British Journal of Anaesthesia, 2004, Vol. 92, No. 1 8-13
© 2004 The Board of Management and Trustees of the British Journal of Anaesthesia
Clinical Investigations |
Does bispectral analysis of the electroencephalogram add anything but complexity?
Department of Anaesthesia, Waikato Hospital, Hamilton, New Zealand
*Corresponding author. E-mail: sleighj@hwl.co.nz
This article is accompanied by Editorial I.
Background. Analysis of the bispectrum of EEG waveforms is a component of the proprietary BIS indexa commonly used commercial monitor of depth of anaesthesia. Does the use of the bispectrum give more information about depth of anaesthesia than the power spectrum?
Methods. We collected and analysed EEG waveforms during induction of general anaesthesia in 39 patients, comparing the changes in bispectral parameter (SynchFastSlow), with an analogous power spectrum-based parameter (PowerFastSlow). Both compare the logarithmic ratio of high frequency components (4047 Hz) with the total (147 Hz). Because the changes in bispectrum are affected by signal amplitude, we also calculated a third parameter (SFSbicoh) from the bicoherence, which is an amplitude-independent statistic.
Results. The SynchFastSlow and PowerFastSlow were correlated (r=0.84) and neither was superior in predicting the awake or anaesthetized state (area under receiver operating characteristic curves = 0.85 vs 0.93). There was no change in the SFSbicoh over the induction period, and it did not correlate with SynchFastSlow (r=0.07).
Conclusions. We could not show that bispectral analysis gave more information than power spectral-based analysis. Most of the changes in the bispectral values result from decreases in the relative high frequency content of the EEG caused by anaesthesia.
Br J Anaesth 2004; 92: 813
![]()
CiteULike
Connotea
Del.icio.us What's this?
This article has been cited by other articles:
![]() |
O. Tirel, E. Wodey, R. Harris, J. Y. Bansard, C. Ecoffey, and L. Senhadji Variation of bispectral index under TIVA with propofol in a paediatric population Br. J. Anaesth., January 1, 2008; 100(1): 82 - 87. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Jeleazcov, J. Schmidt, B. Schmitz, K. Becke, and S. Albrecht EEG variables as measures of arousal during propofol anaesthesia for general surgery in children: rational selection and age dependence Br. J. Anaesth., December 1, 2007; 99(6): 845 - 854. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Jeleazcov, G. Schneider, M. Daunderer, B. Scheller, J. Schuttler, and H. Schwilden The Discriminant Power of Simultaneous Monitoring of Spontaneous Electroencephalogram and Evoked Potentials as a Predictor of Different Clinical States of General Anesthesia Anesth. Analg., October 1, 2006; 103(4): 894 - 901. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Wodey, O. Tirel, J. Y. Bansard, A. Terrier, C. Chanavaz, R. Harris, C. Ecoffey, and L. Senhadji Impact of age on both BIS values and EEG bispectrum during anaesthesia with sevoflurane in children Br. J. Anaesth., June 1, 2005; 94(6): 810 - 820. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Schneider, S. Schoniger, E. Kochs, and J. Sleigh Does bispectral analysis add anything but complexity? BIS sub-components may be superior to BIS for detection of awareness Br. J. Anaesth., October 1, 2004; 93(4): 596 - 597. [Full Text] [PDF] |
||||
![]() |
V. Jantti, S. Alahuhta, J. Barnard, and J. W. Sleigh Spectral entropy--what has it to do with anaesthesia, and the EEG? Br. J. Anaesth., July 1, 2004; 93(1): 150 - 152. [Full Text] [PDF] |
||||
![]() |
L. D. Gugino, L. S. Aglio, and A. Yli-Hankala Monitoring the Electroencephalogram During Bypass Procedures Seminars in Cardiothoracic and Vascular Anesthesia, June 1, 2004; 8(2): 61 - 83. [Abstract] [PDF] |
||||
![]() |
S. Hagihira, M. Takashina, T. Mori, T. Mashimo, J. W. Sleigh, J. Barnard, A. Miller, and D. A. Steyn-Ross Bispectral analysis gives us more information than power spectral-based analysis Br. J. Anaesth., May 1, 2004; 92(5): 772 - 773. [Full Text] [PDF] |
||||
![]() |
D. K. Tempe and L. Satyanarayana Editorial I: Is there any alternative to the Bispectral Index Monitor? Br. J. Anaesth., January 1, 2004; 92(1): 1 - 3. [Full Text] [PDF] |
||||