Skip Navigation

This Article
Right arrow Full Text (PDF)
Right arrow E-Letters: Submit a response to the article
Right arrow Alert me when this article is cited
Right arrow Alert me when E-letters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrowRequest Permissions
Right arrow Disclaimer
Google Scholar
Right arrow Articles by JACOBSON, L.
Right arrow Articles by GREENBAUM, R.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by JACOBSON, L.
Right arrow Articles by GREENBAUM, R.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?

British Journal of Anaesthesia, 1981, Vol. 53, No. 1 97-101
© 1981 The Board of Management and Trustees of the British Journal of Anaesthesia


research-article

A STUDY OF INTRACUFF PRESSURE MEASUREMENTS, TRENDS AND BEHAVIOUR IN PATIENTS DURING PROLONGED PERIODS OF TRACHEAL INTUBATION

L. JACOBSON, F.F.A.R.C.S.* and R. GREENBAUM, F.F.A.R.C.S.

Department of Anaesthetics, Frenchay Hospital Bristol BS16 1LE
*Present address: Department of Anaesthesia, Royal Victoria Infirmary Newcastle-upon-Tyne NE1 4LP

Intracuff pressure (c.p.) was measured continuously in 12 patients during prolonged periods of tracheal intubation. High residual volume-low pressure (floppy) and low residual volume-high pressure (conventional) cuffs were studied. The mean baseline c.p. at minimal occluding pressure was significantly smaller in floppy cuffs than in conventional cuffs. With coughing, the peak pressures in both cuff types were much greater than the baseline pressures but did not differ significantly from each other. On the basis of our observations, we recommend a simplification of the test methods for cuffs and tube collapse. We believe universal testing to 300 mm Hg (40 kPa) to be reasonable and to incorporate a realistic safety factor. Intracuff pressure tends to diminish with time, but there is a poor correlation between the rate of pressure decrease and elapsed time. As this decrease is not accurately predictable, gas should be added to the cuff as required, rather than at regular intervals. Other features of cuff behaviour are discussed and suggestions made to explain some anomalies occasionally observed in the cuff pressure behaviour pattern.


Add to CiteULike CiteULike   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us    What's this?




Disclaimer: Please note that abstracts for content published before 1996 were created through digital scanning and may therefore not exactly replicate the text of the original print issues. All efforts have been made to ensure accuracy, but the Publisher will not be held responsible for any remaining inaccuracies. If you require any further clarification, please contact our Customer Services Department.