference-CI/AP-CI bias 0.69 litre min–1 m–2, LOA –1.22 to 2.62 litre min–1 m–2; TD-CI/AP-CI bias 0.77 litre min–1 m–2, LOA –1.11 to 2.65 litre min–1 m–2.
As suggested by Scheeren and Wiesenack, we also calculated agreement between PiCCO pulse contour values only (PC-CI) and the corresponding FloTrac/Vigileo measurements (AP-CI): PC-CI/AP-CI bias 0.72 litre min–1 m–2, LOA –1.23 to 2.66 litre min–1 m–2.
Re-analysis of subgroups and
CI values did not yield significant changes either.
We regret any confusion concerning the measurements in our study.
References
1 Compton FD, Zukunft B, Hoffmann C, Zidek W, Schaefer JH. Performance of a minimally invasive uncalibrated cardiac output monitoring system (Flotrac/Vigileo) in haemodynamically unstable patients. Br J Anaesth (2008) 100:451–6.
2 Critchley LA, Critchley JA. A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques. J Clin Monit Comput (1999) 15:85–91.[CrossRef][Web of Science][Medline]
3 Hamzaoui O, Monnet X, Richard C, Osman D, Chemla D, Teboul JL. Effects of changes in vascular tone on the agreement between pulse contour and transpulmonary thermodilution cardiac output measurements within an up to 6-hour calibration-free period. Crit Care Med (2008) 36:434–40.[Web of Science][Medline]
4 Bein B, Meybohm P, Cavus E, et al. The reliability of pulse contour-derived cardiac output during hemorrhage and after vasopressor administration. Anesth Analg (2007) 105:107–13.
5 Prasser C, Trabold B, Schwab A, Keyl C, Ziegler S, Wiesenack C. Evaluation of an improved algorithm for arterial pressure-based cardiac output assessment without external calibration. Intensive Care Med (2007) 33:2223–5.[CrossRef][Web of Science][Medline]
6 Mayer J, Boldt J, Wolf MW, Lang J, Suttner S. Cardiac output derived from arterial pressure waveform analysis in patients undergoing cardiac surgery: validity of a second generation device. Anesth Analg (2008) 106:867–72.
7 de Waal EE, Kalkman CJ, Rex S, Buhre WF. Validation of a new arterial pulse contour-based cardiac output device. Crit Care Med (2007) 35:1904–9.[CrossRef][Web of Science][Medline]
8 Mayer J, Boldt J, Schollhorn T, Rohm KD, Mengistu AM, Suttner S. Semi-invasive monitoring of cardiac output by a new device using arterial pressure waveform analysis: a comparison with intermittent pulmonary artery thermodilution in patients undergoing cardiac surgery. Br J Anaesth (2007) 98:176–82.
9 Opdam HI, Wan L, Bellomo R. A pilot assessment of the FloTrac(TM) cardiac output monitoring system. Intensive Care Med (2007) 33:344–9.[CrossRef][Web of Science][Medline]
10 Sander M, Spies CD, Grubitzsch H, Foer A, Muller M, von Heymann C. Comparison of uncalibrated arterial waveform analysis in cardiac surgery patients with thermodilution cardiac output measurements. Crit Care (2006) 10:R164.[CrossRef][Medline]
11 Manecke GR Jr, Auger WR. Cardiac output determination from the arterial pressure wave: clinical testing of a novel algorithm that does not require calibration. J Cardiothorac Vasc Anesth (2007) 21:3–7.[CrossRef][Web of Science][Medline]
12 Prasser C, Bele S, Keyl C, et al. Evaluation of a new arterial pressure-based cardiac output device requiring no external calibration. BMC Anesthesiol (2007) 7:9.[CrossRef][Medline]
13 McGee WT, Horswell JL, Calderon J, et al. Validation of a continuous, arterial pressure-based cardiac output measurement: a multicenter, prospective clinical trial. Crit Care (2007) 11.
14 Breukers RM, Sepehrkhouy S, Spiegelenberg SR, Groeneveld AB. Cardiac output measured by a new arterial pressure waveform analysis method without calibration compared with thermodilution after cardiac surgery. J Cardiothorac Vasc Anesth (2007) 21:632–5.[Web of Science][Medline]
15 Sakka SG, Kozieras J, Thuemer O, van Hout N. Measurement of cardiac output: a comparison between transpulmonary thermodilution and uncalibrated pulse contour analysis. Br J Anaesth (2007) 99:337–42.
16 Lorsomradee S, Lorsomradee S, Cromheecke S, De Hert SG. Uncalibrated arterial pulse contour analysis versus continuous thermodilution technique: effects of alterations in arterial waveform. J Cardiothorac Vasc Anesth (2007) 21:636–43.[Web of Science][Medline]
17 Button D, Weibel L, Reuthebuch O, Genoni M, Zollinger A, Hofer CK. Clinical evaluation of the FloTrac/VigileoTM system and two established continuous cardiac output monitoring devices in patients undergoing cardiac surgery. Br J Anaesth (2007) 99:329–36.
18 Cannesson M, Attof Y, Rosamel P, Joseph P, Bastien O, Lehot JJ. Comparison of FloTracTM cardiac output monitoring system in patients undergoing coronary artery bypass grafting with pulmonary artery cardiac output measurements. Eur J Anaesthesiol (2007) 24:832–9.[Web of Science][Medline]
![]()