British Journal of Anaesthesia, 2001, Vol. 86, No. 1 12-28
© 2001 The Board of Management and Trustees of the British Journal of Anaesthesia
Model-based administration of inhalation anaesthesia. 1. Developing a system model
1 Institute for Anaesthesiology, University of Nijmegen, Geert Grooteplein 10, 6500 HB Nijmegen, The Netherlands *Corresponding author
This paper is the first of a series of reports on a system model for the administration of inhalation anaesthesia. We present the development and basic testing of the model. It is a multiple-gas model; it covers fresh-gas flow rates from basal to more than total ventilation and includes an actual, not an idealized, circle-absorber breathing system featuring a standing bellows ventilator. Kinetics of nitrogen, oxygen, carbon dioxide, nitrous oxide, inhaled anaesthetic agents and helium are described. Their partial pressures sum to the total pressure. Ventilation and cardiac output are treated as continuous, not cyclical. The model of the breathing system was empirically matched to the chosen one (a GMS absorber and 7850 ventilator (Datex-Ohmeda)). Predictions for the wash-in of isoflurane and the uptake of desflurane and isoflurane agree well with observed data. The results obtained by continuously checking total gas pressures, calculating mass balances and simulating the measurement of alveolar space by the closed-circuit helium dilution method support the mathematical credibility of the model. It thus merits further exploration.
Br J Anaesth 2001; 86: 1228
![]()
CiteULike
Connotea
Del.icio.us What's this?
This article has been cited by other articles:
![]() |
J. F. Belda, M. Soro, R. Badenes, A. Meiser, M. L. Garcia, G. Aguilar, and F. J. Marti The Predictive Performance of a Pharmacokinetic Model for Manually Adjusted Infusion of Liquid Sevofluorane for Use with the Anesthetic-Conserving Device (AnaConDa): A Clinical Study Anesth. Analg., April 1, 2008; 106(4): 1207 - 1214. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. G. C. Lerou and J. Mourisse Applying a physiological model to quantify the delay between changes in end-expired concentrations of sevoflurane and bispectral index Br. J. Anaesth., August 1, 2007; 99(2): 226 - 236. [Abstract] [Full Text] [PDF] |
||||
![]() |
Y. Jiang, J. C. Wain, A. W. Chang, W. M. Zapol, and Z. Xie Differential changes of alveolar gas concentrations during anesthetic induction of a patient with an absent right pulmonary artery. Anesth. Analg., August 1, 2006; 103(2): 312 - 5, table of contents. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. R. Kennedy, R. A. French, and S. Gilles The Effect of a Model-Based Predictive Display on the Control of End-Tidal Sevoflurane Concentrations During Low-Flow Anesthesia Anesth. Analg., October 1, 2004; 99(4): 1159 - 1163. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. R. Kennedy, R. A. French, and C. Spencer Predictive Accuracy of a Model of Volatile Anesthetic Uptake Anesth. Analg., December 1, 2002; 95(6): 1616 - 1621. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. G. C. Lerou, R. Verheijen, and L. H. D. J. Booij Model-based administration of inhalation anaesthesia. 4. Applying the system model Br. J. Anaesth., February 1, 2002; 88(2): 175 - 183. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. G. C. Lerou and L. H. D. J. Booij Model-based administration of inhalation anaesthesia. 3. Validating the system model Br. J. Anaesth., January 1, 2002; 88(1): 24 - 37. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. G. C. Lerou and L. H. D. J. Booij Model-based administration of inhalation anaesthesia. 2. Exploring the system model Br. J. Anaesth., January 1, 2001; 86(1): 29 - 37. [Abstract] [Full Text] [PDF] |
||||