Skip Navigation

This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow E-Letters: Submit a response to the article
Right arrow Alert me when this article is cited
Right arrow Alert me when E-letters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrow Search for citing articles in:
ISI Web of Science (4)
Right arrowRequest Permissions
Right arrow Disclaimer
Google Scholar
Right arrow Articles by Proquitté, H.
Right arrow Articles by Schmalisch, G.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Proquitté, H.
Right arrow Articles by Schmalisch, G.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?

British Journal of Anaesthesia, 2003, Vol. 91, No. 5 736-738
© 2003 The Board of Management and Trustees of the British Journal of Anaesthesia


Short Communications

Breathing gas perfluorocarbon measurements using an absorber filled with zeolites

H. Proquitté*, M. Rüdiger, R. R. Wauer and G. Schmalisch

Clinic of Neonatology, CCM, Charité Childrens University Hospital, Schumannstr. 20/21, D-10098 Berlin, Germany

Corresponding author. E-mail: hans.proquitte@charite.de

Background. Perfluorocarbon (PFC) has been widely used in the treatment of respiratory diseases; however, PFC content of the breathing gases remains unknown. Therefore, we developed an absorber using PFC selective zeolites for PFC measurement in gases and investigated its accuracy.

Methods. To generate a breathing gas with different PFC contents a heated flask was rinsed with a constant air flow of 4 litre·min–1 and 1, 5, 10, and 20 ml of PFC were infused over 20 min using an infusor. The absorber was placed on an electronic scale and the total PFC volume was calculated from the weight gain.

Results. Steady-state increase in weight was achieved 3.5 min after stopping the infusion. The calculated PFC volume was slightly underestimated but the measuring error did not exceed –1% for PFC less than 1 ml. The measurement error decreased with increasing PFC volume.

Conclusions. This zeolite absorber is an accurate method to quantitatively determine PFC in breathing gases and can be used as a reference method to validate other PFC sensors.

Br J Anaesth 2003; 91: 736–8


Add to CiteULike CiteULike   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us    What's this?




Disclaimer:
Please note that abstracts for content published before 1996 were created through digital scanning and may therefore not exactly replicate the text of the original print issues. All efforts have been made to ensure accuracy, but the Publisher will not be held responsible for any remaining inaccuracies. If you require any further clarification, please contact our Customer Services Department.