British Journal of Anaesthesia, 2002, Vol. 88, No. 2 175-183
© 2002 The Board of Management and Trustees of the British Journal of Anaesthesia
Clinical Investigations |
Model-based administration of inhalation anaesthesia. 4. Applying the system model
Institute for Anaesthesiology, University of Nijmegen, Geert Grooteplein 10, NL-6500 HB Nijmegen, The Netherlands 1Present address: Department of Anaesthesia, Medisch Spectrum Twente, NL-7500 KA Enschede, The Netherlands*Corresponding author
Background. We developed and tested a simple dosing strategy for rapid induction with isoflurane followed by maintenance under minimal-flow conditions, that is 0.5 litre min1 total fresh gas flow (FGF). An end-expired concentration was to be achieved within 5 min in a desired therapeutic window, that is 0.81.1 vol%, and to be maintained within it for at least 30 min.
Methods. With our new model we computed a three-stage regimen using one fixed vaporizer setting: 3 vol% isoflurane in a FGF of 3 and 1.5 litre min1, each for 3 min, and 0.5 litre min1 thereafter. The ratio of nitrous oxide:oxygen was, consecutively, 2:1, 2:1, and 2:3. We evaluated this scheme in 58 adult patients (body mass 74 (SD 13) kg), mostly during eye and ear, nose, and throat surgery.
Results. Measured oxygen (3345 vol%) and nitrous oxide concentrations (6650 vol%) evolved in accordance with those computed. In five patients with a median of body mass 92 kg (range 76126 kg), inspired oxygen concentrations decreased to less than 30 vol%. End-expired isoflurane concentration entered the window after 2 min (range 1.05.67 min) and attained its maximum, that is 0.96 vol% (0.81.2 vol%), after 3.45 min (1.676.33 min). The mean end-expired concentration was in the desired window from 360 min and an average of 72% of individual measurements were within the window from 330 min. The scheme was adapted in six patients (excluded from analysis) because of hypotension.
Conclusion. The regimen is easily remembered, reliable, and lends itself to alternative strategies, but must be guided by the monitoring of gas and vapour concentrations and haemodynamic variables.
Br J Anaesth 2002; 88: 17583
![]()
CiteULike
Connotea
Del.icio.us What's this?
This article has been cited by other articles:
![]() |
J. F.A. Hendrickx, M. K. Dishart, and A. M. De Wolf Isoflurane and Desflurane Uptake During Liver Resection and Transplantation Anesth. Analg., February 1, 2003; 96(2): 356 - 362. [Abstract] [Full Text] [PDF] |
||||