BJA Advance Access originally published online on November 16, 2007
British Journal of Anaesthesia 2008 100(1):88-94; doi:10.1093/bja/aem320
Electrical velocimetry for measuring cardiac output in children with congenital heart disease


1 Department of Pediatric Cardiology and Intensive Care Medicine
2 Department of Anesthesiology, Medical School Hannover, OE 6730, Carl-Neuberg-Str. 1, 30623 Hannover, Germany
* Corresponding author. E-mail: norozi.kambiz{at}mh-hannover.de
Accepted for publication August 27, 2007.
| Abstract |
|---|
|
|
|---|
Background: The purpose of this study was to evaluate the agreement of cardiac output measurements obtained by electrical velocimetry (COEV) and those that derived from the direct Fick-oxygen principle (COF) in infants and children with congenital heart defects.
Methods: Simultaneous measurements of COEV and COF were compared in 32 paediatric patients, aged 11 days to 17.8 yr, undergoing diagnostic right and left heart catheterization. For non-invasive measurements of cardiac output by electrical velocimetry, which is a variation of impedance cardiography, standard surface electrodes were applied to the left side of the neck and the left side of the thorax at the level of the xiphoid process. Cardiac output determined using direct Fick-oxygen principle was calculated by direct measurement of oxygen consumption (VO2) and invasive determination of the arterio-venous oxygen content difference.
Results: An excellent correlation (r=0.97) was found between COEV and COF (P<0.001). The slope of the regression equation [0.96 (SD 0.04)] was not significantly different from the line of identity. The bias between the two methods (COEV–COF) was 0.01 litre min–1 and the limits of agreement, defined as the bias (2 SD), were –0.47 and +0.45 litre min–1.
Conclusions: COEV demonstrates acceptable agreement with data derived from COF in infants and children with congenital heart disease. The new technique is simple, completely non-invasive, and provides beat-to-beat estimation of CO.
Keywords: heart, cardiac output; heart, congenital heart disease; measurement techniques, electrical velocimetry; measurement techniques, Fick oxygen method; measurement techniques, impedance cardiography
| Introduction |
|---|
|
|
|---|
Comprehensive evaluation of a patients haemodynamic status should include appraisal of cardiac output. Continuous assessment of cardiac output is useful in the intensive care unit for monitoring patients with heart failure or shock and for the titration of cardiovascular drugs and fluids.1
Most currently available techniques for measuring cardiac output, such as dye dilution, thermodilution, and methods based on the Fick principle, are invasive and require strict adherence to rigid protocols for accurate and reproducible results.2–5 Doppler-echocardiography and carbon dioxide re-breathing are commonly used non-invasive techniques in adults, but require an experienced operator.6 7 There have been few studies of cardiac output measurement in infants and children with congenital heart defects.7 With the exception for pulse contour analysis,8 which is invasive, there is no single technique available for measuring cardiac output continuously and accurately in children with congenital heart defects.
Impedance cardiography is a non-invasive method of obtaining continuous measurements of stroke volume and cardiac output.9 Impedance cardiography technology was developed for NASA by Kubicek and colleagues10 in the 1960s and is based upon the assumption that the human thorax is electrically a non-homogeneous bulk conductor and behaviourally conforms to parallel conduction theory when exposed to a field of alternating current.11 In contrast to the original Kubicek equation10 and its modification by Bernstein in the 1980s,12 the formula incorporated into the new impedance cardiometry monitor AESCULON® relates the maximum rate of change of impedance to peak aortic blood acceleration, and derives the mean aortic blood velocity using a transformation.13 The monitor is based on the premise that the orientation of the erythrocytes in the aorta changes quickly from random to alignment in the direction of blood flow upon opening of the aortic valve. The alignment of erythrocytes during early systole and the following increasingly random orientation produces a pulsatile change in electrical conductivity which is reflected in a decrease in thoracic electrical bioimpedance (TEB) during early systole and an increase later. In contrast to former approaches, a recently reported new method, referred to as electrical velocimetry,13 focuses on the changes in the compartment with the greatest conductivity and the major contributing factor to conductivity changes, the blood in the aorta.
In the current study, conducted in children with congenital heart defects, we compared cardiac output measurements obtained by electrical velocimetry (COEV) with cardiac output determined by the gold-standard direct Fick-oxygen method (COF).
| Methods |
|---|
|
|
|---|
Study population
From 1 July 2005 to 31 January 2006, 32 infants, children, and adolescents undergoing haemodynamic evaluation or for an intervention in the cardiac catheter laboratory and requiring general anaesthesia were considered for inclusion. Twelve female and 20 male patients were studied with a mean age of 3.4 yr (range: 12 days to 17.8 yr; median 0.7 yr) and mean weight of 13.8 kg (range: 2.7–54 kg; median 7.1 kg). The study was approved by the institutional ethics committee and written consent was obtained from the patients guardian, and, depending on age, the patient as well.
Measurements
Anaesthetic management, including the use of cuffed and uncuffed endotracheal tubes, was at the discretion of the paediatric anaesthesiology team. Because oxygen consumption (VO2) was being measured and for patient safety, the tube size chosen had to fulfil the requirement of easy insertion through the cricoid cartilage. Uncuffed tubes were required to seal at up to 20 cm H2O with a subsequent air leak at a peak inspiratory airway pressure of 25 cm H2O. This was assessed by means of placing a stethoscope over the patients mouth. After induction of general anaesthesia and tracheal intubation, a spirometry system (Vmax Encore®, Viasis Healthcare, Hoechberg, Germany) was connected to the anaesthetic circuit (Cato®, Draeger, Lubeck, Germany) to measure oxygen consumption (VO2). Electrical velocimetry was applied before insertion of a catheter introducer sheath into the femoral vein. Haemodynamic measurements were performed approximately 20–40 min after induction of anaesthesia when the patient was haemodynamically stable and before the first angiography.
Non-invasive cardiac output measurement by electrical velocimetry (COEV)
Cardiac output measurement using the AESCULON® monitor (Osypka Medical, Berlin, Germany and San Diego, CA, USA) requires placement of four disposable standard surface ECG electrodes connected to the monitor by a cable. The monitor emits a high-frequency (50 kHz) AC current of a constant magnitude (2 mA, rms) through a pair of electrodes, inducing a current field.14 By means of voltage-sensing electrodes placed within the current field, the quasi-static basal transthoracic impedance Z0 (ohms) and the impedance change
Z(t) are calculated by Ohms Law (Z=U/I) from the demodulated basal transthoracic voltage U (volts) and cardiac-synchronous voltage change
U(t), respectively. By electronic differentiation of
Z(t) [dZ(t) in Fig. 1], the first time-derivative is obtained dZ(t)/dt, from which its peak magnitude, dZ(t)/dtmin, and left ventricular ejection time (flow time) are measured and entered into the Bernstein–Osypka SV equation13 15 to compute COEV. The Bernstein–Osypka SV equation is given in its general form as follows:
|
|
|
where SVB–O is the stroke volume from the Bernstein–Osypka equation (ml); VEPT the volume of electrically participating thoracic tissue in ml (VITBV/
n); VITBV (ml) intrathoracic blood volume obtained from body weight (W, kg) by the relationship aWb, where the coefficient a is
0.25 of patient indexed blood volume in ml kg–1 and exponent b
1;
the index of transthoracic aberrant conduction (dimensionless) and n is an exponent between 1 and 2; dZ(t)/dtmin the ohmic mean acceleration (
s–2); Z0 the quasi-static transthoracic base impedance (
); Upon satisfactory initialization of AESCULON®, COEV was continuously displayed and recorded as an average value over 10 valid cardiac cycles. Averaged COEV data were stored on disc by AESCULON® over the period in which VO2 and invasively derived arterio-venous oxygen content measurements were obtained.
Invasive cardiac output by the direct Fick-oxygen method (COF)
To determine COF, blood samples were obtained from the caval veins, right atrium, or pulmonary trunk for mixed venous oxygen saturation (SvO2) and from the aorta or femoral artery for systemic arterial oxygen saturation (SaO2). Samples were collected during the time interval over which COEV was calculated and recorded by AESCULON®, and were analysed by a blood gas analyser (Hemoximeter OSM 3, Radiometer, Copenhagen, Denmark). The Fick principle for measuring CO was determined by dividing the VO2 by the arterio-venous oxygen content difference via the following equation:
|
|
Haemoglobin concentration (Hb) was obtained from the first blood sample, which was taken during heart catheterization. VO2 was determined by spirometry. We used the Vmax Encore Spirometry System according to the manufacturers guidelines, measuring VO2 continuously during a period of
20 min, thus encompassing the time blood samples were taken. We used the mean VO2 from all measurements during the period of blood sampling for calculation of COF. At the beginning and end of blood samples for COF measurement, an event mark was set into the impedance recording to define the corresponding COEV. At the completion of the catheterization procedure, mean COF was compared with the corresponding mean COEV.
Statistical analysis
All results were analysed using GraphPad Prism 4 software (GraphPad Software, Inc., San Diego, CA, USA) on a Windows computer. All results are expressed as mean (SD). Interchangeability or equivalence between COEV and COF was evaluated using two different methods. First, the closeness of association, or correlation, between the two methods, was computed using the Pearson correlation coefficient r2 and applying a linear regression model. The spread of the slope and the ordinate of this relationship were expressed as their standard errors. A P-value of <0.05 was considered statistically significant. Secondly, to assess agreement between COEV and COF, the method of Bland and Altman16 was employed, by computing bias, precision, and limits of agreement.
| Results |
|---|
|
|
|---|
The details of the patients, VO2 index (ml kg–1 m–2), haemoglobin (g dl–1), SaO2 (%), and CHD diagnosis are given in Table 1. In all cases with a VSD, the left-to-right shunt was significant with pulmonary blood flow (Qp) substantially greater than systemic blood flow (Qs), Qp/Qs>2.3.
|
Figure 2 shows a scatter-plot of the data from 32 simultaneously obtained measurements of COEV and COF. An excellent correlation (r2=0.94, P<0.0001) was found between the two techniques. The slope of the regression equation [m=0.96 (SD 0.04)] was not significantly different from unity with b
0. For the subgroup of nine infants with cyanotic heart defects (SaO2<94%), the correlation between COEV and COF was very good (r2=0.80, P=0.007). The lower correlation coefficient of r2=0.80 in this subgroup compared with that in all patients (r2=0.94) is predominantly a statistic effect due to the small number of patients.
|
The results of the Bland and Altman analysis for all patients are shown in Figure 3. The mean difference (bias) between COEV and COF was 0.01 litre min–1 with standard deviation (precision) of 0.23 litre min–1. The upper and lower limits of agreement (±2 SD) were 0.47 and 0.45 litre min–1, respectively.
|
There were no significant correlations between the accuracy of the two CO methods and body weight (r2=0.1, P=0.11) and individual haemoglobin concentration (r=0.002; P=0.38).
| Discussion |
|---|
|
|
|---|
Although infants and children have been studied with impedance cardiography in the past, its use in patients with congenital heart disease has rarely been reported. Using the previous equations, implemented by far less sophisticated devices than AESCULON®, Miles and colleagues17 found a good to very good correlation (r=0.70–0.89) between COF and COICG in young children with a variety of congenital heart diseases. As with our results, correlation and agreement was lower in patients with a ventricular septal defect (VSD). Braden and colleagues18 compared COF with COICG in a group of children with congenital heart disease. Again in this study, correlation was highest (r=0.84) in patients without shunts and was appreciably lower (r=0.70) in patients with a VSD and left-to-right shunt.
To our knowledge, this report is the first study comparing COEV with COF in children with a variety of different heart lesions. The direct Fick-oxygen method is an established, but invasive gold-standard method of measuring cardiac output. Electrical velocimetry is a new technique which has several advantages compared with COF: it is non-invasive, safe, easy to apply, and provides continuous beat-to-beat estimation of cardiac output.
The results from our study suggest that variations of the anatomical position of the great thoracic vessels in congenital heart disease do not substantially affect the accuracy of electrical velocimetry measurements. This is probably due to the fact that in early systole, electrical velocimetry detects the greatest systolic downslope of
Z(t), which is dZ(t)/dtmin. Since it is suggested that dZ(t)/dtmin represents the greatest rate of change of blood resistivity, this value will always determine the magnitude of the stroke volume estimated by electrical velocimetry (SVEV). Since, in the absence of a VSD and left-to-right shunt, the greatest systolic ohmic acceleration is always located in the ascending thoracic aorta and not the pulmonary artery; the actual position of the great vessels within the thorax is probably unimportant. This allows the use of an unfocused transthoracic current field, which interrogates the entire thoracic volume, and especially the intrathoracic blood volume. This is in contrast to Doppler velocimetry, where a precisely focused beam of ultrasound insonating the aortic root is necessary to detect the highest ascending aortic velocities for accurate determination of the systolic velocity integral. The robustness of the electrical velocimetry technique was evident in three patients studied after Norwood and Glenn operations, a circumstance where only one major artery arises from the heart.
Another interesting aspect of our study was the lack of correlation between haemoglobin concentration and the accuracy of electrical velocimetry when compared with the Fick principle determination of cardiac output. Since, by the new theory of electrical velocimetry, dZ(t)/dtmin is dependent on the biphasic orientation of erythrocytes over the cardiac cycle, it seemed plausible that haemoglobin or haematocrit levels might affect the accuracy of electrical velocimetry. Our data show that over a range of haemoglobin concentrations of 8.5–17.3 g dl–1, the agreement between the two methods was unaffected. This is consistent with the results of Quail and colleagues19 and Wallace and colleagues20 who found that the magnitude of impedance cardiography-derived stroke volume and cardiac output was unaffected by haematocrit over a wide range of values. It is also consistent with the observation by Visser and colleagues21 that although the magnitude of
Z(t) is haematocrit-dependent, its maximum systolic upslope [i.e. dZ(t)/dtmin] is not.
The correlation and agreement between cardiac output determined by electrical velocimetry and our gold-standard method was superior to that reported in the studies conducted in adults by Schmidt and colleagues22 and Suttner and colleagues.23 Possible reasons include their choice of reference methods. Schmidt and colleagues22 studied patients undergoing cardiac surgery under anaesthesia before surgery using transoesophageal echocardiography (TOE) as their reference method. They reported excellent correlation (r2=0.86) and limits of agreement of –0.99 to 1.36 litre min–1. TOE-determined stroke volume has several sources for inaccuracy, which include minor errors in aortic valve cross-sectional area measurement24 and errors inherent in determining the systolic velocity integral. In the study by Suttner and colleagues,23 thermodilution was employed as the reference method of cardiac output estimation in patients in intensive care after cardiac surgery. Their patients had the integrity of the chest wall interrupted by median sternotomy during surgery and were inherently more unstable. Despite this, they reported r=0.83 for cardiac index, a bias of 0.01 litre min–1 m–2, and a precision of ±0.57 litre min–1 m–2. Sources of error associated with the thermodilution estimation of cardiac output are well documented,2 especially in ventilated patients, where cardiac output differences over the respiratory cycle can be >20%. The clinical setting, reference method, and results of Suttner and colleagues23 were not significantly different from those of Bernstein and Lemmens.15 They studied a similar, but larger population of patients employing electrical velocimetry from raw data obtained from another proprietary impedance cardiography device. However, despite reference method inaccuracies, results from all these studies substantiate the interchangeability of COEV with their respective gold standards as it is now accepted that limits of agreement between methods of ±30% imply comparability.25
Other concerns regarding the applicability of electrical velocimetry in patients with congenital heart disease include its accuracy in severe aortic stenosis or coarctation, SHONE complex, severe sub-aortic stenosis, severe narrowing of the aortic isthmus, right aortic arch, transposition of the great arteries, and in haemodynamically unstable paediatric patients after corrective surgery. Further studies are needed.
| Conclusions |
|---|
|
|
|---|
In young individuals with congenital heart disease, measurement of COEV by the AESCULON® monitor agreed well with measurement by COF in a steady-state clinical situation. This is true for the range of cardiac output from 0.4 to 4.0 litre min–1. The results were not significantly influenced by the type of heart defect and were stable over the body weight and [Hb] range from 2.7 to 54 kg and 8.5 to 17.3 g dl–1, respectively. Thus, the system seems appropriate for continuous cardiac output measurement, even under the abnormal haemodynamics of infants with a variety of congenital heart lesions.
| Footnotes |
|---|
K.N. and C.B. contributed equally to this paper. | References |
|---|
|
|
|---|
1 Pinsky MR. Why measure cardiac output? Crit Care (2003) 7:114–6.[CrossRef][Web of Science][Medline]
2 Nishikawa T, Dohi S. Errors in the measurement of cardiac output by thermodilution. Can J Anaesth (1993) 40:142–53.[Web of Science][Medline]
3 Dhingra VK, Fenwick JC, Walley KR, Chittock DR, Ronco JJ. Lack of agreement between thermodilution and fick cardiac output in critically ill patients. Chest (2002) 122:990–7.[CrossRef][Web of Science][Medline]
4 Gonzalez J, Delafosse C, Fartoukh M, et al. Comparison of bedside measurement of cardiac output with the thermodilution method and the Fick method in mechanically ventilated patients. Crit Care (2003) 7:171–8.[CrossRef][Web of Science][Medline]
5 Hoeper MM, Maier R, Tongers J, et al. Determination of cardiac output by the Fick method, thermodilution, and acetylene rebreathing in pulmonary hypertension. Am J Respir Crit Care Med (1999) 160:535–41.
6 Beekman RH, Katch V, Marks C, Rocchini AP. Validity of CO2-rebreathing cardiac output during rest and exercise in young adults. Med Sci Sports Exerc (1984) 16:306–10.
7 Wippermann CF, Schranz D, Huth R, Zepp F, Oelert H, Jungst BK. Determination of cardiac output by an angle and diameter independent dual beam Doppler technique in critically ill infants. Br Heart J (1992) 67:180–4.
8 Fakler U, Pauli C, Balling G, et al. Cardiac index monitoring by pulse contour analysis and thermodilution after pediatric cardiac surgery. J Thorac Cardiovasc Surg (2007) 133:224–8.
9 Summers RL, Shoemaker WC, Peacock WF, Ander DS, Coleman TG. Bench to bedside: electrophysiologic and clinical principles of noninvasive hemodynamic monitoring using impedance cardiography. Acad Emerg Med (2003) 10:669–80.[CrossRef][Web of Science][Medline]
10 Kubicek WG, Karnegis JN, Patterson RP, Witsoe DA, Mattson RH. Development and evaluation of an impedance cardiac output system. Aerosp Med (1966) 37:1208–12.[Web of Science][Medline]
11 Visser KR, Lamberts R, Zijlstra WG. Investigation of the parallel conductor model of impedance cardiography by means of exchange transfusion with stroma free haemoglobin solution in the dog. Cardiovasc Res (1987) 21:637–45.[Web of Science][Medline]
12 Bernstein DP. A new stroke volume equation for thoracic electrical bioimpedance: theory and rationale. Crit Care Med (1986) 14:904–9.[Web of Science][Medline]
13 Bernstein DP, Osypka MJ. Apparatus and method for determining an approximation of the stroke volume and the cardiac output of the heart. In: US Patent 6,511,438 B2 (2003).
14 Alexander OW, Huber D, Beck C, et al. Comparison of electrical velocimetry and transpulmonary thermodilution for measuring cardiac output in piglets. Paediatr Anaesth (2007) 17:749–55.[CrossRef][Medline]
15 Bernstein DP, Lemmens HJ. Stroke volume equation for impedance cardiography. Med Biol Eng Comput (2005) 43:443–50.[CrossRef][Web of Science][Medline]
16 Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet (1986) 1:307–10.[CrossRef][Web of Science][Medline]
17 Miles DS, Gotshall RW, Golden JC, Tuuri DT, Beekman RH III, Dillon T. Accuracy of electrical impedance cardiography for measuring cardiac output in children with congenital heart defects. Am J Cardiol (1988) 61:612–6.[CrossRef][Web of Science][Medline]
18 Braden DS, Leatherbury L, Treiber FA, Strong WB. Noninvasive assessment of cardiac output in children using impedance cardiography. Am Heart J (1990) 120:1166–72.[CrossRef][Web of Science][Medline]
19 Quail AW, Traugott FM, Porges WL, White SW. Thoracic resistivity for stroke volume calculation in impedance cardiography. J Appl Physiol (1981) 50:191–5.
20 Wallace AW, Salahieh A, Lawrence A, Spector K, Owens C, Alonso D. Endotracheal cardiac output monitor. Anesthesiology (2000) 92:178–89.[CrossRef][Web of Science][Medline]
21 Visser KR, Lamberts R, Zijlstra WG. Investigation of the origin of the impedance cardiogram by means of exchange transfusion with stroma free haemoglobin solution in the dog. Cardiovasc Res (1990) 24:24–32.
22 Schmidt C, Theilmeier G, Van Aken H, et al. Comparison of electrical velocimetry and transoesophageal Doppler echocardiography for measuring stroke volume and cardiac output. Br J Anaesth (2005) 95:603–10.
23 Suttner S, Schollhorn T, Boldt J, et al. Noninvasive assessment of cardiac output using thoracic electrical bioimpedance in hemodynamically stable and unstable patients after cardiac surgery: a comparison with pulmonary artery thermodilution. Intensive Care Med (2006) 32:2053–8.[CrossRef][Web of Science][Medline]
24 Gray PE, Perrino AC Jr. Hemodynamic-induced changes in aortic valve area: implications for Doppler cardiac output determinations. Anesth Analg (2001) 92:584–9.
25 Critchley LA, Critchley JA. A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques. J Clin Monit Comput (1999) 15:85–91.[CrossRef][Web of Science][Medline]
![]()