British Journal of Anaesthesia, 1978, Vol. 50, No. 9 959-964
© 1978 The Board of Management and Trustees of the British Journal of Anaesthesia
other |
TRACHEAL TUBES AND CONNECTORS USED IN NEONATES-DIMENSIONS AND RESISTANCE TO BREATHING
Departments of Anaesthesia and Respiratory Function, The Hospital for Sick Children Great Ormond Street, London WC1N 3JH.
The dimensions of tracheal tubes and connectors in common use were measured together with the resistance to breathing at flow rates occurring in babies breathing quietly. The outside diameters of rubber tubes varied more than those of plastic tubes. The main factors influencing resistance were the internal diameter and a sudden change in diameter or direction of flow. These are of particular importance in tubes of 2.5 mm i.d. and less and may cause an increase in the work of breathing during spontaneous ventilation. The resistance of small Cole pattern tubes was greater than that of plain tubes of similar o.d.
![]()
CiteULike
Connotea
Del.icio.us What's this?
This article has been cited by other articles:
![]() |
J. Kattwinkel, S. Niermeyer, V. Nadkarni, J. Tibballs, B. Phillips, D. Zideman, P. Van Reempts, and M. Osmond ILCOR Advisory Statement: Resuscitation of the Newly Born Infant : An Advisory Statement From the Pediatric Working Group of the International Liaison Committee on Resuscitation Circulation, April 13, 1999; 99(14): 1927 - 1938. [Full Text] [PDF] |
||||
![]() |
J. Kattwinkel, S. Niermeyer, V. Nadkarni, J. Tibballs, B. Phillips, D. Zideman, P. Van Reempts, and M. Osmond An Advisory Statement From the Pediatric Working Group of the International Liaison Committee on Resuscitation Pediatrics, April 1, 1999; 103(4): 56e - 56. [Abstract] [Full Text] |
||||